TheraSphere Administration Set 990226.SPERecalls & reported problems

Biocompatibles Uk Ltd · Administration Set to be used for infusion of TheraSphere Y90 Glass Microspheres · Class 3 · Implant

All models of TheraSphere Administration Set (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05060116920253), received between Oct 20, 2021 and Aug 12, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Radiation Underexposure” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 23, 2021Class IIOpen, Classified

    Potential for leaks and loosening at the patient catheter connection

    Z-2327-2021 · Root cause: Component change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Radiation Underexposure3
  • Device Contamination with Chemical or Other Material2
  • Adverse Event Without Identified Device or Use Problem1
  • Contamination1
  • Infusion or Flow Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Vascular Dissection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2025Contamination; Device Contamination with Chemical or Other Material; Radiation Underexposure; Contamination; Device Contamination with Chemical or Other Material; Radiation Underexposure

    IT WAS REPORTED THAT FOREIGN MATERIAL WAS OBSERVED IN THE TUBING. TWO THERASPHERE ADMINISTRATION SETS WERE SELECTED FOR USE IN THE DELIVERY OF 6.5GBQ AND 10.5 GBQ THERASPHERE DOSE VIALS IN A Y-90 PROCEDURE. DURING THE FIRST ADMINISTRATION, ALL THE PROPER STEPS WERE FOLLOWED OFF THE CHECKLIST; HOWEVER, WHEN INJECTING, THE PRESSURE VIAL IMMEDIATELY STARTED TO FILL AND A BROWN GLOB WAS OBSERVED WORKING ITS WAY THROUGH THE TUBING. THE PHYSICIAN CONTINUED FLUSHING UNTIL THE RADOS DOSIMETER REACHED 0.0 MR/H. BEFORE INJECTING THE SECOND DOSE INTO ANOTHER SEGMENT OF THE LIVER, THE NON-BOSTON SCIENTIFIC MICROCATHETER WAS CHECKED FOR KINKS, BUT NONE WERE OBSERVED. THE CHECKLIST WAS FOLLOWED AGAIN FOR THE SECOND DOSE ADMINISTRATION, BUT THE SAME THING HAPPENED; HOWEVER, THIS TIME A GRAY/BLACK SPECK WAS OBSERVED WORKING ITS WAY THROUGH THE TUBING. THERE WAS A CLOUDY RESIDUE IN THE TUBING AFTER EACH ADMINISTRATION, NEAR THE NEEDLE INJECTOR. THERE WAS A LOW PERCENT DELIVERY OF 86% FOR BOTH DOSES. THE PROCEDURE WAS COMPLETED WITH THESE DEVICES. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22787426 ↗
  • MalfunctionAug 12, 2025Device Contamination with Chemical or Other Material; Radiation Underexposure; Device Contamination with Chemical or Other Material; Radiation Underexposure

    IT WAS REPORTED THAT FOREIGN MATERIAL WAS OBSERVED IN THE TUBING. TWO THERASPHERE ADMINISTRATION SETS WERE SELECTED FOR USE IN THE DELIVERY OF 6.5GBQ AND 10.5 GBQ THERASPHERE DOSE VIALS IN A Y-90 PROCEDURE. DURING THE FIRST ADMINISTRATION, ALL THE PROPER STEPS WERE FOLLOWED OFF THE CHECKLIST; HOWEVER, WHEN INJECTING, THE PRESSURE VIAL IMMEDIATELY STARTED TO FILL AND A BROWN GLOB WAS OBSERVED WORKING ITS WAY THROUGH THE TUBING. THE PHYSICIAN CONTINUED FLUSHING UNTIL THE RADOS DOSIMETER REACHED 0.0 MR/H. BEFORE INJECTING THE SECOND DOSE INTO ANOTHER SEGMENT OF THE LIVER, THE NON-BOSTON SCIENTIFIC MICROCATHETER WAS CHECKED FOR KINKS, BUT NONE WERE OBSERVED. THE CHECKLIST WAS FOLLOWED AGAIN FOR THE SECOND DOSE ADMINISTRATION, BUT THE SAME THING HAPPENED; HOWEVER, THIS TIME A GRAY/BLACK SPECK WAS OBSERVED WORKING ITS WAY THROUGH THE TUBING. THERE WAS A CLOUDY RESIDUE IN THE TUBING AFTER EACH ADMINISTRATION, NEAR THE NEEDLE INJECTOR. THERE WAS A LOW PERCENT DELIVERY OF 86% FOR BOTH DOSES. THE PROCEDURE WAS COMPLETED WITH THESE DEVICES. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22787425 ↗
  • MalfunctionOct 20, 2021Infusion or Flow Problem; Adverse Event Without Identified Device or Use Problem; Radiation Underexposure; Infusion or Flow Problem; Adverse Event Without Identified Device or Use Problem; Radiation Underexposure

    IT WAS REPORTED THAT A PROCEDURAL RELATED DISSECTION OCCURRED. IN (B)(6) 2020, A MISADMINISTRATION OF THERASPHERE Y-90 MICROSPHERES OCCURRED. THE PHYSICIAN WAS UNABLE TO INJECT THE THERASPHERE FROM THE DELIVERY APPARATUS INTO THE PATIENT. SEVEN DAYS LATER, A REPEAT PROCEDURE WAS ATTEMPTED, HOWEVER, THE INJECTION ATTEMPT FAILED. DURING THIS PROCEDURE, STENOSIS IN THE FEMORAL ARTERY WAS IDENTIFIED. THE PATIENT WAS GIVEN BLOOD THINNERS. THE STENOSIS IS THOUGHT TO BE EITHER THE DISSECTION OF THE ARTERIAL WALL FROM THE NON-BSC CLOSURE DEVICE, OR IT WAS THE DISPLACED CLOSURE DEVICE ITSELF FROM THE FIRST THERASPHERE PROCEDURE. IN LATE (B)(6) 2020, THE PATIENT REPORTED DIFFICULTY IN WALKING, THEIR LEG WOULD GO NUMB AND ESSENTIALLY FEEL DEAD AND HEAVY. THE PATIENT SOUGHT MEDICAL ADVICE. AFTER IMAGING OF THE ARTERY WAS COMPLETED, THE PATIENT LEARNED THAT THE CLOSURE DEVICE HAD COLLAPSED INTO THE ARTERY AND THE ARTERIAL WALL SUFFERED A DISSECTION. IN (B)(6) 2021, A SURGEON WHO SPECIALIZES IN ARTERIAL SURGERY ATTEMPTED TO REPAIR THE DISSECTED ARTERIAL WALL AND REMOVE THE COLLAPSED ANGIOSEAL DEVICE, HOWEVER, THE PATIENT STILL EXPERIENCES LEG NUMBNESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Vascular Dissection

    MAUDE report 12668230 ↗