St. Jude Medical Infinity™ DBS System (REF 6178)Recalls & reported problems
St. Jude Medical, Inc. · 8CH Directional Lead, 30 cm, 0.5 · Class 3 · Implant
All models of St. Jude Medical Infinity (13)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05415067020642), received between Jul 17, 2020 and Jul 17, 2020 — classified as 4 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
5 other recalls cover devices cleared under the same FDA submission (P140009); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture2
- High impedance2
- Material Fragmentation2
Patient problems
- Inadequate Pain Relief2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJul 17, 2020Fracture; High impedance; Fracture; High impedance
RELATED MANUFACTURER REPORT NUMBER: 1627487-2020-22588. IT WAS REPORTED THAT PATIENT EXPERIENCED A RECENT FALL AND HAD LOSS OF EFFECTIVE STIMULATION. LEAD FRACTURE AND HIGH IMPEDANCE WAS CONFIRMED INTRAOPERATIVELY VIA TRIAL CABLE AND EXTERNAL PULSE GENERATOR. LEFT LEAD WAS EXPLANTED, AND A NEW LEAD WAS IMPLANTED HOWEVER IT IS UNKNOWN WHICH LEAD IS THE LEFT LEAD THEREFORE BOTH LEADS ARE BEING REPORTED. IMPEDANCES WERE CHECKED POST PROCEDURE AND ALL VALUES WERE WITHIN NORMAL OPTIMAL RANGE AND EFFECTIVE THERAPY WAS RESUMED. PATIENT WAS STABLE.
Patient problems: Inadequate Pain Relief
MAUDE report 10292344 ↗InjuryJul 17, 2020Fracture; High impedance; Fracture; High impedance
RELATED MANUFACTURER REPORT NUMBER 1627487-2020-22932. IT WAS REPORTED THAT PATIENT EXPERIENCED A RECENT FALL AND HAD LOSS OF EFFECTIVE STIMULATION. LEAD FRACTURE AND HIGH IMPEDANCE WAS CONFIRMED INTRAOPERATIVELY VIA TRIAL CABLE AND EXTERNAL PULSE GENERATOR. LEFT LEAD WAS EXPLANTED, AND A NEW LEAD WAS IMPLANTED HOWEVER IT IS UNKNOWN WHICH LEAD IS THE LEFT LEAD THEREFORE BOTH LEADS ARE BEING REPORTED. IMPEDANCES WERE CHECKED POST PROCEDURE AND ALL VALUES WERE WITHIN NORMAL OPTIMAL RANGE AND EFFECTIVE THERAPY WAS RESUMED. PATIENT WAS STABLE.
Patient problems: Inadequate Pain Relief
MAUDE report 10292328 ↗InjuryJul 17, 2020Material Fragmentation; Material Fragmentation
RELATED MANUFACTURER REPORT NUMBER 1627487-2020-22589. IT WAS REPORTED THAT DURING A LEAD EXPLANT PROCEDURE, PHYSICIAN WAS UNABLE TO EXPLANT THE ENTIRE LEAD AND A PARTIAL OF THE LEAD REMAINS SUBCUTANEOUSLY IMPLANTED IN THE BRAIN TISSUE. IT IS UNKNOWN WHICH LEAD HAD THE ISSUE THEREFORE BOTH LEADS ARE BEING REPORTED. A NEW LEAD WAS SUCCESSFULLY IMPLANTED. PATIENT WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10292313 ↗InjuryJul 17, 2020Material Fragmentation; Material Fragmentation
RELATED MANUFACTURER REPORT NUMBER 1627487-2020-22934. IT WAS REPORTED THAT DURING A LEAD EXPLANT PROCEDURE, PHYSICIAN WAS UNABLE TO EXPLANT THE ENTIRE LEAD AND A PARTIAL OF THE LEAD REMAINS SUBCUTANEOUSLY IMPLANTED IN THE BRAIN TISSUE. IT IS UNKNOWN WHICH LEAD HAD THE ISSUE THEREFORE BOTH LEADS ARE BEING REPORTED. A NEW LEAD WAS SUCCESSFULLY IMPLANTED. PATIENT WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10292307 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P140009. One submission often covers many models, so check the product and lot details in each record.
- Z-0459-2026 · Oct 6, 2025 · Open, ClassifiedLiberta RC DBS IPG (Implantable Pulse Generator), Model Number 62400
- Z-2695-2017 · Jun 2, 2017 · TerminatedInfinity 7, Implantable Pulse Generator, REF 6662, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
- Z-2693-2017 · Jun 2, 2017 · TerminatedInfinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions.
- Z-2694-2017 · Jun 2, 2017 · TerminatedInfinity 5, Implantable Pulse Generator, REF 6661, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
- Z-2696-2017 · Jun 2, 2017 · TerminatedInfinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
Other STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR devices
- Activa® RC 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR2 deaths1,849 reports
- Activa® 37603Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR8 deaths1,093 reports
- Infinity™ 6662St. Jude Medical, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR1 recall3 deaths803 reports
- St. Jude Medical Infinity™ DBS SystemSt. Jude Medical, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR1 recall3 deaths751 reports
- Activa® SC 37603Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR1 recall1 death443 reports
- Infinity™ 7St. Jude Medical, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR2 recalls3 deaths368 reports
- St. Jude Medical Infinity™ DBS SystemSt. Jude Medical, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR406 reports
- Activa® RC 37612Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian tremor2 deaths371 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page