Skip to content

Proclaim™ 5 EliteRecalls & reported problems

St. Jude Medical, Inc. · Implantable Pulse Generator · Class 3 · Implant

All models of Proclaim (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This St. Jude Proclaim 5 Elite spinal cord stimulator is named in 9,650 adverse event reports, including 10 deaths and 8,077 injuries. The reports mostly describe pain relief falling short, battery and wireless problems, and implants removed or replaced after discomfort or lead damage. One open recall, from May 2024, concerns a shorter time between the battery replacement warning and end of service than the labeling states.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
9,661
Deaths reported
10
Injuries reported
8,090
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9,661 adverse event reports that list this exact device (DI 05415067020192), received between Jun 9, 2018 and Sep 29, 2026 — classified as 10 deaths, 8,090 injuries, 1,561 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (1,920 reports).

1 FDA recall record name this device's identifier, 1 still open.

14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 35 reportsDec 2023: 47 reportsJan 2024: 44 reportsFeb 2024: 50 reportsMar 2024: 42 reportsApr 2024: 39 reportsMay 2024: 25 reportsJun 2024: 38 reportsJul 2024: 24 reportsAug 2024: 25 reportsSep 2024: 19 reportsOct 2024: 26 reportsNov 2024: 20 reportsDec 2024: 9 reportsJan 2025: 21 reportsFeb 2025: 19 reportsMar 2025: 21 reportsApr 2025: 14 reportsMay 2025: 16 reportsJun 2025: 19 reportsJul 2025: 15 reportsAug 2025: 19 reportsSep 2025: 12 reportsOct 2025: 11 reportsNov 2025: 16 reportsDec 2025: 11 reportsJan 2026: 18 reportsFeb 2026: 12 reportsMar 2026: 14 reportsApr 2026: 14 reportsMay 2026: 18 reportsJun 2026: 12 reportsJul 2026: 4 reportsAug 2026: 8 reportsSep 2026: 13 reportsOct 2026: 0 reports20242025202650
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 16, 2024Open, Classified

    The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.

    Z-2369-2024 · Root cause: Vendor change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2,355
  • Battery Problem1,920
  • Wireless Communication Problem1,696
  • Insufficient Device Problem Information1,261
  • Premature Elective Replacement Indicator1,120
  • Therapeutic or Diagnostic Output Failure400
  • Application Program Freezes or Fails to Launch329
  • No Device Output187

Patient problems

  • Inadequate Pain Relief5,339
  • No Consequences Or Impact To Patient1,217
  • Failure of Implant961
  • Implant Pain699
  • Pain527
  • No Clinical Signs, Symptoms or Conditions525
  • Discomfort384
  • No Known Impact Or Consequence To Patient335

Recent adverse event reports

The 25 most recent of 9,661 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Insufficient Device Problem Information

    It was reported the patient is going to undergo an explant for an unknown reason. Further information is not available.

    Patient problems: Insufficient Information

    MAUDE report 50936147 ↗
  • InjurySep 29, 2026Insufficient Device Problem Information

    It was reported the patient's IPG reached end of life. It is unknown if this occurred prematurely. Surgical intervention took place where the IPG was replaced and therapy was restored.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50935630 ↗
  • InjurySep 29, 2026Insufficient Device Problem Information

    It was reported patient underwent surgical intervention at an unknown time wherein the entire system was explanted to address an unknown issue.

    Patient problems: Insufficient Information

    MAUDE report 50935148 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Implant Pain

    MAUDE report 50925103 ↗
  • InjurySep 28, 2026Migration

    It was reported that the patient experienced ineffective therapy, uncomfortable stimulation, and pain at the IPG site. Troubleshooting revealed patient's IPG flipped in pocket and the leads migrated. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the system was explanted and replaced to address the issue. Effective therapy was restored post-operatively.

