Why it was recalled
One of the ball studs attaching a gas strut to the canopy may fail. There is potential risk of injury should a ball stud fail when someone is in the NovoTHOR and trying to open or close the canopy
FDA root cause category: Component design/selection
What the company is doing
On November 20, 2023, the firm notified customers via IMPORTANT MEDICAL DEVICE CORRECTION letters sent through email. A service agent representing THOR will contact customers within the next eight weeks to schedule the fitting of replacement ball studs at no cost to the customer. Customers who continue to use the NovoTHOR should be aware that it may potentially cause harm should the failure occur while users/patients are lifting or lowering the canopy. Customers should warn all users of potential risk, ensure someone is present to lift and lower the canopy, ensure the canopy is lifted and lowered gently and slowly, and ensure someone is present in the room or immediately outside the room during use.
Recall details
- Recall number
- Z-0972-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 20, 2023
- Posted by the FDA
- Feb 1, 2024
- Recalling firm
- THOR Photomedicine Ltd
- Quantity
- 61 (41 US; 20 OUS)
- Product code
- ILY
- Distribution
- Domestic distribution to the states of Arizona California Colorado Connecticut Florida Georgia Hawaii Idaho Indiana Maryland New Jersey New York Oregon Pennsylvania Texas Washington Wisconsin. Foreign distribution to Australia, Canada, Italy, Malta, Norway, Switzerland, UAE, and UK.
Product description
NovoTHOR Whole Body Light Pod, Gen 3.0 (1) Product Code S2188 - Reg (2) Product Code S2189 - Reg PRX (3) Product Code S2190 - XL
Code information (models, lots, serial numbers)
(1) UDI-DI 05060494130503 Serial Numbers 4004 4009 4013 4017 4019 4021 4024 4026 4028 4029 4035 4037 4042 4043 4044 4046 4047 4048 4054 4065 4066 4072 4073 4074 4076 4077 4078 4079 4080 4081 4082 4083 4085 4086 4087 4090 4091 4096 (2) UDI-DI 05060494130527 Serial Number 4068 (3) UDI-DI 05060494130510 Serial Numbers 4002 4003 4006 4010 4012 4027 4030 4040 4038 4023 4045 4051 4052 4060 4057 4064 4056 4069 4063 4070 4089 4014
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- NovoTHOR 3.0Thor Photomedicine Limited · Lamp, Infrared, Therapeutic Heating2 recalls0 reports
- NovoTHOR XL 3.0Thor Photomedicine Limited · Lamp, Infrared, Therapeutic Heating2 recalls0 reports
- NovoTHOR PRX 3.0Thor Photomedicine Limited · Lamp, Infrared, Therapeutic Heating1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page