Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05060116920420), received between Aug 28, 2025 and Aug 28, 2025 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Abdominal Pain1
  • High Blood Pressure/ Hypertension1
  • Unintended Radiation Exposure1
  • Vomiting1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED ABDOMINAL PAIN AND VOMITING REQUIRING PRESCRIPTION MEDICATION AS A RESULT OF THE THERASPHERE PROCEDURE. THE PATIENT WAS BEING TREATED FOR UVEAL MELANOMA. A THERASPHERE Y-90 (8.5 GBQ) DOSE VIAL WAS SELECTED FOR USE IN THE RADIOEMBOLIZATION PROCEDURE. THE MULTIFOCAL, SMALL LESIONS WERE LOCATED IN THE BILOBAR LIVER. THE TOTAL TUMOR VOLUME WAS APPROXIMATELY 7.4ML. THERE WAS NOTHING UNUSUAL OBSERVED DURING DEVICE PREPARATION OR THE THERASPHERE PROCEDURE. POST PROCEDURE IMAGING REVEALED THE EXPECTED ACCUMULATION OF MICROSPHERES WITHIN THE LIVER, AS WELL AS ACCUMULATION WITHIN THE AREA OF THE GALLBLADDER. FOLLOWING THE THERASPHERE PROCEDURE, THE PATIENT EXPERIENCED TRANSITORY UPPER ABDOMINAL PAIN, VOMITING, AND A HYPERTENSIVE CRISIS. TO TREAT THE PAIN AND VOMITING, THE PATIENT WAS ADMINISTERED NOVALGIN AND ODANSETRON (ZOFRAN). THE PATIENT WAS ALSO ADMINISTERED ANTIBIOTIC CHOLECYSTITIS PROPHYLAXIS WITH UNACID. THE PAIN WAS REPORTED TO HAVE RESOLVED AFTER ONE DAY. THERE WAS NO REPORTED INTERVENTION FOR THE HYPERTENSIVE CRISIS. THERE WERE NO FURTHER PATIENT COMPLICATIONS.

    Patient problems: Abdominal Pain; High Blood Pressure/ Hypertension; Unintended Radiation Exposure; Vomiting

    MAUDE report 22916931 ↗

Other Microspheres Radionuclide devices

All Microspheres Radionuclide devices →