OrganOx metra D0146Recalls & reported problems

Organox Limited · The OrganOx metra® is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours. The OrganOx metra® device is suitable for liver grafts from donors after brain death (DBD), or liver grafts from donors after circulatory death (DCD) less than or equal to 40 years old, with less than or equal to 20 mins of functional warm ischemic time (time from donor systolic blood pressure less than 50 mmHg), and macrosteatosis less than or equal to 15%, in a near-physiologic, normothermic and functioning state intended for a potential transplant recipient. · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
14
Deaths reported
0
Injuries reported
8
Recalls naming it
1

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 05060462240029), received between Mar 1, 2024 and Nov 13, 2024 — classified as 8 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 4 reportsApr 2024: 4 reportsMay 2024: 4 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 15, 2024Class IIOpen, Classified

    Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket

    Z-0501-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak4
  • Complete Blockage2
  • Contamination2
  • Insufficient Device Problem Information2
  • Failure to Read Input Signal1
  • Low Readings1
  • Restricted Flow rate1
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • Liver Damage/Dysfunction6
  • No Clinical Signs, Symptoms or Conditions6
  • Thrombosis/Thrombus2

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 13, 2024Complete Blockage; Complete Blockage

    THE DEVICE USER REPORTED THAT THEY HAD A LIVER GOING ON PUMP OVERNIGHT. AT 327 AM EST THE DEVICE USER SENT A MESSAGE STATING "LIVER IS BEING DISCARDED. THERE WAS A DEFECT IN THE RECIRCULATING PUMP LINE THAT PREVENTED THE PUMP FROM REFILLING THE RESERVOIR BAG. THAT COMBINED WITH A HEAVY BLEEDER PREVENTED THE LIVER FROM GETTING PROPER PERFUSION." THE DEVICE USER FURTHER DESCRIBED THAT THE TUBING UNDER THE LIVER BOWL WAS SEALED SHUT AND SINCE IT WAS UNDER THE LIVER BOWL, THEY COULD NOT FIGURE OUT WHY THE RECIRCULATING PUMP WAS SLOW. BY THE TIME THE AREA WAS LOCATED, IT HAD BEEN A LONG TIME WITH AN EMPTY RESERVOIR. IT WAS NOTED THAT THE SEAL TOOK SEVERAL MINUTES TO BREAK AND THE TUBING WAS ALMOST COMPLETELY OCCLUDED AS WELL.

    Patient problems: Liver Damage/Dysfunction

    MAUDE report 20679745 ↗
  • InjuryJul 18, 2024Restricted Flow rate; Restricted Flow rate

    THE DEVICE USER (DU) REPORTED THAT THERE WAS LITTLE TO NO FLOW NOTED AFTER THE LIVER WAS CONNECTED TO THE DEVICE. DURING TROUBLESHOOTING A DARKENED AREA AT THE TOP OF THE SOFT-SHELL RESERVOIR WAS NOTED AND AIR WAS OBSERVED IN THE TUBING. THE DECISION WAS MADE TO REMOVE THE LIVER AND COLD FLUSH TO AVOID WARM ISCHAEMIA TIME. FOLLOWING ADDITIONAL TROUBLESHOOTING, THE DU DISCOVERED A CLOT IN THE PORTAL CANNULA. THE DEVICE WAS CLEARED OF AIR AND PREPARATION MODE WAS ACHIEVED. IT WAS CONFIRMED THAT THE DU USED THE RECOMMENDED INFUSIONS, BOLUS DRUGS & PRIMING STEPS. ABOUT ONE HOUR LATER THE DU REPORTED THAT THE DEVICE ENTERED LOW RESERVOIR MODE. THIS WAS RESOLVED BY MANIPULATING THE RECIRCULATION PUMP SENSOR RESULTING IN THE VOLUME IN THE BOWL RECIRCULATING INTO THE RESERVOIR. ULTIMATELY, THE DU DECIDED TO DISCARD THE DONOR LIVER DUE TO FAILURE TO CLEAR LACTATE AND FAILURE TO MAINTAIN PH EVEN THOUGH THE FLOWS WERE FINE ONCE THE CLOT WAS REMOVED.

