Victory® SRRecalls & reported problems
St. Jude Medical, Inc. · Pulse GeneratorSSIR · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05414734009539), received between Jul 30, 2018 and Apr 28, 2025 — classified as 4 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Pacemaker Found in Back-Up Mode” (6 reports).
9 other recalls cover devices cleared under the same FDA submission (P880086); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Pacemaker Found in Back-Up Mode6
- Intermittent Capture1
- Over-Sensing1
- Premature Discharge of Battery1
Patient problems
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient3
- Chest Pain1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 28, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT A PATIENT PRESENTED TO CLINIC FOR FOLLOW UP. DEVICE INTERROGATION REVEALED NON-SUSTAINED RIGHT VENTRICULAR OVERSENSING EPISODES DUE TO POST-PACED T-WAVE OVERSENSING ON THE PACEMAKER (PM). PROGRAMMING CHANGES WERE MADE TO RESOLVE THE ISSUE. THE PATIENT CONDITION WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21922297 ↗MalfunctionNov 6, 2020Premature Discharge of Battery; Premature Discharge of Battery
ADDITIONAL INFORMATION INDICATED THE DEVICE WAS EXPLANTED AND REPLACED TO RESOLVE THE EVENT. THE PATIENT WAS IN STABLE CONDITION FOLLOWING THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10804656 ↗InjuryNov 5, 2020Pacemaker Found in Back-Up Mode; Pacemaker Found in Back-Up Mode
ADDITIONAL INFORMATION RECEIVED INDICATED THE DEVICE WAS EXPLANTED AND REPLACED TO RESOLVE THE EVENT. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10793053 ↗InjuryOct 5, 2020Pacemaker Found in Back-Up Mode; Pacemaker Found in Back-Up Mode
IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED TO A FOLLOW-UP IN-CLINIC WITH THEIR PACEMAKER EXHIBITING BACKUP VVI MODE. IT WAS SUSPECTED TO BE CAUSED BY THE DEVICE REACHING END OF LIFE. DEVICE WAS EXPLANTED AND REPLACED ON (B)(6) 2020. PATIENT WAS STABLE AFTER THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10632457 ↗InjuryNov 1, 2019Pacemaker Found in Back-Up Mode; Pacemaker Found in Back-Up Mode
IT WAS REPORTED THAT THE PATIENT'S DEVICE WENT INTO BACKUP VVI MODE. FURTHER INFORMATION WAS REQUESTED BUT NOT RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9267631 ↗MalfunctionNov 20, 2018Pacemaker Found in Back-Up Mode; Pacemaker Found in Back-Up Mode
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY ROOM WITH COMPLAINTS OF CHEST PAIN. UPON INTERROGATION, IT WAS DISCOVERED THE PACEMAKER WAS IN BACKUP VVI OPERATION. PROGRAMMING CHANGES COULD NOT BE MADE, AS THE BATTERY HAD REACHED ITS ELECTIVE REPLACEMENT INDICATOR NORMALLY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: Chest Pain
MAUDE report 8091119 ↗InjurySep 6, 2018Pacemaker Found in Back-Up Mode; Pacemaker Found in Back-Up Mode
DURING A CLINIC FOLLOW UP, IT WAS NOTED THAT THE DEVICE WENT INTO BACK UP MODE. THE DEVICE WAS EXPLANTED AND EXCHANGED TO RESOLVE THE EVENT AND THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7852122 ↗MalfunctionJul 30, 2018Intermittent Capture; Pacemaker Found in Back-Up Mode; Intermittent Capture; Pacemaker Found in Back-Up Mode
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A ROUTINE GENERATOR CHANGE PROCEDURE. DURING EXPOSURE TO ELECTROCAUTERY, THE PULSE GENERATOR REVERTED TO BACKUP OPERATION. INTERMITTENT LOSS OF CAPTURE WAS NOTED DUE TO THE CHANGE TO UNIPOLAR PACING. FURTHER INTERROGATION WAS NOT POSSIBLE DUE TO THE DEVICE'S BACKUP OPERATION. THE DEVICE WAS REMOVED AND REPLACED. THE PATIENT WAS STABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7733367 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P880086. One submission often covers many models, so check the product and lot details in each record.
- Z-1497-2022 · Jun 17, 2022 · Open, ClassifiedMerlin.net model MN5000 Software
- Z-1495-2022 · Jun 17, 2022 · Open, ClassifiedMerlin PCS 3650 programmer Model 3330 software
- Z-1496-2022 · Jun 17, 2022 · Open, ClassifiedMerlin 2 PCS MER37000 programmer Model MER3400 software
- Z-0430-2012 · Sep 23, 2011 · TerminatedAnthem CRT-P models PM3110 and PM3210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.
- Z-0429-2012 · Sep 23, 2011 · TerminatedImplantable Pacemaker Accent DR models PM2110 and PM2210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.
- Z-0298-2007 · Oct 6, 2006 · TerminatedSt. Jude Medical Identity SR Model 5172, Identity DR Model 5370, & Identity XL DR Model 5376; Pulse Generators (pacemakers)
- Z-0296-2007 · Oct 6, 2006 · TerminatedSt. Jude Medical APS III Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3500/3510
- Z-0297-2007 · Oct 6, 2006 · TerminatedSt. Jude Medical Merlin PCS Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3650
- Z-0052-04 · Jul 31, 2003 · TerminatedPulse Generator
Other Programmer, pacemaker devices
- Viva™ XT CRT-D DTBA1D1Medtronic, Inc. · Programmer, pacemaker1 recall25 deaths893 reports
- Medtronic CareLink® Programmer 2090WMedtronic, Inc. · Programmer, pacemaker3 recalls489 reports
- Medtronic CareLink® Programmer 2090WMedtronic, Inc. · Programmer, pacemaker3 recalls208 reports
- Viva™ Quad XT CRT-D DTBA1Q1Medtronic, Inc. · Programmer, pacemaker2 recalls6 deaths236 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Programmer, pacemaker2 recalls198 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Programmer, pacemaker3 recalls108 reports
- Viva™ Quad S CRT-D DTBB1QQMedtronic, Inc. · Programmer, pacemaker2 recalls1 death38 reports
- Viva™ Quad XT CRT-D DTBA1QQMedtronic, Inc. · Programmer, pacemaker1 recall47 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page