Assurity™ PM2240Recalls & reported problems
St. Jude Medical, Inc. · DR Pacemaker DDDR · Class 3 · Implant
All models of Assurity (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This St. Jude Medical Assurity pacemaker has 4,070 adverse event reports, including 84 deaths, 1,834 injuries and 2,151 malfunctions. The reports mostly describe batteries running down early, the pacemaker switching to back-up mode and failed attempts to check the device. 1 open recall, from March 2021, covers some pacemakers where moisture can get in because of an epoxy manufacturing issue.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 4,076
- Deaths reported
- 84
- Injuries reported
- 1,829
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4,076 adverse event reports that list this exact device (DI 05414734507073), received between Jun 11, 2018 and Sep 30, 2026 — classified as 84 deaths, 1,829 injuries, 2,162 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (803 reports).
1 FDA recall record name this device's identifier, 1 still open.
9 other recalls cover devices cleared under the same FDA submission (P880086); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Mar 15, 2021Open, Classified
Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the device header and feedthrough
Z-1530-2021 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem830
- Premature Discharge of Battery803
- Defective Device761
- Pacemaker Found in Back-Up Mode525
- Failure to Interrogate450
- Over-Sensing249
- Under-Sensing119
- Pacing Problem116
Patient problems
- No Clinical Signs, Symptoms or Conditions1,926
- Insufficient Information920
- No Known Impact Or Consequence To Patient520
- Unspecified Infection247
- Arrhythmia139
- Discomfort105
- No Consequences Or Impact To Patient95
- Pocket Erosion73
Recent adverse event reports
The 25 most recent of 4,076 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50955067 ↗MalfunctionSep 30, 2026Premature Discharge of Battery
It was reported that the patient's pacemaker exhibited an Electronic Performance Indicator (EPI) alert, resulting in premature battery depletion. Additional information was unavailable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950080 ↗MalfunctionSep 28, 2026Over-Sensing; Pacing Problem
It was reported that the patient presented remotely via Merlin.net. It was noted that pacemaker exhibited far-field R-wave over-sensing, which resulted in inappropriate mode switch. No intervention was performed. There were no patient consequences.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50911135 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient's pacemaker, right atrial (RA) lead and right ventricular (RV) lead were explanted on (b)(6) 2026 due to infection. Temporary pacemaker was placed until the infection was resolved; the patient was subsequently implanted with a new pacemaker system on (b)(6) 2026. The patient was in stable condition.
Patient problems: Unspecified Infection
MAUDE report 50910666 ↗InjurySep 24, 2026Premature Discharge of Battery
Following the Electronics Performance Indicator (EPI), an alert was received by the clinician and device was explanted. The patient condition was not reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889193 ↗InjurySep 16, 2026Premature Discharge of Battery; Failure to Interrogate
It was reported that the patient's pacemaker was unable to be interrogated due to premature battery depletion. It was suspected that the pacemaker had a header issue. The patient was stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50792458 ↗MalfunctionSep 16, 2026Unexpected Mode Switch
It was reported during in clinic follow up that the patient's pacemaker went into backup mode due to event queue overflow. Programming changes were made and the device was restored. There were no patient consequences.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50790211 ↗MalfunctionSep 14, 2026Loose or Intermittent Connection
It was reported that the patient presented for a pacemaker implant procedure. During the procedure, the pacemaker's set screw could not be tightened. As a result, the pacemaker was not implanted. The physician completed the procedure using an alternate pacemaker. No adverse patient conditions were reported. The patient was discharged following the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50748000 ↗InjurySep 11, 2026Unexpected Mode Switch
Patient problems: Insufficient Information
MAUDE report 50744669 ↗MalfunctionSep 11, 2026Premature Discharge of Battery
New information indicates that the pacemaker was explanted and replaced. The patient condition was stable, before, during and after the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50744241 ↗MalfunctionSep 10, 2026No Apparent Adverse Event
Patient problems: Insufficient Information
MAUDE report 50721624 ↗MalfunctionSep 9, 2026Premature Discharge of Battery
Following the Electronics Performance Indicator (EPI), an alert was received by the clinician and awaiting device explant. Further information was requested, but not yet available. The patient condition was not reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50695272 ↗MalfunctionSep 9, 2026Failure to Disconnect
