Agilis™ G408318Recalls & reported problems

St. Jude Medical, Inc. · NxT Steerable Introducer · Class 2

All models of Agilis (9)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Medical Agilis NxT steerable introducer is a tube used to guide catheters into the heart during rhythm procedures. Its 333 reports mostly describe air or blood leaking from its valve or side port, with the sheath usually replaced during the procedure. The reports mention 7 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
333
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 333 adverse event reports that list this exact device (DI 05414734206112), received between Jun 15, 2018 and Sep 24, 2026 — classified as 7 injuries, 326 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (152 reports).

1 other recall cover devices cleared under the same FDA submission (K061363, K241370, K243493); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 4 reportsFeb 2024: 3 reportsMar 2024: 7 reportsApr 2024: 8 reportsMay 2024: 5 reportsJun 2024: 4 reportsJul 2024: 4 reportsAug 2024: 5 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 3 reportsJan 2025: 6 reportsFeb 2025: 3 reportsMar 2025: 3 reportsApr 2025: 5 reportsMay 2025: 4 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 6 reportsSep 2025: 2 reportsOct 2025: 7 reportsNov 2025: 5 reportsDec 2025: 4 reportsJan 2026: 3 reportsFeb 2026: 7 reportsMar 2026: 3 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak152
  • Gas/Air Leak150
  • Material Separation14
  • Adverse Event Without Identified Device or Use Problem5
  • Detachment of Device or Device Component5
  • Peeled/Delaminated4
  • Fracture3
  • Difficult to Insert2

Patient problems

  • No Clinical Signs, Symptoms or Conditions257
  • No Consequences Or Impact To Patient71
  • Arrhythmia4
  • Foreign Body In Patient2
  • Air Embolism1
  • Cardiac Perforation1
  • Embolism/Embolus1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 333 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884214 ↗
  • MalfunctionJun 2, 2026Gas/Air Leak

    DURING A PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, AND AIR LEAK WAS NOTED DURING FLUSHING OF THE SIDEPORT OF THE AGILIS 8.5 FR SHEATH. THE TACTIFLEX SE CATHETER WAS INTRODUCED INTO THE AGILIS 8.5 FR SHEATH WHEN THE AIR LEAK WAS NOTED. THE AGILIS 8.5 FR SHEATH WAS REPLACED; THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT HEALTH CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25359917 ↗
  • MalfunctionJun 1, 2026Gas/Air Leak

    DURING AN ATRIAL FIBRILLATION PROCEDURE, AN AIR LEAK WAS NOTED ON THE AGILIS 8.5 FR SHEATH WHEN EXCHANGING THE CATHETER. THE AGILIS 8.5 FR CATHETER WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT HEALTH CONSEQUENCES. THE AGILIS 8.5 FR SHEATH WAS DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25349002 ↗
  • MalfunctionApr 30, 2026Gas/Air Leak

    DURING AN ATRIAL TACHYCARDIA PROCEDURE, AIR ASPIRATED FROM THE SHEATH. THE SHEATH WAS INSERTED INTO THE LEFT ATRIUM. NEAR THE END OF THE PROCEDURE, WHEN ATTEMPTING TO PERFORM AN AIR FLUSH OF THE SHEATH, AIR WAS ASPIRATED. ASPIRATION WAS PERFORMED SEVERAL TIMES FROM THE SIDE PORT; HOWEVER, THE AIR COULD NOT BE FULLY REMOVED. THE SHEATH WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25039926 ↗
  • MalfunctionApr 14, 2026Fluid/Blood Leak

    DURING THE PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, BLOOD LEAKED FROM THE SHEATH VALVE WHEN OPERATING THE ABLATION CATHETER THROUGH THE AGILIS SHEATH AND WHEN THE CATHETER WAS WITHDRAWN. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24872346 ↗
  • MalfunctionMar 30, 2026Fluid/Blood Leak

    DURING ATRIAL FIBRILLATION PROCEDURE, BLOOD LEAKED FROM THE HEMOSTASIS VALVE. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24715899 ↗
  • MalfunctionMar 20, 2026Fluid/Blood Leak

    DURING THE ATRIAL FIBRILLATION ABLATION PROCEDURE, BLOOD LEAKAGE WAS NOTED FROM THE HEMOSTASIS VALVE. AFTER THE CATHETER WAS INSERTED INTO THE SHEATH INTO THE LEFT ATRIUM, BLOOD LEAKAGE WAS NOTED FROM THE HEMOSTASIS VALVE. THE SHEATH WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649778 ↗
  • MalfunctionMar 6, 2026Gas/Air Leak

