Skip to content

CardioMEMS™ GuidewireRecalls & reported problems

Cardiomems, Inc. · 0.018 in / 260 cm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05414734509565), received between Mar 28, 2019 and Mar 28, 2019 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break2
  • Difficult to Advance1

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 28, 2019Break; Break

    IT WAS REPORTED DURING THE SENSOR IMPLANT PROCEDURE, THE GUIDEWIRE SNAPPED INSIDE THE SWAN WHILE THE PHYSICIAN WAS TRYING TO ADVANCE INTO THE IMPLANT SITE. THE ISSUE WAS OBSERVED PRIOR TO THE INTRODUCTION OF THE CARDIOMEMS DELIVERY SYSTEM. REPORTEDLY, THE PHYSICIAN WAS ABLE TO SNARE THE TWO PIECED OUT AND SUCCESSFULLY IMPLANT THE SENSOR AND CALIBRATE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8460725 ↗
  • InjuryMar 28, 2019Break; Difficult to Advance; Break; Difficult to Advance

    IT WAS REPORTED DURING THE SENSOR IMPLANT PROCEDURE, THE GUIDEWIRE STOPPED ADVANCING THROUGH THE SWAN AND THE DISTAL TIP BROKE OFF DURING PULL BACK. THE ENTIRE WIRE WAS ABLE TO BE RETRIEVED. REPORTEDLY, THE PHYSICIAN WAS ABLE TO COMPLETE THE IMPLANT WITH A NEW GUIDEWIRE WITHOUT ANY PATIENT CONSEQUENCES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8460667 ↗