Octrode™ 3186 (UDI 05415067017246)Recalls & reported problems

St. Jude Medical, Inc. · Lead Kit, 60cm Length · Class 3 · Implant

All models of Octrode (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This is a St. Jude Medical Octrode spinal cord stimulator lead kit, an implanted pain relief device. Its reports mostly mention inadequate pain relief, leads that migrate, high electrical resistance and fractures, with some saying surgery may follow. There are 32,014 reports, including 30 deaths and 31,908 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32,096
Deaths reported
30
Injuries reported
31,988
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32,096 adverse event reports that list this exact device (DI 05415067017246), received between Jun 11, 2018 and Sep 30, 2026 — classified as 30 deaths, 31,988 injuries, 77 malfunctions.

The most frequently reported specific device problem is “Migration” (9,882 reports).

14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 358 reportsDec 2023: 340 reportsJan 2024: 424 reportsFeb 2024: 420 reportsMar 2024: 336 reportsApr 2024: 362 reportsMay 2024: 344 reportsJun 2024: 371 reportsJul 2024: 364 reportsAug 2024: 356 reportsSep 2024: 361 reportsOct 2024: 437 reportsNov 2024: 363 reportsDec 2024: 326 reportsJan 2025: 327 reportsFeb 2025: 339 reportsMar 2025: 459 reportsApr 2025: 253 reportsMay 2025: 386 reportsJun 2025: 441 reportsJul 2025: 409 reportsAug 2025: 350 reportsSep 2025: 215 reportsOct 2025: 273 reportsNov 2025: 385 reportsDec 2025: 438 reportsJan 2026: 356 reportsFeb 2026: 340 reportsMar 2026: 436 reportsApr 2026: 356 reportsMay 2026: 271 reportsJun 2026: 210 reportsJul 2026: 158 reportsAug 2026: 96 reportsSep 2026: 290 reportsOct 2026: 0 reports202420252026459
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration9,882
  • High impedance9,185
  • Adverse Event Without Identified Device or Use Problem5,162
  • Therapeutic or Diagnostic Output Failure4,512
  • Fracture2,444
  • Impedance Problem988
  • Low impedance482
  • Inappropriate/Inadequate Shock/Stimulation320

Patient problems

  • Inadequate Pain Relief24,906
  • No Clinical Signs, Symptoms or Conditions2,766
  • Failure of Implant2,237
  • Undesired Nerve Stimulation810
  • No Consequences Or Impact To Patient779
  • Post Operative Wound Infection733
  • Unspecified Infection541
  • Cerebrospinal Fluid Leakage340

Recent adverse event reports

The 25 most recent of 32,096 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50955028 ↗
  • InjurySep 30, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50955026 ↗
  • InjurySep 30, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50954525 ↗
  • InjurySep 30, 2026Migration

    Patient problems: Inadequate Pain Relief

    MAUDE report 50953852 ↗
  • InjurySep 30, 2026Migration

    Patient problems: Inadequate Pain Relief

    MAUDE report 50953840 ↗
  • InjurySep 30, 2026Therapeutic or Diagnostic Output Failure

    Patient problems: Inadequate Pain Relief

    MAUDE report 50949226 ↗
  • InjurySep 30, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50947567 ↗
  • InjurySep 30, 2026Migration

    It was reported that the patient lead had migrated and was confirmed with imaging. As a result, surgical intervention was undertaken wherein the lead was explanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946025 ↗
  • InjurySep 29, 2026Therapeutic or Diagnostic Output Failure

    Patient problems: Inadequate Pain Relief

    MAUDE report 50941202 ↗
  • InjurySep 29, 2026Inappropriate/Inadequate Shock/Stimulation

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 50941201 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Gastrointestinal Problem

    MAUDE report 50941199 ↗
  • InjurySep 29, 2026Therapeutic or Diagnostic Output Failure

    Patient problems: Inadequate Pain Relief

    MAUDE report 50940954 ↗
  • InjurySep 29, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50939588 ↗
  • InjurySep 29, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50939583 ↗
  • InjurySep 29, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50939149 ↗
  • InjurySep 29, 2026Fracture

    Patient problems: Inadequate Pain Relief

    MAUDE report 50939147 ↗
  • OtherSep 29, 2026Therapeutic or Diagnostic Output Failure

    Patient problems: Inadequate Pain Relief

    MAUDE report 50938918 ↗
  • InjurySep 29, 2026High impedance

    It was reported patient experienced increasing pain and stated that she was unable to adjust the stimulation intensity due to high impedances. Surgical intervention was undertaken wherein the lead was explanted and replaced to address the issue. Therapy was restored post-op.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50938812 ↗
  • InjurySep 29, 2026High impedance

    It was reported that the patient experienced ineffective therapy following a fall. Troubleshooting revealed high impedances on the entirety of both leads. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50938084 ↗
  • InjurySep 29, 2026High impedance

    It was reported that the patient experienced ineffective therapy following a fall. Troubleshooting revealed high impedances on the entirety of both leads. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50938081 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient's IPG site opened and the system was protruding. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the system was explanted to address the issue.

    Patient problems: Erosion

    MAUDE report 50937942 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient's IPG site opened and the system was protruding. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the system was explanted to address the issue.

    Patient problems: Erosion

    MAUDE report 50937941 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced discomfort at the occipital lead site. As such, surgical intervention took place on (b)(6) 2026 wherein the leads were repositioned to address the issue.

    Patient problems: Implant Pain

    MAUDE report 50937617 ↗
  • InjurySep 29, 2026High impedance

    It was reported that patient's system was unable to enter MRI mode. Troubleshooting revealed high impedances on the left lead. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the lead was replaced to address the issue. Therapy was restored post-operatively.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937615 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced discomfort at the occipital lead site. As such, surgical intervention took place on (b)(6) 2026 wherein the leads were repositioned to address the issue.

    Patient problems: Implant Pain

    MAUDE report 50937607 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2507-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →

Sources and updates