    DURING A PAROXYSMAL ATRIAL FIBRILLATION PROCEDURE, AIR WAS ASPIRATED FROM THE SHEATH. THE SHEATH WAS INSERTED INTO THE LEFT ATRIUM AND THE ABLATION CATHETER WAS INTRODUCED. DURING RF DELIVERY, AIR EVACUATION WAS PERFORMED VIA THE SIDE PORT, AND AIR WAS ASPIRATED. MULTIPLE ATTEMPTS WERE MADE TO ASPIRATE THE AIR WITH NO RESOLUTION. THE SHEATH WAS REPLACED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24531188 ↗
  • MalfunctionFeb 26, 2026Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, AIR WAS ASPIRATED FROM THE SHEATH. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24462218 ↗
  • MalfunctionFeb 24, 2026Gas/Air Leak

    DURING AN ATRIAL FIBRILLATION PROCEDURE, AFTER INSERTING THE SHEATH INTO THE PATIENT AND DURING THE PREPARATION PHASE BEFORE INTRODUCING THE CATHETER, AIR WAS ASPIRATED. THE SHEATH WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24430067 ↗
  • MalfunctionFeb 19, 2026Fluid/Blood Leak

    DURING THE SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, THE VALVE LEAKED BLOOD AFTER INSERTING THE INTRODUCER INTO THE PATIENT. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24390211 ↗
  • MalfunctionFeb 19, 2026Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, THE SHEATH WAS INSERTED INTO THE LEFT ATRIUM, ASPIRATION WAS PERFORMED FOR FLUSHING PRIOR TO CATHETER INSERTION, AND AIR WAS ASPIRATED. SEVERAL ATTEMPTS WERE MADE TO ASPIRATE THE AIR, BUT THE AIR COULD NOT BE FULLY REMOVED. THE SHEATH WAS REPLACED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24389492 ↗
  • MalfunctionFeb 19, 2026Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, AFTER THE AGILIS INTRODUCER SHEATH WAS INSERTED INTO THE PATIENT, AIR WAS ASPIRATED BEFORE CATHETER INSERTION. THE AGILIST INTRODUCER SHEATH WAS REPLACED AND THE PROCEDURE COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24389477 ↗
  • MalfunctionFeb 17, 2026Gas/Air Leak

    DURING THE DIAGNOSTIC PROCEDURE, AIR WAS ASPIRATED FROM THE VALVE. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24368824 ↗
  • MalfunctionFeb 2, 2026Gas/Air Leak

    DURING AN ATRIAL FIBRILLATION PROCEDURE, AN AIR LEAK OCCURRED. THE SHEATH WAS INSERTED INTO THE LEFT ATRIUM. MAPPING WAS PERFORMED USING THE NON-ABBOTT CATHETER (ORION BY BOSTON SCIENTIFIC). DURING RF DELIVERY WITH THE NON-ABBOTT CATHETER (STABLEPOINT BY BOSTON SCIENTIFIC), AIR WAS ASPIRATED FROM THE SIDE PORT OF THE SHEATH. DESPITE MULTIPLE ATTEMPTS TO ASPIRATE THE AIR, IT COULD NOT BE COMPLETELY REMOVED. THE SHEATH WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24238066 ↗
  • MalfunctionJan 27, 2026Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, AIR ENTERED THE HEMOSTASIS VALVE. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24192232 ↗
  • MalfunctionJan 15, 2026Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, AFTER INSERTING AND WITHDRAWING MAPPING CATHETERS THROUGH THE INTRODUCER, INTERMITTENT AIR ASPIRATION WAS OBSERVED. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. THE HOSPITAL DISCARDED THE REPORTED INTRODUCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24080777 ↗
  • MalfunctionJan 13, 2026Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, AIR WAS ASPIRATED. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. THE HOSPITAL DISCARDED THE REPORTED INTRODUCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24053430 ↗
  • MalfunctionDec 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, THE ABLATION CATHETER WAS INSERTED INTO THE AGILIS SHEATH AND THEN MANIPULATED, AND THE PHYSICIAN NOTICED THAT BLOOD LEAKED FROM THE VALVE OF THE AGILIS SHEATH WHEN REMOVING IT FROM THE PATIENT¿S BODY. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23863393 ↗
  • MalfunctionDec 16, 2025Material Separation; Material Separation

