Quattrode™ 3166 (UDI 05414734406109)Recalls & reported problems
St. Jude Medical, Inc. · Lead Wide Spaced, 60 cm · Class 2 · Implant
All models of Quattrode (17)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This St. Jude Medical Quattrode spinal cord stimulator lead is used for pain relief. Reports often describe pain relief being ineffective, along with leads migrating, high electrical resistance and fractures, and some mention surgery to reposition or remove leads. The 590 reports include injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 587
- Deaths reported
- 0
- Injuries reported
- 587
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 587 adverse event reports that list this exact device (DI 05414734406109), received between Jun 19, 2018 and Jul 17, 2026 — classified as 587 injuries.
The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (130 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem290
- Therapeutic or Diagnostic Output Failure130
- High impedance68
- Migration52
- Fracture29
- Impedance Problem16
- Failure to Deliver Energy11
- Inappropriate/Inadequate Shock/Stimulation10
Patient problems
- Inadequate Pain Relief349
- Erosion37
- Implant Pain28
- Pain28
- No Clinical Signs, Symptoms or Conditions25
- Discomfort24
- Post Operative Wound Infection23
- Unspecified Infection21
Recent adverse event reports
The 25 most recent of 587 reports, in the reporter's own words as published by the FDA.
InjuryJul 17, 2026Migration
IT WAS REPORTED THAT THE PATIENT'S IPG MIGRATED AND THE LEADS WERE PULLING AT THEIR NECK. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE IPG WAS REPOSITIONED, AND ALL LEADS WERE SUTURED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25862050 ↗InjuryJul 17, 2026Migration
IT WAS REPORTED THAT THE PATIENT'S IPG MIGRATED AND THE LEADS WERE PULLING AT THEIR NECK. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE IPG WAS REPOSITIONED, AND ALL LEADS WERE SUTURED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25862045 ↗InjuryJul 17, 2026Migration
IT WAS REPORTED THAT THE PATIENT'S IPG MIGRATED AND THE LEADS WERE PULLING AT THEIR NECK. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE IPG WAS REPOSITIONED, AND ALL LEADS WERE SUTURED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25862036 ↗InjuryJun 22, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENTS OCCIPITAL LEAD INCISION SITE WAS NOT HEALING PROPERLY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS AND ANCHORS WERE EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Wound Dehiscence
MAUDE report 25588230 ↗InjuryJun 22, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENTS OCCIPITAL LEAD INCISION SITE WAS NOT HEALING PROPERLY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS AND ANCHORS WERE EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Wound Dehiscence
MAUDE report 25588226 ↗InjuryJun 15, 2026Migration
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. X-RAY IMAGING REVEALED MIGRATION OF ONE OF THE SUPRAORBITAL LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEAD WAS REPOSITIONED TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH SUPRAORBITAL LEAD MIGRATED.
Patient problems: Inadequate Pain Relief
MAUDE report 25512536 ↗InjuryJun 14, 2026Use of Device Problem
IT WAS REPORTED THAT DURING THE PATIENT'S SYSTEM EXPLANT THE LEADS WERE CUT AND THEN LEFT IMPLANTED IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 25497869 ↗InjuryJun 14, 2026Use of Device Problem
IT WAS REPORTED THAT DURING THE PATIENT'S SYSTEM EXPLANT THE LEADS WERE CUT AND THEN LEFT IMPLANTED IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 25497864 ↗InjuryApr 23, 2026Failure to Deliver Energy
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO AUTO-REDUCING. TROUBLESHOOTING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN TWO NEW LEADS WERE PLACED, AND THE EXISTING LEADS WERE UNABLE TO BE EXPLANTED [RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-01965].
Patient problems: Inadequate Pain Relief
MAUDE report 24971456 ↗InjuryApr 15, 2026High impedance
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. TROUBLESHOOTING REVELED HIGH IMPEDANCES ON BOTH OCCIPITAL LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS WERE REPLACED. DURING SURGERY, X-RAY IMAGING REVEALED BOTH SUPRA-ORBITAL LEADS WERE ACCIDENTALLY PULLED. AS A RESULT, THE SUPRA-ORBITAL LEADS WERE REPOSITIONED DURING THAT PROCEDURE.
Patient problems: Inadequate Pain Relief
MAUDE report 24885802 ↗InjuryApr 15, 2026Use of Device Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. TROUBLESHOOTING REVELED HIGH IMPEDANCES ON BOTH OCCIPITAL LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS WERE REPLACED. DURING SURGERY, X-RAY IMAGING REVEALED BOTH SUPRA-ORBITAL LEADS WERE ACCIDENTALLY PULLED. AS A RESULT, THE SUPRA-ORBITAL LEADS WERE REPOSITIONED DURING THAT PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24885790 ↗InjuryApr 15, 2026Use of Device Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. TROUBLESHOOTING REVELED HIGH IMPEDANCES ON BOTH OCCIPITAL LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE OCCIPITAL LEADS WERE REPLACED. DURING SURGERY, X-RAY IMAGING REVEALED BOTH SUPRA-ORBITAL LEADS WERE ACCIDENTALLY PULLED. AS A RESULT, THE SUPRA-ORBITAL LEADS WERE REPOSITIONED DURING THAT PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24885789 ↗InjuryMar 30, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN PATIENTS SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. THE INVESTIGATION WAS UNABLE TO DETERMINE THE LEAD THAT IS ASSOCIATED WITH THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 24734971 ↗InjuryMar 30, 2026Failure to Deliver Energy
IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO AUTO REDUCING ON ONE LEAD DUE TO HIGH IMPEDANCES. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS RESTORED. NOTE: IT IS UNKNOWN WHICH LEAD IS RELATED TO THIS ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 24719880 ↗InjuryMar 16, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH LEAD CAUSED INEFFECTIVE THERAPY.
