Quattrode™ 3166 (UDI 05414734406109)Recalls & reported problems

St. Jude Medical, Inc. · Lead Wide Spaced, 60 cm · Class 2 · Implant

Quick summary

This St. Jude Medical Quattrode spinal cord stimulator lead is used for pain relief. Reports often describe pain relief being ineffective, along with leads migrating, high electrical resistance and fractures, and some mention surgery to reposition or remove leads. The 590 reports include injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
594
Deaths reported
0
Injuries reported
594
Recalls naming it
0

Summary

The FDA's MAUDE database contains 594 adverse event reports that list this exact device (DI 05414734406109), received between Jun 19, 2018 and Sep 22, 2026 — classified as 594 injuries.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (130 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 9 reportsFeb 2024: 4 reportsMar 2024: 3 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 3 reportsSep 2024: 1 reportsOct 2024: 4 reportsNov 2024: 3 reportsDec 2024: 3 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 4 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 5 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 2 reportsDec 2025: 4 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 5 reportsApr 2026: 5 reportsMay 2026: 0 reportsJun 2026: 5 reportsJul 2026: 3 reportsAug 2026: 1 reportsSep 2026: 5 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem290
  • Therapeutic or Diagnostic Output Failure130
  • High impedance73
  • Migration54
  • Fracture29
  • Impedance Problem16
  • Failure to Deliver Energy11
  • Inappropriate/Inadequate Shock/Stimulation10

Patient problems

  • Inadequate Pain Relief355
  • Erosion37
  • Implant Pain28
  • Pain28
  • No Clinical Signs, Symptoms or Conditions26
  • Discomfort24
  • Post Operative Wound Infection23
  • Unspecified Infection21

Recent adverse event reports

The 25 most recent of 594 reports, in the reporter's own words as published by the FDA.

  • InjurySep 22, 2026Migration

    It was reported that the patient's right Supra-orbital lead had migrated. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the lead was repositioned to address the issue. X-ray imaging revealed the right occipital lead was accidentally moved during the procedure. As a result, the patient returned to surgery on (b)(6) 2026 wherein the right occipital lead was repositioned to address the issue. Effective therapy was restored post operatively. It is unknown which leads were the right Supra-orbital and right occipital lead.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50860383 ↗
  • InjurySep 21, 2026Migration

    It was reported that following recent weight loss, the patient experienced pain at the IPG site. Additionally, the right Occipital lead migrated and bent and another lead migrated. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the IPG was repositioned, the right Occipital lead was explanted and replaced, and the other lead was repositioned to address the issue. It is unknown which lead migrated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846752 ↗
  • InjurySep 16, 2026High impedance

    It was reported that the patient experienced ineffective therapy. Troubleshooting revealed high impedances on the leads. Reprogramming has been unable to resolve the issue. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50793875 ↗
  • InjurySep 16, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50793873 ↗
  • InjurySep 16, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50793861 ↗
  • InjuryAug 18, 2026High impedance

    It was reported that the patient experienced ineffective therapy. Troubleshooting reveled high impedances on all leads. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50311741 ↗
  • InjuryJul 17, 2026Migration

    IT WAS REPORTED THAT THE PATIENT'S IPG MIGRATED AND THE LEADS WERE PULLING AT THEIR NECK. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE IPG WAS REPOSITIONED, AND ALL LEADS WERE SUTURED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25862050 ↗
  • InjuryJul 17, 2026Migration

    IT WAS REPORTED THAT THE PATIENT'S IPG MIGRATED AND THE LEADS WERE PULLING AT THEIR NECK. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE IPG WAS REPOSITIONED, AND ALL LEADS WERE SUTURED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25862045 ↗
  • InjuryJul 17, 2026Migration

    IT WAS REPORTED THAT THE PATIENT'S IPG MIGRATED AND THE LEADS WERE PULLING AT THEIR NECK. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE IPG WAS REPOSITIONED, AND ALL LEADS WERE SUTURED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25862036 ↗
  • InjuryJun 22, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENTS OCCIPITAL LEAD INCISION SITE WAS NOT HEALING PROPERLY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS AND ANCHORS WERE EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Wound Dehiscence

    MAUDE report 25588230 ↗
  • InjuryJun 22, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENTS OCCIPITAL LEAD INCISION SITE WAS NOT HEALING PROPERLY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS AND ANCHORS WERE EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Wound Dehiscence

