VIZIGO D138501Recalls & reported problems

Biosense Webster Inc · Class 2

All models of VIZIGO (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Biosense Webster Vizigo sheath is a catheter introducer used to guide devices into the heart. Its 2,014 adverse event reports mostly describe material separation, blood leaks, and backflow, and some mention cardiac tamponade, air in the bloodstream or bleeding. The reports include 2 deaths and 90 injuries. There are 2 open recalls, from 2023 for carton labels that do not match the inner pouch and from 2022 for water-damaged packaging that could affect sterility.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,012
Deaths reported
2
Injuries reported
90
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2,012 adverse event reports that list this exact device (DI 10846835016253), received between Jul 27, 2018 and Sep 30, 2026 — classified as 2 deaths, 90 injuries, 1,920 malfunctions.

The most frequently reported specific device problem is “Material Separation” (795 reports).

2 FDA recall records name this device's identifier, 2 still open.

5 other recalls cover devices cleared under the same FDA submission (K170997); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 17 reportsJan 2024: 18 reportsFeb 2024: 26 reportsMar 2024: 33 reportsApr 2024: 28 reportsMay 2024: 19 reportsJun 2024: 21 reportsJul 2024: 25 reportsAug 2024: 24 reportsSep 2024: 20 reportsOct 2024: 14 reportsNov 2024: 19 reportsDec 2024: 15 reportsJan 2025: 21 reportsFeb 2025: 15 reportsMar 2025: 18 reportsApr 2025: 9 reportsMay 2025: 18 reportsJun 2025: 16 reportsJul 2025: 13 reportsAug 2025: 19 reportsSep 2025: 17 reportsOct 2025: 11 reportsNov 2025: 16 reportsDec 2025: 29 reportsJan 2026: 23 reportsFeb 2026: 26 reportsMar 2026: 26 reportsApr 2026: 23 reportsMay 2026: 22 reportsJun 2026: 13 reportsJul 2026: 7 reportsAug 2026: 14 reportsSep 2026: 20 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202633
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 21, 2023Class IIOpen, Classified

    Incorrect labeling; specifically, the product name, lot number and product code on the carton label for Bi-Directional Guiding Sheaths does not match the inner label pouch

    Z-1319-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jun 29, 2022Class IIOpen, Classified

    Outside packaging for a device indicated for introducing cardiovascular catheters into the heart, sustained water damage, which may have affected the sterile barrier of the product inside. Devices with a compromised sterile barrier could be a source for infection.

    Z-1677-2022 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Separation795
  • Fluid/Blood Leak425
  • Backflow241
  • Detachment of Device or Device Component201
  • Device-Device Incompatibility181
  • Obstruction of Flow105
  • Break93
  • Patient Device Interaction Problem84

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,780
  • No Consequences Or Impact To Patient135
  • Cardiac Tamponade37
  • Air Embolism16
  • Blood Loss11
  • Hemorrhage/Blood Loss/Bleeding11
  • Ischemic Heart Disease8
  • Cardiac Perforation7

Recent adverse event reports

The 25 most recent of 2,012 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954730 ↗
  • MalfunctionSep 30, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953603 ↗
  • MalfunctionSep 30, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951315 ↗
  • MalfunctionSep 30, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950871 ↗
  • MalfunctionSep 30, 2026Fluid/Blood Leak

    It was reported that a patient underwent a supraventricular tachycardia (SVT) ablation procedure with a VIZIGO and a leakage issue occurred.It was reported that during the operation, fluid (saline solution) escaped from an unintended location of the tubing; the hub area according to a picture that was received. A second device was used to complete the operation. There was no adverse event reported on patient.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944784 ↗
  • MalfunctionSep 29, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935472 ↗
  • MalfunctionSep 28, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920599 ↗
  • MalfunctionSep 25, 2026Backflow; Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903386 ↗
  • MalfunctionSep 25, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902466 ↗
  • MalfunctionSep 24, 2026Backflow; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889835 ↗
  • MalfunctionSep 23, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50862929 ↗
  • MalfunctionSep 15, 2026Fluid/Blood Leak