Patient problems: Inadequate Pain Relief
MAUDE report 24607799 ↗InjuryMar 5, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED PATIENT EXPERIENCED SWELLING AT THE SITE OF THEIR LEADS AND DRAINAGE DUE TO OPEN WOUNDS. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. PATIENT WAS PRESCRIBED ORAL ANTIBIOTICS.
Patient problems: Swelling/ Edema
MAUDE report 24525923 ↗InjuryMar 5, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED PATIENT EXPERIENCED SWELLING AT THE SITE OF THEIR LEADS AND DRAINAGE DUE TO OPEN WOUNDS. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. PATIENT WAS PRESCRIBED ORAL ANTIBIOTICS.
Patient problems: Swelling/ Edema
MAUDE report 24525888 ↗InjuryJan 28, 2026Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED PATIENT EXPERIENCE DISCOMFORT AT THE SUPRA-ORBITAL LEAD SITES EVEN WHEN STIMULATION WAS OFF. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Undesired Nerve Stimulation
MAUDE report 24206324 ↗InjuryJan 28, 2026Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED PATIENT EXPERIENCE DISCOMFORT AT THE SUPRA-ORBITAL LEAD SITES EVEN WHEN STIMULATION WAS OFF. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Undesired Nerve Stimulation
MAUDE report 24206316 ↗InjuryJan 13, 2026High impedance
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO HIGH IMPEDANCES ON TWO LEADS. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH LEADS ARE AT FAULT.
Patient problems: Inadequate Pain Relief
MAUDE report 24060528 ↗InjuryJan 12, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY AND DISCOMFORT AT THE IPG SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 24046626 ↗InjuryDec 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT EXPERIENCED PAIN AT THE LEADS, IPG, AND EXTENSION SITES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Implant Pain
MAUDE report 23712228 ↗InjuryDec 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT EXPERIENCED PAIN AT THE LEADS, IPG, AND EXTENSION SITES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Implant Pain
MAUDE report 23712224 ↗InjuryDec 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT EXPERIENCED PAIN AT THE LEADS, IPG, AND EXTENSION SITES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Implant Pain
MAUDE report 23712223 ↗InjuryDec 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT EXPERIENCED PAIN AT THE LEADS, IPG, AND EXTENSION SITES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Implant Pain
MAUDE report 23712222 ↗
FDA clearance and classification
Cleared through Special 510(k) K072462 on Oct 11, 2007 (decision: Substantially Equivalent). Submitted by Advanced Neuromodulation Systems as “Wide Spaced Quattrode Leads”. The FDA review took 41 days. FDA record ↗
The FDA regulates product code GZB (stimulator, spinal-cord, implanted (pain relief)) under 21 CFR 882.5880 as a Class II device. 21 CFR 882.5880 ↗
Common questions about the Quattrode™ 3166
Has the Quattrode™ 3166 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05414734406109), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Quattrode™ 3166?
The FDA's MAUDE database lists 587 adverse event reports for this device, received between Jun 19, 2018 and Jul 17, 2026. In those reports, the most common specific device problems are “Therapeutic or Diagnostic Output Failure” (130), “High impedance” (68) and “Migration” (52); the most common patient problems are “Inadequate Pain Relief” (349), “Erosion” (37) and “Implant Pain” (28). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Quattrode™ 3166?
None of the reports are classified as deaths, 587 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Quattrode™ 3166 FDA approved?
It was cleared through Special 510(k) K072462 on Oct 11, 2007. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Quattrode™ 3166 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Quattrode™ 3166, and is it still sold?
The FDA's device database lists St. Jude Medical, Inc. as the company responsible for it (brand Quattrode™, model 3166). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) devices
- Octrode™ 3086St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)30 deaths1,533 reports
- Swift-Lock™ 1192St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)215 reports
- NA 3386St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)154 reports
- NA 3346St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)117 reports
- Quattrode® 3163St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)81 reports
- Quattrode™ 3163St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)52 reports
- Freedom-8A Receiver Kit Channel A US w/ Receiver FR8A-RCV-A0Stimwave Technologies Incorporated · Stimulator, Spinal-Cord, Implanted (Pain Relief)1 death52 reports
- Freedom System Accessories PDBT-915-2KCuronix LLC · Stimulator, Spinal-Cord, Implanted (Pain Relief)34 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page