Livewire™ 401940Recalls & reported problems

St. Jude Medical Cardiovascular Division · Steerable Electrophysiology Catheter · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05414734202688), received between May 23, 2019 and Oct 27, 2023 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Signal Artifact/Noise” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Signal Artifact/Noise3
  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Cardiac Perforation1
  • Stroke/CVA1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 27, 2023Signal Artifact/Noise; Signal Artifact/Noise

    RELATED MANUFACTURING REF: (B)(4). DURING A PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA (PSVT) PROCEDURE, THE ENSITE DWS COULD NOT DISPLAY THE ELECTRODE IMAGE WHICH CAUSED A DELAY. THE DWS AND AMPLIFIER WERE RESTARTED AND RECONNECTED WITH NO RESOLUTION. THE PATCHES WERE THEN REPLACED WHICH RESOLVED THE ISSUE. ADDITIONALLY, ELECTRODES 7-8 OF THE CATHETER DID NOT DISPLAY AS EXPECTED IN THE RIGHT ATRIUM. THE CABLE AND THE POLE IN THE JUNCTION BOX WERE CHANGED WITH NO RESOLUTION. THE CATHETER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18022628 ↗
  • MalfunctionJun 21, 2023Signal Artifact/Noise; Signal Artifact/Noise

    RELATED MANUFACTURE REF: 2182269-2023-00028. DURING WOLFE PARKINSON WHITE (WPW) PROCEDURE, DISTORTION WAS NOTED ON ELECTRODE 2 OF THE CATHETER. A SECOND CATHETER FROM THE SAME LOT WAS USED AND THE SAME ISSUE WAS NOTED ON ELECTRODE 9. THE CABLES WERE REPLACED AND AMPERE INPUTS WERE CHANGED WITH NO RESOLUTION. A THIRD CATHETER WAS USED TO COMPLETE THE PROCEDURE WITH NO CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17171884 ↗
  • MalfunctionJun 21, 2023Signal Artifact/Noise; Signal Artifact/Noise

    RELATED MANUFACTURE REF: (B)(4). DURING WOLFE PARKINSON WHITE (WPW) PROCEDURE, DISTORTION WAS NOTED ON ELECTRODE 2 OF THE CATHETER. A SECOND CATHETER FROM THE SAME LOT WAS USED AND THE SAME ISSUE WAS NOTED ON ELECTRODE 9. THE CABLES WERE REPLACED AND AMPERE INPUTS WERE CHANGED WITH NO RESOLUTION. A THIRD CATHETER WAS USED TO COMPLETE THE PROCEDURE WITH NO CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17171866 ↗
  • InjuryMay 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REFERENCE NUMBER: 3005334138-2020-00200, 3008452825-2020-00251, 3005334138-2020-00201, 3008452825-2020-00252. TWO DAYS FOLLOWING AN ABLATION PROCEDURE, THE PATIENT DEVELOPED A CEREBROVASCULAR ACCIDENT. THE CIRCUMFERENTIAL FOUR PULMONARY VEIN ISOLATION AND CAVOTRICUSPID ISTHMUS ABLATION PROCEDURE WAS PERFORMED ON (B)(6) 2020 WITH NO REPORTED ISSUES. THE PATIENT WAS ANTICOAGULATED DURING THE PROCEDURE WITH AN ACTIVE CLOTTING TIME VALUE OF GREATER THAN 300, AND THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICE. ON (B)(6) 2020 A HEAD CT AND MRI WERE PERFORMED, REVEALING AN ACUTE-TO-SUBACUTE CEREBRAL INFARCTION OVER THE RIGHT FRONTAL AND TEMPORAL OPERCULA REGION. IT WAS SUSPECTED THAT THERE WAS A THROMBUS IN THE CORTICAL BIFURCATION OF THE INFERIOR SEGMENT OF THE RIGHT MIDDLE CEREBRAL ARTERY (MCA). THE PATIENT WAS HOSPITALIZED FOR IV FLUIDS AND ANTICOAGULANTS, REMAINING IN STABLE CONDITION. A REPEAT HEAD CT REVEALED NO INTRACRANIAL HEMORRHAGE AND THE PATIENT WAS DISCHARGED ON (B)(6) 2020 WITH NO RESIDUAL NEUROLOGICAL DEFICITS.

    Patient problems: Stroke/CVA

    MAUDE report 10088465 ↗
  • InjuryMay 23, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REFERENCE: 3005334138-2019-00256. DURING A SUPRAVENTRICULAR TACHYCARDIA ABLATION PROCEDURE A CARDIAC TAMPONADE OCCURRED. DURING THE PROCEDURE, FOLLOWING ABLATION OF THE POSTEROLATERAL LA, THE PATIENT BECAME HYPOTENSIVE AND A PERICARDIOCENTESIS WAS PERFORMED. TRANSOESOPHAGEAL ECHO CONFIRMED THE EFFUSION AND SURGICAL INTERVENTION WAS PERFORMED AND INCLUDED PLACING PATCHES AROUND THE PROXIMAL CORONARY SINUS OS TO STABILIZE THE PATIENT. THE PATIENT WAS STABILIZED. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICES.

    Patient problems: Cardiac Perforation

    MAUDE report 8635811 ↗

Other CATHETER, STEERABLE devices

All CATHETER, STEERABLE devices →