VIZIGO D138501Recalls & reported problems

Biosense Webster Inc · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Biosense Webster Vizigo sheath is a catheter introducer used to guide devices into the heart. Its 2,014 adverse event reports mostly describe material separation, blood leaks, and backflow, and some mention cardiac tamponade, air in the bloodstream or bleeding. The reports include 2 deaths and 90 injuries. There are 2 open recalls, from 2023 for carton labels that do not match the inner pouch and from 2022 for water-damaged packaging that could affect sterility.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,005
Deaths reported
2
Injuries reported
90
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2,005 adverse event reports that list this exact device (DI 10846835016253), received between Jul 27, 2018 and Sep 30, 2026 — classified as 2 deaths, 90 injuries, 1,913 malfunctions.

The most frequently reported specific device problem is “Material Separation” (794 reports).

2 FDA recall records name this device's identifier, 2 still open.

5 other recalls cover devices cleared under the same FDA submission (K170997); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 33 reportsDec 2023: 17 reportsJan 2024: 18 reportsFeb 2024: 26 reportsMar 2024: 33 reportsApr 2024: 28 reportsMay 2024: 19 reportsJun 2024: 21 reportsJul 2024: 25 reportsAug 2024: 24 reportsSep 2024: 20 reportsOct 2024: 14 reportsNov 2024: 19 reportsDec 2024: 15 reportsJan 2025: 21 reportsFeb 2025: 15 reportsMar 2025: 18 reportsApr 2025: 9 reportsMay 2025: 18 reportsJun 2025: 16 reportsJul 2025: 13 reportsAug 2025: 19 reportsSep 2025: 17 reportsOct 2025: 11 reportsNov 2025: 16 reportsDec 2025: 29 reportsJan 2026: 23 reportsFeb 2026: 26 reportsMar 2026: 26 reportsApr 2026: 23 reportsMay 2026: 22 reportsJun 2026: 13 reportsJul 2026: 7 reportsAug 2026: 10 reportsSep 2026: 17 reportsOct 2026: 0 reports20242025202633
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 21, 2023Open, Classified

    Incorrect labeling; specifically, the product name, lot number and product code on the carton label for Bi-Directional Guiding Sheaths does not match the inner label pouch

    Z-1319-2023 · Root cause: Under Investigation by firm · FDA record ↗

  2. Jun 29, 2022Open, Classified

    Outside packaging for a device indicated for introducing cardiovascular catheters into the heart, sustained water damage, which may have affected the sterile barrier of the product inside. Devices with a compromised sterile barrier could be a source for infection.

    Z-1677-2022 · Root cause: Material/Component Contamination · FDA record ↗

Most reported problems

Device problems

  • Material Separation794
  • Fluid/Blood Leak420
  • Backflow240
  • Detachment of Device or Device Component201
  • Device-Device Incompatibility181
  • Obstruction of Flow105
  • Break94
  • Patient Device Interaction Problem84

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,773
  • No Consequences Or Impact To Patient135
  • Cardiac Tamponade37
  • Air Embolism16
  • Blood Loss11
  • Hemorrhage/Blood Loss/Bleeding10
  • Ischemic Heart Disease8
  • Cardiac Perforation7

Recent adverse event reports

The 25 most recent of 2,005 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954730 ↗
  • MalfunctionSep 30, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953603 ↗
  • MalfunctionSep 30, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951315 ↗
  • MalfunctionSep 30, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950871 ↗
  • MalfunctionSep 30, 2026Fluid/Blood Leak

    It was reported that a patient underwent a supraventricular tachycardia (SVT) ablation procedure with a VIZIGO and a leakage issue occurred.It was reported that during the operation, fluid (saline solution) escaped from an unintended location of the tubing; the hub area according to a picture that was received. A second device was used to complete the operation. There was no adverse event reported on patient.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944784 ↗
  • MalfunctionSep 30, 2026Material Separation

    It was reported that a patient underwent an Atrial Fibrillation (AFib) ablation procedure with a VIZIGO and the hemostatic valve was dislodged inside the hub. It was reported that hemostatic valve of Vizigo was damaged. The while gasket was found abnormal. Change another sheath to complete the operation. There was no adverse event reported on patient. Additional information received indicating the hemostatic valve was dislodged inside the hub and not dislodged outside of the hub. The brim cap/hub did not detach from the sheath. No air entered the patient¿s body. There was no blood loss.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941966 ↗
  • MalfunctionSep 29, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935472 ↗
  • MalfunctionSep 28, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920599 ↗
  • MalfunctionSep 28, 2026Difficult to Advance