    MAUDE report 25588226 ↗
  • InjuryJun 15, 2026Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. X-RAY IMAGING REVEALED MIGRATION OF ONE OF THE SUPRAORBITAL LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEAD WAS REPOSITIONED TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH SUPRAORBITAL LEAD MIGRATED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25512536 ↗
  • InjuryJun 14, 2026Use of Device Problem

    IT WAS REPORTED THAT DURING THE PATIENT'S SYSTEM EXPLANT THE LEADS WERE CUT AND THEN LEFT IMPLANTED IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 25497869 ↗
  • InjuryJun 14, 2026Use of Device Problem

    IT WAS REPORTED THAT DURING THE PATIENT'S SYSTEM EXPLANT THE LEADS WERE CUT AND THEN LEFT IMPLANTED IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 25497864 ↗
  • InjuryApr 23, 2026Failure to Deliver Energy

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO AUTO-REDUCING. TROUBLESHOOTING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN TWO NEW LEADS WERE PLACED, AND THE EXISTING LEADS WERE UNABLE TO BE EXPLANTED [RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-01965].

    Patient problems: Inadequate Pain Relief

    MAUDE report 24971456 ↗
  • InjuryApr 15, 2026High impedance

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. TROUBLESHOOTING REVELED HIGH IMPEDANCES ON BOTH OCCIPITAL LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS WERE REPLACED. DURING SURGERY, X-RAY IMAGING REVEALED BOTH SUPRA-ORBITAL LEADS WERE ACCIDENTALLY PULLED. AS A RESULT, THE SUPRA-ORBITAL LEADS WERE REPOSITIONED DURING THAT PROCEDURE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24885873 ↗
  • InjuryApr 15, 2026High impedance

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. TROUBLESHOOTING REVELED HIGH IMPEDANCES ON BOTH OCCIPITAL LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS WERE REPLACED. DURING SURGERY, X-RAY IMAGING REVEALED BOTH SUPRA-ORBITAL LEADS WERE ACCIDENTALLY PULLED. AS A RESULT, THE SUPRA-ORBITAL LEADS WERE REPOSITIONED DURING THAT PROCEDURE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24885802 ↗
  • InjuryApr 15, 2026Use of Device Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. TROUBLESHOOTING REVELED HIGH IMPEDANCES ON BOTH OCCIPITAL LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS WERE REPLACED. DURING SURGERY, X-RAY IMAGING REVEALED BOTH SUPRA-ORBITAL LEADS WERE ACCIDENTALLY PULLED. AS A RESULT, THE SUPRA-ORBITAL LEADS WERE REPOSITIONED DURING THAT PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885790 ↗
  • InjuryApr 15, 2026Use of Device Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. TROUBLESHOOTING REVELED HIGH IMPEDANCES ON BOTH OCCIPITAL LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS WERE REPLACED. DURING SURGERY, X-RAY IMAGING REVEALED BOTH SUPRA-ORBITAL LEADS WERE ACCIDENTALLY PULLED. AS A RESULT, THE SUPRA-ORBITAL LEADS WERE REPOSITIONED DURING THAT PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885789 ↗
  • InjuryMar 30, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN PATIENTS SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. THE INVESTIGATION WAS UNABLE TO DETERMINE THE LEAD THAT IS ASSOCIATED WITH THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24734971 ↗
  • InjuryMar 30, 2026Failure to Deliver Energy

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO AUTO REDUCING ON ONE LEAD DUE TO HIGH IMPEDANCES. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS RESTORED. NOTE: IT IS UNKNOWN WHICH LEAD IS RELATED TO THIS ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24719880 ↗
  • InjuryMar 16, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH LEAD CAUSED INEFFECTIVE THERAPY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24607799 ↗
  • InjuryMar 5, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT EXPERIENCED SWELLING AT THE SITE OF THEIR LEADS AND DRAINAGE DUE TO OPEN WOUNDS. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. PATIENT WAS PRESCRIBED ORAL ANTIBIOTICS.

    Patient problems: Swelling/ Edema

    MAUDE report 24525923 ↗
  • InjuryMar 5, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT EXPERIENCED SWELLING AT THE SITE OF THEIR LEADS AND DRAINAGE DUE TO OPEN WOUNDS. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. PATIENT WAS PRESCRIBED ORAL ANTIBIOTICS.

    Patient problems: Swelling/ Edema

    MAUDE report 24525888 ↗
  • InjuryJan 28, 2026Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED PATIENT EXPERIENCE DISCOMFORT AT THE SUPRA-ORBITAL LEAD SITES EVEN WHEN STIMULATION WAS OFF. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 24206324 ↗

Other STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) devices

All STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.