    THIS REPORT IS TO ADVISE OF A SIDEPORT TUBING DETACHMENT THAT WAS NOTED DURING ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23822690 ↗
  • MalfunctionDec 4, 2025Gas/Air Leak; Gas/Air Leak

    DURING THE PREMATURE VENTRICULAR CONTRACTIONS PROCEDURE, AIR LEAKED FROM THE SHEATH. THE SHEATH WAS INSERTED INTO THE HEART AND THE DILATOR WAS REMOVED. WHEN NEGATIVE PRESSURE WAS APPLIED FOR FLUSHING, AIR WAS ASPIRATED INTO THE SYRINGE. AIR WAS ASPIRATED SEVERAL TIMES, BUT IT COULD NOT BE COMPLETELY REMOVED. THE SHEATH WAS REPLACED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23720629 ↗
  • MalfunctionDec 1, 2025Gas/Air Leak; Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, AIR WAS ASPIRATED FROM THE INTRODUCER. THE INTRODUCER WAS REPLACED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23684989 ↗
  • MalfunctionNov 25, 2025Gas/Air Leak; Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, THERE WAS AN AIR LEAK FROM THE SHEATH. THE SHEATH WAS INSERTED INTO THE RIGHT ATRIUM, AND WHEN FLUSHING BEFORE INSERTING THE CATHETER OR SEPTAL PUNCTURE NEEDLE, AIR WAS ASPIRATED. AIR WAS ASPIRATED MULTIPLE TIMES, BUT THE AIR COULD NOT BE COMPLETELY REMOVED. THE SHEATH WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645378 ↗
  • MalfunctionNov 20, 2025Fluid/Blood Leak; Fluid/Blood Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, THE HEMOSTASIS VALVE LEAKED BLOOD. THE SHEATH WAS INSERTED INTO THE RIGHT ATRIUM AND WHEN THE DILATOR WAS REMOVED PRIOR TO THE INSERTION OF THE CATHETER AND SEPTAL PUNCTURE NEEDLE, BLOOD LEAKED FROM THE VALVE. THE SHEATH WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23605552 ↗
  • MalfunctionNov 5, 2025Gas/Air Leak; Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE, IT WAS NOTED THAT AIR WAS ASPIRATED DURING ASPIRATION WHEN THE CATHETERS WERE EXCHANGED. EVEN AFTER ASPIRATING AIR SEVERAL TIMES, AIR COULD NOT BE REMOVED COMPLETELY. THE SHEATH WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23480516 ↗

FDA clearance and classification

Cleared through Special 510(k) K243493 on Dec 13, 2024 (decision: Substantially Equivalent). Submitted by Abbott Medical as “Agilis NxT Steerable Introducer Dual-Reach”. The FDA review took 31 days. FDA record ↗

Cleared through Special 510(k) K241370 on Jun 13, 2024 (decision: Substantially Equivalent). Submitted by Abbott Medical as “Agilis™ NxT Steerable Introducer Dual-Reach™ (I-V2-MED)”. The FDA review took 30 days. FDA record ↗

Cleared through Special 510(k) K061363 on Jul 12, 2006 (decision: Substantially Equivalent). Submitted by St Jude Medical as “Agilis Nxt Steerable Introducer”. The FDA review took 57 days. FDA record ↗

The FDA regulates product code DYB (introducer, catheter) under 21 CFR 870.1340 as a Class II device. 21 CFR 870.1340 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K061363, K241370, K243493. One submission often covers many models, so check the product and lot details in each record.

  • Z-2348-2024 · Apr 30, 2024 · Open, Classified
    St. Jude Medical Agilis NxT Steerable Introducer, REF 408309

Common questions about the Agilis™ G408318

Has the Agilis™ G408318 been recalled?

No FDA recall record in our data names this exact device identifier (DI 05414734206112), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the Agilis™ G408318?

The FDA's MAUDE database lists 333 adverse event reports for this device, received between Jun 15, 2018 and Sep 24, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (152), “Gas/Air Leak” (150) and “Material Separation” (14); the most common patient problems are “Arrhythmia” (4), “Foreign Body In Patient” (2) and “Air Embolism” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Agilis™ G408318?

None of the reports are classified as deaths, 7 as injuries and 326 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Agilis™ G408318 FDA approved?

It was cleared through Special 510(k) K061363 on Jul 12, 2006; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Agilis™ G408318, and is it still sold?

The FDA's device database lists St. Jude Medical, Inc. as the company responsible for it (brand Agilis™, model G408318). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.