Skip to content

Infinity™ 6660Recalls & reported problems

St. Jude Medical, Inc. · Implantable Pulse Generator · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Medical Infinity 6660 is an implanted, battery-powered stimulator used to treat Parkinsonian tremor. Many of the 313 reports describe batteries wearing out or the device being removed and replaced, and some describe poor pain relief. There were 278 injury reports and 5 deaths, and no recalls name this device. Reports fell to 8 in the latest 12 months from 23 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
312
Deaths reported
5
Injuries reported
277
Recalls naming it
0

Summary

The FDA's MAUDE database contains 312 adverse event reports that list this exact device (DI 05415067020246), received between Jun 20, 2018 and Mar 13, 2026 — classified as 5 deaths, 277 injuries, 30 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (99 reports).

5 other recalls cover devices cleared under the same FDA submission (P140009); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 6 reportsFeb 2024: 2 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 8 reportsJun 2024: 5 reportsJul 2024: 4 reportsAug 2024: 3 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 3 reportsFeb 2025: 5 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 3 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Battery Problem99
  • Adverse Event Without Identified Device or Use Problem75
  • Insufficient Device Problem Information34
  • Wireless Communication Problem27
  • Premature Elective Replacement Indicator24
  • Migration23
  • No Apparent Adverse Event18
  • High impedance10

Patient problems

  • Inadequate Pain Relief116
  • Failure of Implant60
  • No Clinical Signs, Symptoms or Conditions50
  • No Consequences Or Impact To Patient34
  • Movement Disorder30
  • Unspecified Infection20
  • Twiddlers Syndrome17
  • Implant Pain16

Recent adverse event reports

The 25 most recent of 312 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 13, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SURGICAL INTERVENTION ON AN UNKNOWN DATE WHEREIN THE SYSTEM WAS EXPLANTED FOR AN UNKNOWN REASON.

    Patient problems: Insufficient Information

    MAUDE report 24591418 ↗
  • InjuryDec 31, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SURGICAL INTERVENTION ON (B)(6) 2025 IN WHICH THEIR IPGS WERE EXPLANTED AND REPLACED FOR AN UNKNOWN REASON.

    Patient problems: Insufficient Information

    MAUDE report 23936739 ↗
  • InjuryDec 12, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SURGICAL INTERVENTION ON AN UNKNOWN DATE WHEREIN THE IPG WAS REPLACED FOR AN UNKNOWN REASON.

    Patient problems: Insufficient Information

    MAUDE report 23792118 ↗
  • InjuryDec 4, 2025Impedance Problem; Impedance Problem

    IT WAS REPORTED THAT THE PATIENT'S SYSTEM HAD IMPEDANCES ISSUES AND THE IPG WAS UNABLE TO ENTER MRI MODE. SURGICAL INTERVENTION WAS UNDERTAKEN AND THE IPG WAS EXPLANTED AND REPLACED, RESOLVING THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23717684 ↗
  • InjuryNov 24, 2025High impedance; High impedance

    IT WAS REPORTED THE PATIENT IS EXPERIENCING INEFFECTIVE THERAPY DUE TO HIGH IMPEDANCES AND UNABLE TO SET IPG TO MRI MODE. IN TURN, SURGICAL INTERVENTION TOOK PLACE WHEREIN THE EXTENSION AND IPG WERE EXPLANTED AND REPLACED. EFFECTIVE THERAPY RESTORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23631929 ↗
  • InjuryNov 11, 2025Low impedance; Low impedance

    IT WAS REPORTED THAT THE PATIENT HAS LOW IMPEDANCE ON THE RIGHT SIDE. PHYSICIAN BELIEVES IT IS CAUSED BY FLUID IN THE IPG HEADER. SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23526450 ↗
  • InjuryNov 6, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    ADDITIONAL INFORMATION REPORTED PATIENT'S IPG WAS EXPLANTED DUE TO IPG REACHING END OF LIFE. IT WAS UNKNOWN IF THIS OCCURRED PREMATURELY. PATIENT EXTENSION WAS EXPLANTED DUE TO HIGH IMPEDANCES AND FRACTURE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23486249 ↗
  • InjurySep 9, 2025Migration; Migration