    It was reported that a patient underwent a cardiac ablation procedure with a VIZIGO sheath and a hemostatic valve leak issue was observed.\\Leaking valve with air entrainment during aspiration and persistent back bleed. Replaced with new Vizigo sheath. No patient or user harm reported. No significant delay.\\Additional information was received. The specific section where the issue was observed was likely hemostatic valve related. This part had a leakage issue. The hemostatic valve did not dislodge. The brim cap did not become detached from the sheath. The hub did not become detached from the sheath. The sheath was being used on the patient. Impossible to know for sure, but unlikely that air entered the patient¿s body as aspiration taking place when fault noted. No patient consequence reported. It did not require treatment. The approximate volume of blood that was lost was a few ml. Not used with needle.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50766168 ↗
  • MalfunctionSep 14, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50765230 ↗
  • MalfunctionSep 12, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50745262 ↗
  • MalfunctionSep 12, 2026Difficult to Advance; Material Too Soft/Flexible

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50744674 ↗
  • MalfunctionSep 11, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50741702 ↗
  • MalfunctionSep 11, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50737554 ↗
  • MalfunctionSep 9, 2026Coagulation in Device or Device Ingredient; Material Protrusion/Extrusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50698538 ↗
  • MalfunctionSep 3, 2026Backflow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50599049 ↗
  • MalfunctionSep 3, 2026Fluid/Blood Leak; Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50596888 ↗
  • MalfunctionAug 31, 2026Difficult to Advance; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50533790 ↗
  • MalfunctionAug 26, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50458839 ↗
  • MalfunctionAug 25, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50439366 ↗
  • MalfunctionAug 20, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50356380 ↗
  • MalfunctionAug 18, 2026Material Too Soft/Flexible

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50311992 ↗

FDA clearance and classification

Cleared through 510(k) K170997 on Jul 25, 2017 (decision: Substantially Equivalent). Submitted by Biosense Webster, Inc. as “CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath”. The FDA review took 113 days. FDA record ↗

The FDA regulates product code DYB (introducer, catheter) under 21 CFR 870.1340 as a Class II device. 21 CFR 870.1340 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K170997. One submission often covers many models, so check the product and lot details in each record.

  • Z-1290-2024 · Feb 16, 2024 · Open, Classified
    CARTO VIZIGO Bi-Directional Guiding Sheath, REF D128502 The CARTO VIZIGOTM Bi-Directional Guiding Sheath is designed to provide accessibility and maneuverability in the cardiac a
  • Z-1320-2023 · Feb 21, 2023 · Open, Classified
    Biosense Webster Carto Vizigo Bi-Directional Guiding Sheath, 8.5F(Medium), REF D-1385-02-S, RX CE 2797, Sterile
  • Z-1287-2021 · Feb 26, 2021 · Completed
    Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (medium) C3, REF: D138502, Rx Only, CE 2797, STERILE EO, UDI: (01)10846835016277 - Product Usage: provide access
  • Z-1286-2021 · Feb 26, 2021 · Completed
    Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (Large) C3, REF: D138503, Rx Only, CE 2797, STERILE EO, udi: (01)10846835016260 - Product Usage: provide accessi
  • Z-1285-2021 · Feb 26, 2021 · Completed
    Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (SMALL) C3, REF: D138501, Rx Only, CE 2797, STERILE EO, udi: (01)10846835016253 - Product Usage: provide accessi

Common questions about the VIZIGO D138501

Has the VIZIGO D138501 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1319-2023 (Class II), began on Feb 21, 2023: Incorrect labeling; specifically, the product name, lot number and product code on the carton label for Bi-Directional Guiding Sheaths does not match the inner label pouch. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the VIZIGO D138501?

The FDA's MAUDE database lists 2,012 adverse event reports for this device, received between Jul 27, 2018 and Sep 30, 2026. In those reports, the most common specific device problems are “Material Separation” (795), “Fluid/Blood Leak” (425) and “Backflow” (241); the most common patient problems are “Cardiac Tamponade” (37), “Air Embolism” (16) and “Blood Loss” (11). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VIZIGO D138501?

2 reports are classified as deaths, 90 as injuries and 1,920 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the VIZIGO D138501 FDA approved?

It was cleared through 510(k) K170997 on Jul 25, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the VIZIGO D138501, and is it still sold?

The FDA's device database lists Biosense Webster Inc as the company responsible for it (brand VIZIGO, model D138501). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page