    Patient problems: Liver Damage/Dysfunction

    MAUDE report 19780106 ↗
  • MalfunctionMay 15, 2024Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    DEVICE USER (DU) REPORTED THAT THERE WAS A SMALL HOLE ON THE PLASTIC SIDE OF THE VASCULAR CANNULAS PACKAGING. AS ARESULT, THE ITEMS WERE DEEMED UNSTERILE AND ANOTHER ANOTHER KIT WAS OPENED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19323793 ↗
  • MalfunctionMay 15, 2024Fluid/Blood Leak; Fluid/Blood Leak

    DEVICE USER (DU) REPORTED THE DEVICE HAD A SMALL LEAK AT THE OXYGENATOR. A VIDEO OF THE LEAK AND ADDITIONAL PHOTOS OF THE TOP OF THE OXYGENATOR INDICATED THAT IT WAS COMPROMISED. A CLINICAL SPECIALIST (CS) ADVISED THE DU TO RESOLVE THE ISSUE BY CHANGING THE DISPOSABLE SET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19320525 ↗
  • InjuryMay 14, 2024Complete Blockage; Complete Blockage

    THE DEVICE USER REPORTED THAT MESSAGE CODE 270 (OCCLUDED TUBING) DISPLAYED ON THE DEVICE SCREEN. THE DEVICE USER WAS ABLE TO DETERMINE THAT A SYRINGE INFUSION LINE FOR HEPARIN WAS BEING OCCLUDED BY A PLASTIC FILM IN TWO LOCATIONS. THE FIRST LOCATION WAS WHERE THE TIP OF THE SYRINGE IS SCREWED INTO THE LUER LOCK. THERE WAS PLASTIC FILM THAT WAS OCCLUDING THE SYRINGE LINE FROM BEING PRIMED. THE SECOND LOCATION WAS FURTHER UP THE INFUSION LINE WHERE THE INFUSION TUBING WAS STUCK TOGETHER. IT WAS DECIDED TO DISCARD THE DONOR LIVER DUE TO LACK OF VIABILITY. THE DEVICE USER BELIEVES THAT THE OCCULDED TUBING CONTRIBUTED TO THE LIVER NOT BEING VIABLE FOR TRANSPLANTATION.

    Patient problems: Liver Damage/Dysfunction

    MAUDE report 19315704 ↗
  • MalfunctionMay 13, 2024Contamination; Contamination

    IT WAS REPORTED THAT A HAIR STRAND WAS FOUND INSIDE THE STERILE PACKAGING OF ONE OF THE SILICONE BILE TUBES. IT WAS INSIDE THE SECOND PACKET, ATTACHED TO THE TUBE. THE DEVICE WAS USED DURING A TRAINING CASE SO IT WAS NOT OF CONCERN FROM A PATIENT PERSPECTIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19307034 ↗
  • MalfunctionApr 26, 2024Fluid/Blood Leak; Fluid/Blood Leak

    DEVICE USER (DU) NOTICED PERFUSATE LEAKING AT THE OXYGENATOR CONNECTION. PRIMING WAS PAUSED WHILE THE CONNECTION WAS ASSESSED. DU WAS ABLE TO MANUALLY TIGHTEN THE CONNECTION WITH INITIAL RESOLUTION. PRIMING WAS THEN RESUMED BUT THE LEAK CONTINUED. THE DISPOSABLE SET WAS CHANGED. NO FURTHER ISSUE OCCURRED AFTER A NEW DISPOSABLE SET WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19192139 ↗
  • MalfunctionApr 25, 2024

    WHILE PREPARING THE ORGAN OX PUMP FOR LIVER DONOR ORGAN, IT WAS NOTED THAT THERE WAS A FAULTY CONNECTION FROM THE ONE TIME USE CASSETTE. A SLOW LEAK AT THE LEVEL OF THE OXYGENATOR FROM THE CANNULA ORIGINATING FROM THE LIVER BOWL NOT RESPONSIVE TO MANUAL TIGHTENING WAS DISCOVERED UPON INFUSION OF PERFUSATE. PERFUSATE WAS THEN INFUSED BACK INTO BAGS USING STERILE TECHNIQUE UNDER GRAVITY AND A NEW CASSETTE WAS INSTALLED AND CONTINUED PREP AS NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19180738 ↗
  • InjuryApr 4, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    A SURGEON REPORTED THAT "UPON REMOVAL OF THE ARTERIAL CANNULA, THERE WAS A THROMBOSIS. THE LIVER WAS STILL TRANSPLANTED, BUT IT DID INCREASE THE TIME ON ICE BY APPROXIMATELY 30 MINUTES".

