Quattrode® 3163Recalls & reported problems

St. Jude Medical, Inc. · Lead Wide Spaced, 30 cm · Class 2 · Implant

Quick summary

This St. Jude Medical Quattrode spinal cord stimulator lead is an implanted pain-relief device. Reports mostly describe ineffective therapy and lead migration, sometimes requiring surgery. There are 81 reports: no deaths and 81 injuries; no recalls name this device. Reports fell to 0 in the last 12 months from 3.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
81
Deaths reported
0
Injuries reported
81
Recalls naming it
0

Summary

The FDA's MAUDE database contains 81 adverse event reports that list this exact device (DI 05414734401654), received between Oct 18, 2018 and Sep 5, 2025 — classified as 81 injuries.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (40 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Therapeutic or Diagnostic Output Failure40
  • Migration11
  • Adverse Event Without Identified Device or Use Problem9
  • Insufficient Device Problem Information8
  • High impedance6
  • Inappropriate/Inadequate Shock/Stimulation3
  • Fracture2
  • Impedance Problem2

Patient problems

  • Inadequate Pain Relief64
  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient4
  • Insufficient Information2
  • Scar Tissue2
  • Undesired Nerve Stimulation2
  • Device Overstimulation of Tissue1
  • Erosion1

Recent adverse event reports

The 25 most recent of 81 reports, in the reporter's own words as published by the FDA.

  • InjurySep 5, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS SUCH, SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 22984570 ↗
  • InjuryMay 13, 2025High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO HIGH IMPEDANCES ON BOTH LEADS. SURGICAL INTERVENTION MAY TAKE PLACE IN THE FUTURE TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 22012764 ↗
  • InjuryJan 2, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THE PATIENT IS NOT RECEIVING EFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2024 WHEREIN THE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH LEAD CAUSED THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 21061753 ↗
  • InjurySep 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE LEAD WAS SHOWING THROUGH THE SKIN ABOUT 1MM AND CAUSING PAIN/DISCOMFORT. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED. THERAPY WAS RESTORED POST-OP.

    Patient problems: Erosion; Implant Pain

    MAUDE report 20297421 ↗
  • InjuryFeb 7, 2024Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED UNCOMFORTABLE STIMULATION DUE TO LEAD MIGRATION. SURGICAL INTERVENTION MAY BE UNDERTAKEN AT A LATER DATE TO ADDRESS THE ISSUE. INVESTIGATION WAS UNABLE TO DETERMINE WHICH OF THE LEADS MIGRATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Undesired Nerve Stimulation

    MAUDE report 18659516 ↗
  • InjuryNov 2, 2023Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE STIMULATION. SURGICAL INTERVENTION TOOK PLACE WHERE THE SYSTEM WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THE IPG WAS ELECTIVELY REPLACED. EFFECTIVE STIMULATION WAS RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 18062320 ↗
  • InjurySep 6, 2023Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2023-04229. IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION WITH THEIR PERIPHERAL SCS SYSTEM. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN AT A LATER DATE TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH LEAD IS LIABLE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17689326 ↗
  • InjuryJun 30, 2023Migration; Migration

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2023-03103, 1627487-2023-03104. IT WAS REPORTED THE PATIENT EXPERIENCED LEAD MIGRATION AND AN INOPERABLE IPG DUE TO NOT RECHARGING THE IPG AS RECOMMENDED. AS SUCH, THE IPG BECAME INOPERABLE. SURGICAL INTERVENTION TOOK PLACE WHEREIN THE IPG AND LEADS WERE EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17239806 ↗
  • InjuryJun 21, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2023-02839, 1627487-2023-02954, 1627487-2023-02955. IT WAS REPORTED THAT THE LEAD WAS EXPLANTED ON AN UNKNOWN DATE FOR AN UNKNOWN REASON. NO OTHER INFORMATION IS KNOWN AT THIS TIME.

    Patient problems: Insufficient Information

    MAUDE report 17174961 ↗
  • InjuryMay 31, 2023High impedance; High impedance

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2023-02570. IT WAS REPORTED THAT BOTH OF THE PATIENT'S LEADS HAD HIGH IMPEDANCES AS A RESULT THE PATIENT LOST EFFECTIVE THERAPY. SURGICAL INTERVENTION IS PLANNED TO ADDRESS THIS ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17035657 ↗
  • InjuryMay 23, 2023Therapeutic or Diagnostic Output Failure; Migration; Therapeutic or Diagnostic Output Failure; Migration

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2023-02838. ADDITIONAL INFORMATION INDICATES THAT THE LEADS HAD MIGRATED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2023 WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16985770 ↗
  • InjuryApr 5, 2023Migration; Migration

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2023-01518. IT WAS REPORTED THAT THE PATIENT EXPERIENCED A MINOR FALL AND AS A RESULT THE PATIENT'S LEADS HAD MIGRATED AND WAS EXPERIENCING INEFFECTIVE STIMULATION. SURGICAL INTERVENTION MAY TAKE PLACE AT A FUTURE DATE TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16686280 ↗
  • InjuryJan 11, 2023High impedance; High impedance