    It was reported that a patient underwent an Atrial Fibrillation (AFib) ablation procedure with a VIZIGO and while trying to cross the septum after going transseptal, the VIZIGO sheath would not cross. The Regional Technology Specialist (RTS) reported that the physician tried different angles without resolution. The RTS reported that they pulled the sheath from the patient, inspected it, but found no deformities. They replaced the VIZIGO sheath and the issue was resolved. The procedure continued. No adverse patient consequence was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916931 ↗
  • MalfunctionSep 25, 2026Backflow; Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903386 ↗
  • MalfunctionSep 25, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902466 ↗
  • MalfunctionSep 24, 2026Backflow; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889835 ↗
  • MalfunctionSep 17, 2026Break

    It was reported that a patient underwent a pulmonary vein isolation ablation procedure with a Vizigo sheath and the physician noticed the tip was frayed. The doctor noticed the VIZIGO¿ sheath was not advancing to the location he wanted it to with ease. He exited the body and noticed the sheath was frayed at the tip. The sheath was replaced and the issue was resolved. The procedure continued and was completed. No patient consequences were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50806192 ↗
  • MalfunctionSep 9, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50697424 ↗
  • MalfunctionSep 4, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50600190 ↗
  • MalfunctionSep 3, 2026Material Protrusion/Extrusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50597231 ↗
  • MalfunctionSep 3, 2026Fluid/Blood Leak; Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50596888 ↗
  • MalfunctionAug 28, 2026Mechanical Problem; Contamination /Decontamination Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50492452 ↗
  • MalfunctionAug 26, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50458839 ↗
  • MalfunctionAug 20, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50356380 ↗
  • MalfunctionAug 18, 2026Material Too Soft/Flexible

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50311992 ↗
  • MalfunctionAug 18, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50290947 ↗
  • MalfunctionAug 7, 2026Device-Device Incompatibility

    It was reported that a patient underwent a cardiac ablation procedure with a VIZIGO and catheter for which Biosense Webster¿s Product Analysis Lab (PAL) identified the hemostatic valve dislodged inside the hub of the device. It was initially reported by the customer that the dilatator could not go through the valve of the VIZIGO Sheath. Issue happened before the operation. The sheath was not used on the patient, therefore, no patient hazard/consequence and the procedure could be finished normally.There was no damage to the sheath or dilator noticed and no occlusion was noticed when irrigating the sheath. No observable blockage. The customer's reported impeded sheath issue is not considered to be MDR Reportable since the potential risk that it could cause or contribute to a serious injury or death to the operator or patient is remote. On 13-Jul-2026, the BWI PAL revealed that a visual inspection of the returned device found the hemostatic valve dislodged inside the hub of the device. This findings was reviewed and assessed as an MDR Reportable Malfunction since the integrity of the device has been compromised.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50182985 ↗
  • MalfunctionAug 7, 2026Fluid/Blood Leak

    It was reported that a patient underwent an Atrial Fibrillation (AFib) ablation procedure with a VIZIGO and a fluid leakage issue occurred. It was reported that during the operation, fluid (saline solution) escaped from an unintended location of the sheath. A second device was used to complete the operation. There was no adverse event reported on patient.The issue was suspected to be in the hemostatic valve area. The brim cap did not detach and the hub did not detach from the sheath. No air entered the patient.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50181006 ↗
  • MalfunctionAug 5, 2026Expiration Date Error

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50150205 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170997. One submission often covers many models, so check the product and lot details in each record.

  • Z-1290-2024 · Feb 16, 2024 · Open, Classified
    CARTO VIZIGO Bi-Directional Guiding Sheath, REF D128502 The CARTO VIZIGOTM Bi-Directional Guiding Sheath is designed to provide accessibility and maneuverability in the cardiac a
  • Z-1320-2023 · Feb 21, 2023 · Open, Classified
    Biosense Webster Carto Vizigo Bi-Directional Guiding Sheath, 8.5F(Medium), REF D-1385-02-S, RX CE 2797, Sterile
  • Z-1287-2021 · Feb 26, 2021 · Completed
    Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (medium) C3, REF: D138502, Rx Only, CE 2797, STERILE EO, UDI: (01)10846835016277 - Product Usage: provide access
  • Z-1286-2021 · Feb 26, 2021 · Completed
    Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (Large) C3, REF: D138503, Rx Only, CE 2797, STERILE EO, udi: (01)10846835016260 - Product Usage: provide accessi
  • Z-1285-2021 · Feb 26, 2021 · Completed
    Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (SMALL) C3, REF: D138501, Rx Only, CE 2797, STERILE EO, udi: (01)10846835016253 - Product Usage: provide accessi