NovoTHOR XL 3.0Recalls & reported problems
Thor Photomedicine Limited · S2190 NovoTHOR XL · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05060494130510), received between Aug 4, 2026 and Aug 4, 2026 — classified as 1 injury.
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 24, 2025Class IIOpen, Classified
The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
Z-2417-2025 · Root cause: Process control · Recall details · FDA record ↗
- Nov 20, 2023Class IIOpen, Classified
One of the ball studs attaching a gas strut to the canopy may fail. There is potential risk of injury should a ball stud fail when someone is in the NovoTHOR and trying to open or close the canopy
Z-0972-2024 · Root cause: Component design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Skin Inflammation/ Irritation1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 4, 2026Adverse Event Without Identified Device or Use Problem
A patient attended a UK clinic for treatment using the NovoTHOR whole body photobiomodulation system for management of severe eczema symptoms.In (b)(6) 2025 the patient underwent two 20-minute treatment sessions. Following these sessions the patient reported that her skin felt hot and irritated. The symptoms settled shortly afterwards. The patient noted that other factors may have coincided with these symptoms.In (b)(6) 2026 the patient returned for further treatment. At that time she had unstable eczema and had recently received systemic treatment for this condition. She underwent two 12-minute treatment sessions which were reported to aggravate facial skin symptoms.During a third 12-minute session the patient reported a sensation of heat and skin agitation within approximately one minute of entering the device but chose to continue the session for approximately eight minutes. Following this session the patient reported a marked flare of skin symptoms including redness, itching and burning sensations. The patient reported that the discomfort required analgesic medication to allow sleep.The patient subsequently consulted a dermatologist and was prescribed oral steroids. The symptom
Patient problems: Skin Inflammation/ Irritation
MAUDE report 50125818 ↗
FDA clearance and classification
The FDA regulates product code ILY (lamp, infrared, therapeutic heating) under 21 CFR 890.5500 as a Class II device. 21 CFR 890.5500 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NovoTHOR XL 3.0
Has the NovoTHOR XL 3.0 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2417-2025 (Class II), began on Jul 24, 2025: The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the NovoTHOR XL 3.0?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Aug 4, 2026. That report lists patient problem “Skin Inflammation/ Irritation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NovoTHOR XL 3.0?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NovoTHOR XL 3.0 FDA approved?
Devices of this type (product code ILY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Thor Photomedicine Limited.
Who makes the NovoTHOR XL 3.0, and is it still sold?
The FDA's device database lists Thor Photomedicine Limited as the company responsible for it (brand NovoTHOR XL, model 3.0). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Lamp, Infrared, Therapeutic Heating devices
- NovoTHOR 3.0Thor Photomedicine Limited · Lamp, Infrared, Therapeutic Heating2 recalls0 reports
- NovoTHOR PRX 3.0Thor Photomedicine Limited · Lamp, Infrared, Therapeutic Heating1 recall0 reports
- NovoTHOR Gen 3.0 Reg PRXThor Photomedicine Limited · Lamp, Infrared, Therapeutic Heating1 recall0 reports
- CHATTANOOGA 27812Djo, LLC · Lamp, infrared, therapeutic heating3 reports
- 810nm 1W Cluster (5x200mW) 2.0Thor Photomedicine Limited · Lamp, Infrared, Therapeutic Heating3 reports
- CHATTANOOGA 27840Djo, LLC · Lamp, infrared, therapeutic heating2 reports
- LightForce FXi LTC-1500Litecure LLC · Lamp, Infrared, Therapeutic Heating1 report
- 660nm 75mW Dental Probe 2.3Thor Photomedicine Limited · Lamp, Infrared, Therapeutic Heating1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page