It was reported that the patient presented for a generator change procedure. During the procedure, it was noted that the right ventricular (RV) lead and right atrial (RA) lead could not be removed from the pacemaker header. The physician injected saline into the pacemaker header in an attempt to remove the leads. The RA lead was eventually removed; however, the insulation of the RV lead was damaged, and the RV lead could not be removed from the pacemaker header. The physician subsequently capped the RV lead. The existing RA lead and a new RV lead were connected to the new pacemaker, and the old pacemaker was explanted. The patient remained in stable condition throughout the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50685317 ↗MalfunctionSep 9, 2026Failure to Disconnect
It was reported that the patient presented to hospital for generator change procedure. During the procedure, it was noted that both the right atrial (RA) and right ventricular (RV) leads failed to remove from the header of pacemaker. The pacemaker was explanted and replaced; however, both the RA and RV leads remained implanted. The patient was in stable condition.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50683626 ↗MalfunctionSep 8, 2026No Apparent Adverse Event
Patient problems: Insufficient Information
MAUDE report 50680423 ↗InjurySep 8, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50679868 ↗MalfunctionSep 8, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50679800 ↗MalfunctionSep 8, 2026Premature Discharge of Battery
New information indicates that the pacemaker was explanted and replaced on (b)(6) 2026.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50679622 ↗MalfunctionSep 4, 2026Over-Sensing
It was reported that patient presented to clinic for follow up. Device evaluation identified oversensing due to noise on the pacemaker (PM), atrial lead, and right ventricular (RV) lead. Subsequent interrogation of the PM revealed recorded episodes of high ventricular rate (HVR). No intervention or procedure was performed. The patient was stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50604690 ↗MalfunctionSep 3, 2026No Apparent Adverse Event
The product was returned without a complaint associated to the product and there is no allegation of a malfunction for this product.
Patient problems: Insufficient Information
MAUDE report 50594816 ↗InjurySep 2, 2026Failure to Interrogate; No Apparent Adverse Event; No Pacing
Additional information received indicated the patient was asymptomatic and stable throughout the procedure. Corrected information received indicated the PM was explanted and replaced electively to implant an Aveir leadless system.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 50576774 ↗InjurySep 2, 2026Premature Discharge of Battery
Following the Electronics Performance Indicator (EPI), an alert was received by the clinician and device was explanted. The patient condition was stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50572060 ↗InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem
It was reported the device is subject to the Assurityand EndurityPacemakers Header Anomaly Advisory issued by Abbott on 15 March 2021.The pacemaker was explanted and replaced.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50552419 ↗MalfunctionSep 1, 2026No Apparent Adverse Event
This report is to advise of an event observed during analysis.
Patient problems: Insufficient Information
MAUDE report 50550319 ↗InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem
It was reported the device is subject to the Assurity and Endurity Pacemakers Header Anomaly Advisory issued by Abbott on 15 March 2021. The patient's pacemaker was prophylactically explanted and replaced. The patient remained stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50535910 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P880086. One submission often covers many models, so check the product and lot details in each record.
- Z-1495-2022 · Jun 17, 2022 · Open, ClassifiedMerlin PCS 3650 programmer Model 3330 software
- Z-1497-2022 · Jun 17, 2022 · Open, ClassifiedMerlin.net model MN5000 Software
- Z-1496-2022 · Jun 17, 2022 · Open, ClassifiedMerlin 2 PCS MER37000 programmer Model MER3400 software
- Z-0429-2012 · Sep 23, 2011 · TerminatedImplantable Pacemaker Accent DR models PM2110 and PM2210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.
- Z-0430-2012 · Sep 23, 2011 · TerminatedAnthem CRT-P models PM3110 and PM3210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.
- Z-0296-2007 · Oct 6, 2006 · TerminatedSt. Jude Medical APS III Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3500/3510
- Z-0297-2007 · Oct 6, 2006 · TerminatedSt. Jude Medical Merlin PCS Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3650
- Z-0298-2007 · Oct 6, 2006 · TerminatedSt. Jude Medical Identity SR Model 5172, Identity DR Model 5370, & Identity XL DR Model 5376; Pulse Generators (pacemakers)
- Z-0052-04 · Jul 31, 2003 · TerminatedPulse Generator
Other Pulse generator, permanent, implantable devices
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
- Zephyr™ XL DRSt. Jude Medical, Inc. · Pulse generator, permanent, implantable24 deaths944 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record