Ellipse™ CD2411-36QRecalls & reported problems

St. Jude Medical, Inc. · DR Implantable Cardioverter Defibrillator VVED DDDR · Class 3 · Implant

All models of Ellipse (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Medical Ellipse DR implantable defibrillator monitors heart rhythm and can deliver shocks. The 1,495 reports most often describe the device reading extra signals, unexpected resets, and inappropriate shocks, and some mention infections. The reports list 42 deaths and 509 injuries. Reports in the last 12 months number 39, compared with 81 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,498
Deaths reported
42
Injuries reported
509
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,498 adverse event reports that list this exact device (DI 05414734507615), received between Jun 11, 2018 and Sep 30, 2026 — classified as 42 deaths, 509 injuries, 947 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (592 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 12 reportsDec 2023: 8 reportsJan 2024: 8 reportsFeb 2024: 12 reportsMar 2024: 10 reportsApr 2024: 10 reportsMay 2024: 10 reportsJun 2024: 15 reportsJul 2024: 11 reportsAug 2024: 10 reportsSep 2024: 7 reportsOct 2024: 8 reportsNov 2024: 6 reportsDec 2024: 6 reportsJan 2025: 8 reportsFeb 2025: 8 reportsMar 2025: 7 reportsApr 2025: 8 reportsMay 2025: 10 reportsJun 2025: 7 reportsJul 2025: 11 reportsAug 2025: 5 reportsSep 2025: 4 reportsOct 2025: 1 reportsNov 2025: 7 reportsDec 2025: 4 reportsJan 2026: 5 reportsFeb 2026: 3 reportsMar 2026: 2 reportsApr 2026: 5 reportsMay 2026: 5 reportsJun 2026: 1 reportsJul 2026: 2 reportsAug 2026: 3 reportsSep 2026: 5 reportsOct 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing592
  • Adverse Event Without Identified Device or Use Problem237
  • Inappropriate or Unexpected Reset189
  • Inappropriate/Inadequate Shock/Stimulation183
  • Incorrect Interpretation of Signal104
  • Under-Sensing87
  • Signal Artifact/Noise51
  • Pacing Problem46

Patient problems

  • No Clinical Signs, Symptoms or Conditions629
  • No Known Impact Or Consequence To Patient386
  • Unspecified Infection155
  • Therapeutic Effects, Unexpected62
  • Arrhythmia59
  • No Consequences Or Impact To Patient58
  • Discomfort52
  • Insufficient Information45

Recent adverse event reports

The 25 most recent of 1,498 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951236 ↗
  • MalfunctionSep 17, 2026Over-Sensing

    Patient problems: Insufficient Information

    MAUDE report 50797110 ↗
  • MalfunctionSep 15, 2026Intermittent Capture

    Patient problems: Insufficient Information

    MAUDE report 50774222 ↗
  • MalfunctionSep 14, 2026Over-Sensing; Pacing Problem

    It was reported that a potential myopotentials oversensing resulting in episode of automatic mode switch (AMS) was observed via a review of remote transmission. No intervention was made at this time. No patient condition was reported.

    Patient problems: Insufficient Information

    MAUDE report 50753777 ↗
  • MalfunctionSep 8, 2026Signal Artifact/Noise

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50677630 ↗
  • MalfunctionAug 16, 2026Over-Sensing; Pacing Problem

    It was reported that the patient presented remotely via Merlin.net. It was noted that the Implantable Cardioverter Defibrillator (ICD) exhibited Far R-wave Oversensing resulting in inappropriate mode switch (AMS). No intervention was performed. The patient's condition is unknown.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50268017 ↗
  • MalfunctionAug 13, 2026Failure to Capture

    It was reported that the patient's Implantable Cardioverter Defibrillator (ICD) exhibited loss of capture. No additional information was received.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50240243 ↗
  • MalfunctionAug 12, 2026Failure to Capture

    It was reported that the patient presented remotely via merlin.net. Review of transmission revealed that the Implantable Cardioverter Defibrillator (ICD) exhibited loss of capture. No intervention was performed, and the patient will continue to be monitored. The patient was in stable condition.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50217073 ↗
  • MalfunctionJul 31, 2026Low impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50105053 ↗
  • MalfunctionJul 6, 2026Over-Sensing

