Sentry IVC Filter 146155-01Recalls & reported problems
Novate Medical Limited · The Sentry IVC Filter is indicated for the prevention of recurrent Pulmonary Embolism via percutaneous placement in the inferior vena cava in patients at transient risk of PE, in the following situations: • Pulmonary thromboembolism when anticoagulants are contraindicated. • Failure of anticoagulant therapy in thromboembolic diseases. • Emergency treatment following massive PE where anticipated benefits of conventional therapy are reduced. · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05391529380069), received between Apr 25, 2019 and Aug 7, 2020 — classified as 3 injuries.
The most frequently reported specific device problem is “Difficult to Open or Close” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Difficult to Open or Close1
- Malposition of Device1
- Use of Device Problem1
Patient problems
- No Code Available2
- Perforation of Vessels1
- Thrombosis1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 7, 2020Use of Device Problem; Malposition of Device; Difficult to Open or Close; Use of Device Problem; Malposition of Device; Difficult to Open or Close
IT WAS REPORTED THAT A PATIENT WAS IMPLANTED WITH A SENTRY IVC FILTER FOR DVT THROUGH THE RIGHT COMMON FEMORAL VEIN. A VENOGRAM WAS PERFORMED PRIOR TO PLACEMENT. THE DEVICE WAS DEPLOYED IN THE IVC, INFRA RENAL AND THE DISTAL PORTION OF THE FILTER DID NOT OPEN AS EXPECTED. THE NEXT DAY A CTA WAS PERFORMED AND REVEALED THAT THE DISTAL PART OF THE FILTER WAS OUTSIDE OF THE IVC. THE PATIENT IS STABLE AND THE PHYSICIAN IS PLANNING POSSIBLE INTERVENTION. IT WAS FURTHER REPORTED THAT THIS EVENT WAS A RESULT OF USE ERROR AND NOT DEVICE FAILURE. THE PHYSICIAN DID NOT USE A GUIDEWIRE WHILE ADVANCING THE SHEATH 3CM FORWARD RIGHT BEFORE THE FILTER WAS DEPLOYED. THE MOVEMENT OF THE SHEATH RESULTED IN PERFORATION OF THE IVC. VENOGRAM IMAGING SHOWS THE TOP CROWN WAS CONSTRAINED DUE TO BEING EXTRAVASCULAR. THE PATIENT WAS TREATED AT A DIFFERENT INSTITUTION WHERE THE FILTER WAS SUCCESSFULLY EXPLANTED.
Patient problems: No Code Available; Perforation of Vessels
MAUDE report 10382129 ↗InjuryNov 13, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS EVENT IS DERIVED FROM A LITERATURE REVIEW OF FILTER-RELATED COMPLICATIONS PRESENTED IN "INFERIOR VENA CAVA FILTER EVALUATION AND MANAGEMENT FOR THE DIAGNOSTIC RADIOLOGIST: A COMPREHENSIVE REVIEW INCLUDING INFERIOR VENA CAVA FILTER-RELATED COMPLICATIONS AND PRESERVE TRIAL FILTERS" ARTICLE BY AHMED ET AL 2019. IN FILTER-ASSOCIATED COMPLICATIONS SECTION, A CORONAL CONTRAST ENHANCED COMPUTED TOMOGRAPHY IMAGE DEMONSTRATES ACUTE THROMBOTIC OCCLUSION OF A BTG SENTRY CONVERTIBLE FILTER. THROMBUS IS ALSO NOTED EXTENDING CEPHALAD TO THE CONICAL APEX OF THE FILTER.
Patient problems: Thrombosis
MAUDE report 9316015 ↗InjuryApr 25, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PHYSICIAN PERFORMED A VENOGRAM ON A PATIENT WHO HAD A SENTRY IVC FILTER PLACED PREVIOUSLY IN HIS LAB. THE INITIAL VERBATIM DESCRIPTION WAS; "THIS DEVELOPED CLOT IN THE FILTER AFTER IT PARTIALLY OPENED AND YES IT'S OPEN AND I DID SOME TPA". THE PHYSICIAN PROVIDED AN IMAGE OF THE VENOGRAM. THE VENOGRAM SHOWED THAT THE SENTRY HAD BIO-CONVERTED AND THAT THERE WAS A FILLING DEFECT (REPORTED AS CLOT) IN ADJACENT TO 2-3 OF THE FILTER'S ARMS. THE DATE THE SENTRY WAS PLACED WAS NOT REPORTED. DURING FOLLOW-UP CONVERSATIONS, THE PHYSICIAN STATED THAT, AS PER HIS PRACTICE, HE HAND INJECTS 4 MG OF TPA, IMPLANT FILTER, THEN DISCHARGE THE PATIENT WITH A PRESCRIPTION FOR ELIQUIS. IT IS NOT KNOWN IF THIS PATIENT FILLED THE PRESCRIPTION. THE INDICATION FOR FILTER PLACEMENT IN THIS PATIENT IS UNKNOWN. THE PHYSICIAN ALSO STATED THAT HE PERFORMS ROUTINE FOLLOW-UP FOR ALL OF HIS IVC FILTER PATIENT'S VENOGRAMS ON ALL OF HIS IVC FILTER PATIENTS WITHIN 2-3 POST-IMPLANT. THIS VENOGRAM WAS NOT PERFORMED DUE TO ANY NEW THROMBOEMBOLIC RISK FACTORS. A FOLLOW-UP VENOGRAM WAS PERFORMED ON (B)(6) 2019 VIA JUGULAR ACCESS. A CATHETER WAS PLACED THROUGH THE SENTRY IVC FILTER TO PERFORM THE VE
Patient problems: No Code Available
MAUDE report 8550275 ↗
FDA clearance and classification
The FDA regulates product code DTK (filter, intravascular, cardiovascular) under 21 CFR 870.3375 as a Class II device. 21 CFR 870.3375 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Sentry IVC Filter 146155-01
Has the Sentry IVC Filter 146155-01 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05391529380069), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Sentry IVC Filter 146155-01?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Apr 25, 2019 and Aug 7, 2020. In those reports, the most common specific device problems are “Difficult to Open or Close” (1), “Malposition of Device” (1) and “Use of Device Problem” (1); the most common patient problems are “Perforation of Vessels” (1) and “Thrombosis” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Sentry IVC Filter 146155-01?
None of the reports are classified as deaths, 3 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Sentry IVC Filter 146155-01 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Sentry IVC Filter 146155-01, and is it still sold?
The FDA's device database lists Novate Medical Limited as the company responsible for it (brand Sentry IVC Filter, model 146155-01). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Filter, Intravascular, Cardiovascular devices
- Denali® Vena Cava Filter DL900FBard Peripheral Vascular, Inc. · FILTER, INTRAVASCULAR, CARDIOVASCULAR22 deaths530 reports
- DENALI® Vena Cava Filter DL900JBard Peripheral Vascular, Inc. · FILTER, INTRAVASCULAR, CARDIOVASCULAR15 deaths491 reports
- Option Elite Retrievable Vena Cava Filter System 352506070EArgon Medical Devices, Inc. · Filter, Intravascular, Cardiovascular2 recalls3 deaths340 reports
- Gunther Tulip G33016Cook Incorporated · FILTER, INTRAVASCULAR, CARDIOVASCULAR3 deaths348 reports
- Gunther Tulip G33017Cook Incorporated · FILTER, INTRAVASCULAR, CARDIOVASCULAR3 deaths286 reports
- Cook Celect G34309William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall1 death230 reports
- Gunther Tulip G52916William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall2 deaths189 reports
- Cook Celect G34502William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall2 deaths165 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page