    Patient problems: Implant Pain

    MAUDE report 50916637 ↗
  • InjurySep 25, 2026Insufficient Device Problem Information

    Patient problems: Insufficient Information

    MAUDE report 50902938 ↗
  • InjurySep 22, 2026Wireless Communication Problem

    It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the IPG became unable to communicate with external devices. Troubleshooting was unable to resolve the issue. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50856010 ↗
  • InjurySep 22, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that following significant weight loss, the patient experienced pain at the IPG site. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Implant Pain

    MAUDE report 50855405 ↗
  • InjurySep 22, 2026Wireless Communication Problem

    It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the IPG became unable to communicate with external devices. Troubleshooting has been unable to resolve the issue. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50849989 ↗
  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    It was reported patient experienced discomfort at implant site due to size and location of IPG. Surgical intervention may be pending to address the issue.

    Patient problems: Implant Pain

    MAUDE report 50828433 ↗
  • InjurySep 16, 2026Insufficient Device Problem Information

    Patient problems: Insufficient Information

    MAUDE report 50796632 ↗
  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced discomfort at the IPG site. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Implant Pain

    MAUDE report 50791915 ↗
  • InjurySep 14, 2026Insufficient Device Problem Information

    It was reported that the patient¿s IPG was explanted and replaced with a competitor¿s IPG. The reason for explant and date of explant is unknown.

    Patient problems: Insufficient Information

    MAUDE report 50760895 ↗
  • InjuryAug 27, 2026Insufficient Device Problem Information

    It was reported that the patient had their IPG explanted on an unknown date for an unknown reason.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50476309 ↗
  • InjuryAug 26, 2026Insufficient Device Problem Information

    It was reported that the patient had their system explanted on an unknown date for an unknown reason.

    Patient problems: Insufficient Information

    MAUDE report 50463541 ↗
  • InjuryAug 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient developed an infection. As a result, the SCS system was surgically removed. Date of explant is unknown. Patient has now completely recovered

    Patient problems: Unspecified Infection

    MAUDE report 50260376 ↗
  • InjuryAug 6, 2026Insufficient Device Problem Information

    It was reported that the patient had their system explanted in (b)(6) 2024 for an unknown reason.

    Patient problems: Insufficient Information

    MAUDE report 50164456 ↗
  • InjuryAug 4, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that following recent weight loss, the patient experienced pain at the IPG site. As a result, surgical intervention was undertaken on (b)(6)2026 wherein the IPG was explanted and replaced to address the issue.

    Patient problems: Implant Pain

    MAUDE report 50131538 ↗
  • InjuryAug 4, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced an infection. As a result, surgical intervention was undertaken at an unknown date in 2025, during which the system was explanted to address the issue.

    Patient problems: Unspecified Infection

    MAUDE report 50129395 ↗
  • InjuryAug 4, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced an infection. As a result, surgical intervention was undertaken at an unknown date in 2025, during which the system was explanted to address the issue.

    Patient problems: Unspecified Infection

    MAUDE report 50129384 ↗
  • InjuryAug 3, 2026Insufficient Device Problem Information

    It was reported that the patient had their IPG explanted on an unknown date for an unknown reason.

    Patient problems: Insufficient Information

    MAUDE report 50114375 ↗
  • InjuryJul 31, 2026Wireless Communication Problem

    It was reported that following an unrelated surgical procedure wherein the device was placed into surgery mode, the IPG became unable to communicate with external devices. Troubleshooting was unable to resolve the issue. As a result, surgical intervention maybe undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50104335 ↗
  • InjuryJul 29, 2026Temperature Problem

    It was reported that during an MRI while MRI mode was enabled patient experienced a burning sensation at the IPG site. As a result, MRI was aborted to address the issue. Symptoms resolved shortly after.

    Patient problems: Burning Sensation

    MAUDE report 50073191 ↗
  • InjuryJul 27, 2026Insufficient Device Problem Information

    It was reported that the patient had their system explanted on an unknown date for an unknown reason.

    Patient problems: Insufficient Information

    MAUDE report 50031690 ↗
  • InjuryJul 13, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED DISCOMFORT AT THE IPG SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS EXPLANTED AND REPLACED AT A NEW POCKET SITE.

    Patient problems: Implant Pain

    MAUDE report 25803891 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2507-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page