    Patient problems: Thrombosis/Thrombus

    MAUDE report 19043401 ↗
  • InjuryApr 4, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    A SURGEON REPORTED THAT "UPON REMOVAL OF THE ARTERIAL CANNULA, THERE WAS A THROMBOSIS. THE LIVER WAS STILL TRANSPLANTED, BUT IT DID INCREASE THE TIME ON ICE BY APPROXIMATELY 30 MINUTES".

    Patient problems: Thrombosis/Thrombus

    MAUDE report 19040953 ↗
  • MalfunctionMar 20, 2024Contamination; Contamination

    THE DEVICE USER REPORTED VISUALIZING A PIECE OF HAIR IN THE CANNULA PACKAGING. A CLINICAL SPECIALIST INFORMED THE DEVICE USER TO USE A NEW SET OF CANNULAS FROM A NEW DISPOSABLE SET SO THAT THE PROCEDURE COULD BE COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18946471 ↗
  • InjuryMar 13, 2024Fluid/Blood Leak; Fluid/Blood Leak

    A LIVER WAS PLACED ON THE DEVICE. THE DEVICE USER CONTACTED A CLINICAL SPECIALIST (CS) SUBSEQUENTLY BECAUSE A LEAK WAS OBSERVED UNDERNEATH THE LIVER BOWL APPROXIMATELY 20 MINUTES INTO THE PERFUSION. THE CS ADVISED THE DEVICE USER TO PUT BONE WAX OVER THE AREA AND REPLACE ZIP TIE ON THE CONNECTION, BUT THIS WAS UNSUCCESSFUL. ALL THE BLOOD LEAKED OUT OF THE PUMP ONTO THE DEVICE. ATTEMPT TO COLD FLUSH THE LIVER WAS UNSUCCESSFUL. DEVICE USER ALLEGED THAT THE DEVICE FAILED. LIVER WAS DISCARDED.

    Patient problems: Liver Damage/Dysfunction

    MAUDE report 18896427 ↗
  • InjuryMar 13, 2024Fluid/Blood Leak; Fluid/Blood Leak

    A LIVER WAS PLACED ON THE DEVICE. THE DEVICE USER CONTACTED A CLINICAL SPECIALIST (CS) SUBSEQUENTLY BECAUSE A LEAK WAS OBSERVED UNDERNEATH THE LIVER BOWL APPROXIMATELY 20 MINUTES INTO THE PERFUSION. THE CS ADVISED THE DEVICE USER TO PUT BONE WAX OVER THE AREA AND REPLACE ZIP TIE ON THE CONNECTION, BUT THIS WAS UNSUCCESSFUL. ALL THE BLOOD LEAKED OUT OF THE PUMP ONTO THE DEVICE. ATTEMPT TO COLD FLUSH THE LIVER WAS UNSUCCESSFUL. DEVICE USER ALLEGED THAT THE DEVICE FAILED. LIVER WAS DISCARDED.

    Patient problems: Liver Damage/Dysfunction

    MAUDE report 18894325 ↗
  • InjuryMar 1, 2024Failure to Read Input Signal; Low Readings; Failure to Read Input Signal; Low Readings

    DEVICE USER REPORTED THAT THE DEVICE FAILED TO GO INTO LIVER ON BOARD MODE AFTER A DONOR LIVER WAS PLACED ON THE DEVICE. AS A RESULT, THE MARGINAL LIVER WAS DISCARDED.

    Patient problems: Liver Damage/Dysfunction

    MAUDE report 18821974 ↗

Other Normothermic Machine Perfusion System For The Preservation Of Donor Livers Prior To Transplantation devices

All Normothermic Machine Perfusion System For The Preservation Of Donor Livers Prior To Transplantation devices →