    IT WAS REPORTED THAT BOTH IPGS MIGRATED. IT WAS ALSO REPORTED THAT FOLLOWING AN UNRELATED SURGICAL PROCEDURE WHEREIN THE DEVICE WAS NOT PLACED INTO SURGERY MODE, ONE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. TROUBLESHOOTING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23000780 ↗
  • InjuryAug 24, 2025Migration; Migration

    IT WAS REPORTED THAT FOLLOWING AN UNRELATED SURGICAL PROCEDURE WHEREIN THE DEVICE WAS NOT PLACED INTO SURGERY MODE, THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. IT WAS ALSO REPORTED THAT THE IPG MIGRATED AND PATIENT EXPERIENCED PAIN AT THE IPG SITE. TROUBLESHOOTING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Implant Pain; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22881704 ↗
  • InjuryAug 24, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT FOLLOWING AN UNRELATED SURGICAL PROCEDURE WHEREIN THE DEVICE WAS NOT PLACED INTO SURGERY MODE, THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. IT WAS ALSO REPORTED THAT THE IPG MIGRATED AND PATIENT EXPERIENCED PAIN AT THE IPG SITE. TROUBLESHOOTING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Failure of Implant; Movement Disorder

    MAUDE report 22881703 ↗
  • InjuryAug 4, 2025High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT HAD HIGH IMPEDANCES ON LEFT CONTACT. IT IS UNCLEAR IF THE HIGH IMPEDANCES WERE PRESENT ON THE LEAD EXTENSION OR THE IPG. AS A RESULT, SURGICAL INTERVENTION TOOK PLACE ON (B)(6) 2025 WHEREIN BOTH THE LEFT LEAD EXTENSION AND IPG WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Movement Disorder

    MAUDE report 22709983 ↗
  • InjuryJul 30, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT THE PATIENT'S SYSTEMS HAD LOW IMPEDANCES ON THE LEFT SIDE, AND HIGH IMPEDANCES ON THE RIGHT SIDE. SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN IT WAS DISCOVERED THAT THE LEFT EXTENSION PULLED OUT FROM THE IPG, AND THE RIGHT LEAD PULLED OUT FROM THE EXTENSION. THE LEFT EXTENSION WAS REINSERTED INTO THE IPG, AND THE RIGHT LEAD WAS REINSERTED INTO THE RIGHT EXTENSION TO RESOLVE THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22671703 ↗
  • InjuryJul 11, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT PATIENTS IPG ENTERED SERVICE APP MODE FOR AN UNKNOWN REASON, IPG WAS RECOVERED, AND THERAPY WAS RESTORED. PATIENTS IPG THEN ENTER SERVICE APP MODE AGAIN, SEVERAL DAYS LATER, AGAIN FOR AN UNKNOWN REASON. AT THIS TIME THE DEVICE IS UNABLE COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 22484944 ↗
  • InjuryJul 10, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PATIENT¿S LEFT IPG HAD REACHED END OF LIFE. IT IS UNKNOWN IF THIS OCCURRED PREMATURELY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 22475713 ↗
  • InjuryJun 4, 2025Separation Failure; Separation Failure

    IT WAS REPORTED THAT DUE DURING A NORMAL END OF LIFE IPG REPLACEMENT PROCEDURE ON (B)(6) 2025, THE EXTENSION STUCK IN THE IPG HEADER. THE EXTENSION COULD NOT BE DISCONNECTED FROM THE HEADER. AS SUCH, THE EXTENSION WAS EXPLANTED ALONG WITH THE IPG TO ADDRESS THE ISSUE. THERAPY WAS CONFIRMED POST OP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22144922 ↗
  • InjuryMay 9, 2025Battery Problem; Wireless Communication Problem; Battery Problem; Wireless Communication Problem

    IT WAS REPORTED THE PATIENT'S IPG HAD CONNECTION ISSUES WITH EXTERNAL DEVICES. IN TURN, SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THIS SITUATION.

