Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI B00950480), received between Nov 14, 2025 and Nov 14, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Infusion or Flow Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2025Infusion or Flow Problem; Infusion or Flow Problem

    EVENT DESCRIPTION THE OCS LIVER CONSOLE EXHIBITED AN ABNORMAL CLICKING NOISE DURING THE CASE, FOLLOWED BY A SUDDEN AND COMPLETE LOSS OF PUMP FLOW. ATTEMPTS TO INCREASE FLOW WERE UNSUCCESSFUL. THE DEVICE WAS RESET MULTIPLE TIMES, AND THE MODULE WAS REMOVED AND REINSTALLED. NORMAL FUNCTION WAS RESTORED AFTER REINSTALLATION. THE OCS FUNCTIONED NORMALLY AND THE LACTATE LEVELS TRENDED DOWN AND REMAINED LOW FOR REMAINDER OF THE CASE. THE INTERRUPTION LASTED APPROXIMATELY 21 MINUTES. THE LIVER WAS NOT TRANSPLANTED. BIOPSY FINDINGS INDICATED FOCAL APOPTOSIS - 10% FIBROSIS (STAGE 2- LEFT, STAGE 1- RIGHT), MACROSTEOTOSIS 30% (BOTH LOBES). APOPTOSIS MAY HAVE BEEN ASSOCIATED WITH THE REPORTED MALFUNCTION. INVESTIGATION SUMMARY THE REPORTED PUMP FAILURE WAS CONFIRMED. ROOT CAUSE ANALYSIS DETERMINED THAT THE FAILURE RESULTED FROM INADEQUATE ADHESION OF THE RETAINING COMPOUND BETWEEN THE BALL SCREW AND THE MOTOR SHAFT. THIS CONDITION ALLOWED THE BALL SCREW TO ROTATE FREELY, PREVENTING PROPER PUMP OPERATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23560410 ↗

Other Normothermic Machine Perfusion System For The Preservation Of Donor Livers Prior To Transplantation devices

All Normothermic Machine Perfusion System For The Preservation Of Donor Livers Prior To Transplantation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page