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2023-00125. IT WAS REPORTED THE PATIENT'S IPG WAS INOPERABLE. SURGICAL INTERVENTION WAS UNDERTAKEN TO REPLACE THE IPG. POST OPERATIVELY IT WAS REPORTED ONE OF THE PATIENT'S LEADS HAD ALL HIGH IMPEDANCES. EFFECTIVE THERAPY WAS ESTABLISHED POST-OPERATIVELY. PATIENT IS TO FOLLOW-UP WITH PHYSICIAN AS THE NEXT COURSE OF ACTION.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16141806 ↗
  • InjuryJan 8, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A LEAD REVISION PROCEDURE A PART OF THE LEAD HAD FRACTURED AND A FRAGMENT OF THE LEAD WITH THE DISTAL CONTACTS REMAINS IMPLANTED IN THE SOFT TISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16119442 ↗
  • InjuryJul 1, 2022Migration; Migration

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2022-03697. IT WAS REPORTED THE PATIENT'S LEADS MIGRATED. THE ISSUE WAS CONFIRMED VIA X-RAY. AS A RESULT, SURGICAL INTERVENTION OCCURRED WHEREIN THE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 14883303 ↗
  • InjuryJun 15, 2022High impedance; Battery Problem: High Impedance; High impedance; Battery Problem: High Impedance

    RELATED MANUFACTURER NUMBER: 1627487-2022-03409. IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION DUE TO HIGH IMPEDANCE. SURGERY MAY TAKE PLACE LATER. IT IS UNKNOWN WHICH LEAD PRESENTS THE ISSUE, THEREFORE BOTH LEADS ARE BEING REPORTED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 14709053 ↗
  • InjuryMay 15, 2022Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2022-02777, 1627487-2022-02779, AND 1627487-2022-02780. IT WAS REPORTED THAT THE PATIENT HAD A FALL IN MARCH 2022 AND LOST EFFECTIVE THERAPY. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION DURING WHICH THE LEADS WERE EXPLANTED AND REPLACED, RESOLVING THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 14404070 ↗
  • InjuryMay 5, 2022Impedance Problem; Impedance Problem

    RELATED MANUFACTURER NUMBER: 1627487-2022-02325, AND 1627487-2022-02327. IT WAS REPORTED THE PATIENT EXPERIENCED AUTOREDUCING AND LATER LOST THERAPY AS THE IPG REACHED END OF SERVICE (EOS). IT IS UNKNOWN IF THE IPG DEPLETED PREMATURELY. AS A RESULT, SURGICAL INTERVENTION MAY OCCUR LATER TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 14292915 ↗
  • InjuryMar 23, 2022Migration; Migration

    RELATED MANUFACTURER REFERENCE NUMBER:1627487-2022-01541, 3006705815-2022-01441. IT WAS REPORTED ONE OF THE PATIENT'S LEADS HAD MIGRATED AND THE IPG HAS FLIPPED IN THE POCKET. SURGICAL INTERVENTION MAY OCCUR LATER TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 13858872 ↗
  • InjuryMar 4, 2022Fracture; Fracture

    RELATED MANUFACTURER REFERENCE NUMBER : 1627487-2022-01230. IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY. DIAGNOSTICS REVEALED THAT BOTH THE LEADS WERE FRACTURED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHERE IN THE BOTH THE LEADS WERE EXPLANTED AND REPLACED ADDRESSING THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 13666225 ↗
  • InjuryMar 4, 2022Fracture; Fracture

    RELATED MANUFACTURER REFERENCE NUMBER : 1627487-2022-01183. IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY. DIAGNOSTICS REVEALED THAT BOTH THE LEADS WERE FRACTURED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHERE IN THE BOTH THE LEADS WERE EXPLANTED AND REPLACED ADDRESSING THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 13666221 ↗
  • InjuryJan 31, 2022Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2022-00558, 1627487-2022-00559, 1627487-2022-00560. IT WAS REPORTED THE PATIENTS PERIPHERAL SYSTEM WAS NOT PROVIDING EFFECTIVE THERAPY, THEREFORE THE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY RESTORED POST-OP.

    Patient problems: Inadequate Pain Relief

    MAUDE report 13399995 ↗
  • InjuryJan 31, 2022Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2022-00558, 1627487-2022-00559, 1627487-2022-00561. IT WAS REPORTED THE PATIENTS PERIPHERAL SYSTEM WAS NOT PROVIDING EFFECTIVE THERAPY, THEREFORE THE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY RESTORED POST-OP.

    Patient problems: Inadequate Pain Relief

    MAUDE report 13399950 ↗
  • InjuryJan 31, 2022Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2022-00559, 1627487-2022-00560, 1627487-2022-00561. IT WAS REPORTED THE PATIENTS PERIPHERAL SYSTEM WAS NOT PROVIDING EFFECTIVE THERAPY, THEREFORE THE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY RESTORED POST-OP.

    Patient problems: Inadequate Pain Relief

    MAUDE report 13399894 ↗
  • InjuryJan 31, 2022Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2022-00558, 1627487-2022-00560, 1627487-2022-00561. IT WAS REPORTED THE PATIENTS PERIPHERAL SYSTEM WAS NOT PROVIDING EFFECTIVE THERAPY, THEREFORE THE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY RESTORED POST-OP.

    Patient problems: Inadequate Pain Relief

    MAUDE report 13399871 ↗

Other STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) devices

All STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) devices →