    Patient problems: Failure of Implant; Inadequate Pain Relief; Movement Disorder; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21998463 ↗
  • InjuryApr 21, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT DIAGNOSTICS INDICATED THE ELECTIVE REPLACEMENT INDICATOR (ERI) TRIGGERED EARLIER THAN INTENDED. THE DEVICE HAD THE APPROPRIATE LEVEL OF BATTERY VOLTAGE TO PROVIDE THERAPY. THE DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 21877648 ↗
  • InjuryApr 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED INDICATES, PATIENT'S INFECTION HAS RESOLVED.

    Patient problems: Pocket Erosion; Unspecified Infection

    MAUDE report 21870430 ↗
  • InjuryMar 6, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PATIENT¿S IPG HAD REACHED END OF LIFE. IT IS UNKNOWN IF THIS OCCURRED PREMATURELY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 TO REPLACE THE IPG. THERAPY WAS RESTORED POST-OPERATIVELY.

    Patient problems: Failure of Implant; Movement Disorder

    MAUDE report 21540848 ↗
  • InjuryFeb 26, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED A LOW BATTERY WARNING. IT IS UNKNOWN IF THIS OCCURRED PREMATURELY. AS A RESULT, SURGICAL INTERVENTION OCCURRED ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED RESOLVING THE ISSUE.

    Patient problems: Failure of Implant

    MAUDE report 21472962 ↗
  • InjuryFeb 23, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PATIENT'S IPGS WERE REACHING END OF LIFE. SURGICAL INTERVENTION WAS UNDERTAKEN AND THE IPGS WERE EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 21442833 ↗
  • InjuryFeb 23, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PATIENT'S IPGS WERE REACHING END OF LIFE. SURGICAL INTERVENTION WAS UNDERTAKEN AND THE IPGS WERE EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 21442831 ↗
  • InjuryFeb 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT EXPERIENCED ANXIETY AND COMMITTED SUICIDE AFTER A DBS IMPLANT. PATIENT DID RECEIVE ADEQUATE PSYCHIATRIC CARE AND SPECIFIC MEDICATIONS.

    Patient problems: Anxiety; Suicidal Ideation

    MAUDE report 21381778 ↗
  • InjuryFeb 3, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PATIENT¿S IPG HAD REACHED END OF LIFE. IT IS UNKNOWN IF THIS OCCURRED PREMATURELY. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Failure of Implant; Movement Disorder

    MAUDE report 21300644 ↗
  • InjuryJan 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE BILATERAL IPG SITES IN THEIR ABDOMEN. AS A RESULT, SURGICAL INTERVENTION TOOK PLACE ON (B)(6) 2025 WHEREIN BOTH THE IPGS WERE EXPLANTED AND REPLACED WITH A SINGLE DBS IPG TO ADDRESS THE ISSUE.

    Patient problems: Implant Pain

    MAUDE report 21279582 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140009. One submission often covers many models, so check the product and lot details in each record.

  • Z-0459-2026 · Oct 6, 2025 · Open, Classified
    Liberta RC DBS IPG (Implantable Pulse Generator), Model Number 62400
  • Z-2693-2017 · Jun 2, 2017 · Terminated
    Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions.
  • Z-2696-2017 · Jun 2, 2017 · Terminated
    Infinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
  • Z-2695-2017 · Jun 2, 2017 · Terminated
    Infinity 7, Implantable Pulse Generator, REF 6662, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
  • Z-2694-2017 · Jun 2, 2017 · Terminated
    Infinity 5, Implantable Pulse Generator, REF 6661, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I

Other STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR devices

All STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page