The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05414734054201), received between Sep 13, 2018 and Oct 29, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture1
- Gradient Increase1
- Obstruction of Flow1
Patient problems
- Corneal Pannus1
- Mitral Regurgitation1
- No Clinical Signs, Symptoms or Conditions1
- Thrombus1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 29, 2022Fracture; Fracture
IT WAS REPORTED THAT ON (B)(6) 2022, A 21MM SJM MASTERS SERIES HEMODYNAMIC PLUS VALVE WAS IMPLANTED INTO A PATIENT. AFTER THE VALVE WAS IMPLANTED, THE LEAFLET ON THE VALVE BROKE. THE VALVE WAS REMOVED AND REPLACED WITH 21MM EPIC SUPRA VALVE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15694781 ↗InjurySep 13, 2018Gradient Increase; Obstruction of Flow; Gradient Increase; Obstruction of Flow
A FORMER PREMATURE INFANT WAS OPERATED ON IN (B)(6) 2018 FOR CONGENITAL MITRAL INSUFFICIENCY. THE PATIENT HAD TWO FAILED ATTEMPTS TO REPAIR THE MITRAL VALVE WHICH WERE UNSUCCESSFUL DUE TO EXTREMELY FRAGILE AND RAPIDLY REORGANIZING ANNULAR TISSUE. ON (B)(6) 2018, A 21 MM MASTERS SERIES MECHANICAL HEART VALVE WAS PLACED IN THE MITRAL POSITION AND THE PATIENT RECOVERED WELL. ON (B)(6) 2018, THE PATIENT WAS READMITTED FOR RESPIRATORY DYSFUNCTION, INITIALLY TREATED AS ASPIRATION PNEUMONIA. UPON REVIEW OF THE ECHOCARDIOGRAM, IT WAS NOTED THAT THE MITRAL PROSTHESIS HAD THROMBOSED AND WAS NOT FUNCTIONING. THE VALVE WAS FIXED IN A SEMI-OPEN POSITION CAUSING EXTREMELY HIGH DIASTOLIC GRADIENTS AND A LARGE REGURGITATION. ON (B)(6) 2018, THE PATIENT'S CONDITION DETERIORATED AND THE PATIENT WAS PLACED ON V-A ECMO AND ATTEMPTS AT THROMBOLYSIS WERE PERFORMED WITHOUT IMPACT ON PRODUCT PERFORMANCE. ON (B)(6) 2018, THE DECISION TO SURGICALLY INTERVENE WAS MADE AND AS THE FACILITY DID NOT HAVE ANOTHER MECHANICAL HEART VALVE AVAILABLE, A 23 MM TRIFECTA GT VALVE WAS IMPLANTED. UPON EXPLANT OF THE THROMBOSED MHV, PANNUS AND THROMBUS WERE NOTED. THIS INTERVENTION WAS CONSIDERED A LIFESAVING PROCEDURE AND
Patient problems: Corneal Pannus; Mitral Regurgitation; Thrombus
MAUDE report 7872255 ↗
Other HEART-VALVE, MECHANICAL devices
- On-X Mitral Heart Valve with Conform-X Sewing Ring ONXMC-25/33On-X Life Technologies, Inc. · Heart-Valve, Mechanical1 recall1 death34 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths70 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL5 deaths66 reports
- SJM Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths44 reports
- CARBOMEDICS SUPRA-ANNULAR (TOP HAT) S5-021Corcym Srl · HEART-VALVE, MECHANICAL7 deaths37 reports
- Regent™ 21AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths35 reports
- Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL2 deaths33 reports
- Regent™ 19AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL31 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page