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET. IT WAS DISCOVERED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED POST-PACED T-WAVE OVERSENSING. PROGRAMMING CHANGES WERE MADE, AND THE PATIENT DID NOT EXPERIENCE ANY CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25725406 ↗
  • MalfunctionJun 16, 2026Inappropriate or Unexpected Reset; Compatibility Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC FOR A MAGNETIC RESONANCE IMAGING (MRI) PROCEDURE. THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS PUT INTO MRI MODE PRIOR TO PROCEDURE BUT WAS NOT TAKEN OUT OF MRI MODE AFTERWARDS, CAUSING THE ICD TO INAPPROPRIATELY GO INTO BACKUP OPERATION. PROGRAMMING CHANGES WERE MADE TO RESET THE DEVICE. THE PATIENT WAS STABLE THROUGHOUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25519463 ↗
  • MalfunctionMay 26, 2026Data Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR DISPLAYED AN ERROR MESSAGE. PROGRAMMING CHANGES WERE PERFORMED TO RESTORE THE DEVICE BACK TO CHRONIC SETTINGS. FURTHER INFORMATION WAS REQUESTED HOWEVER, WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25290408 ↗
  • InjuryMay 21, 2026Premature Discharge of Battery

    IT WAS REPORTED DURING IN CLINIC FOLLOW UP THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED PREMATURE BATTERY DEPLETION. THE PRODUCT WAS EXPLANTED AND SUCCESSFULLY REPLACED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25244076 ↗
  • InjuryMay 13, 2026Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT A PATIENT PRESENTED WITH INAPPROPRIATE ANTI-TACHYCARDIA PACING NOTED DUE TO INCORRECT INTERPRETATION OF SIGNAL. PROGRAMMING CHANGES WERE RECOMMENDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25165131 ↗
  • MalfunctionMay 8, 2026Over-Sensing

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED POST-PACED T-WAVE OVERSENSING (PPTWOS). NO INTERVENTION WAS REPORTED AND PATIENT STATUS WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25122678 ↗
  • InjuryMay 5, 2026Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE DEVICE DELIVERED INAPPROPRIATE SHOCKS DUE TO AN INCORRECT INTERPRETATION OF SIGNAL. PROGRAMMING CHANGES WERE MADE. THE PATIENT WAS ASYMPTOMATIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25089699 ↗
  • MalfunctionApr 29, 2026Under-Sensing

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED INTERMITTENT ATRIAL UNDER-SENSING. NO INTERVENTION WAS REPORTED, AND THE PATIENT WILL CONTINUE TO BE MONITORED. THERE WERE NO PATIENT CONSEQUENCES NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25034888 ↗
  • MalfunctionApr 23, 2026Signal Artifact/Noise; Over-Sensing

    IT WAS REPORTED THAT NOISE WAS OBSERVED ON THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24972166 ↗
  • MalfunctionApr 14, 2026Over-Sensing; Pacing Asynchronously

    IT WAS REPORTED THAT A PATIENT PRESENTED WITH OVER-SENSING OF COMPETITIVE ATRIAL PACING NOTED ON THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR. NO PROGRAMMING OR INTERVENTION WAS PLANNED OR PERFORMED. NO INFORMATION WAS PROVIDED REGARDING PATIENT CONSEQUENCES BUT THE PATIENT WAS PRESUMED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24874920 ↗
  • MalfunctionApr 11, 2026Inappropriate or Unexpected Reset

    IT WAS REPORTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS FOUND TO BE IN BACKUP VVI MODE UPON INTERROGATION. NO INTERVENTION WAS PERFORMED, AND THE PATIENT'S STATUS WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24848010 ↗
  • InjuryApr 8, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AND REPLACED DUE TO PATIENT DISCOMFORT. THE PATIENT REMAINED STABLE.

    Patient problems: Discomfort

    MAUDE report 24820933 ↗
  • MalfunctionMar 30, 2026Failure to Capture

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET. REVIEW OF THE TRANSMISSION REVEALED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR EXHIBITED LOSS OF CAPTURE. FURTHER INFORMATION WAS REQUESTED HOWEVER, WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24714934 ↗
  • MalfunctionMar 9, 2026Under-Sensing

    NEW INFORMATION RECEIVED NOTES THE PATIENT WAS ASYMPTOMATIC, PROGRAMMING CHANGES WERE MADE, THE PATIENT WAS TO CONTINUE WITH FOLLOW UP IN CLINIC AND WAS STABLE BEFORE, DURING AND AFTER THE EVENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24548360 ↗
  • InjuryFeb 20, 2026Inappropriate/Inadequate Shock/Stimulation

    NEW INFORMATION RECEIVED NOTES THAT THE EPISODES OF INAPPROPRIATE SHOCK DELIVERED BY THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WERE NOTED IN CLINIC. NO INTERVENTION WAS PERFORMED. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24404967 ↗
  • InjuryFeb 20, 2026Unexpected Mode Switch

    NEW INFORMATION RECEIVED NOTES THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AND REPLACED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24403183 ↗

Other Pulse generator, permanent, implantable devices

All Pulse generator, permanent, implantable devices →