Axium™ MN10450-50A (UDI 05415067025531)Recalls & reported problems

Spinal Modulation, Inc. · 50cm SlimTip Implant Lead Kit · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Spinal Modulation Axium lead, used with a nerve stimulator for pain relief, has 1,827 adverse event reports on record, with no deaths and 1,825 injuries. Most reports describe the pain relief becoming ineffective, with leads moving out of place, fracturing, or showing high electrical resistance, and some of these led to surgery to replace them. Some reports describe part of a lead being left in the patient during removal. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,836
Deaths reported
0
Injuries reported
1,834
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,836 adverse event reports that list this exact device (DI 05415067025531), received between Jun 11, 2018 and Sep 28, 2026 — classified as 1,834 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Migration” (358 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 4 reportsFeb 2024: 2 reportsMar 2024: 7 reportsApr 2024: 0 reportsMay 2024: 4 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 4 reportsNov 2024: 0 reportsDec 2024: 4 reportsJan 2025: 4 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 4 reportsJul 2025: 7 reportsAug 2025: 4 reportsSep 2025: 2 reportsOct 2025: 2 reportsNov 2025: 2 reportsDec 2025: 7 reportsJan 2026: 6 reportsFeb 2026: 2 reportsMar 2026: 7 reportsApr 2026: 4 reportsMay 2026: 2 reportsJun 2026: 2 reportsJul 2026: 4 reportsAug 2026: 1 reportsSep 2026: 6 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem678
  • Migration358
  • High impedance296
  • Therapeutic or Diagnostic Output Failure242
  • Fracture195
  • Impedance Problem44
  • Material Fragmentation35
  • Insufficient Device Problem Information34

Patient problems

  • Inadequate Pain Relief1,519
  • No Consequences Or Impact To Patient87
  • No Clinical Signs, Symptoms or Conditions73
  • Undesired Nerve Stimulation31
  • Fall29
  • No Known Impact Or Consequence To Patient28
  • Failure of Implant27
  • Pain22

Recent adverse event reports

The 25 most recent of 1,836 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Migration

    It was reported that the patient experienced ineffective therapy. X-ray imaging revealed migration of two leads. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the Left L4-L5 was explanted, and surgical intervention was undertaken on (b)(6) 2026 wherein the Left L3-L4 was explanted and replaced to address the issue. It is unknown which leads migrated.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50923567 ↗
  • InjurySep 28, 2026Migration

    Patient problems: Inadequate Pain Relief

    MAUDE report 50923564 ↗
  • InjurySep 14, 2026Migration

    It was reported that the patient had ineffective therapy. X-ray imaging revealed migration of leads. As a result, the entire system was explanted to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50757001 ↗
  • InjurySep 14, 2026Migration

    It was reported that the patient had ineffective therapy. X-ray imaging revealed migration of leads. As a result, the entire system was explanted to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50757000 ↗
  • InjurySep 14, 2026Migration

    It was reported that the patient had ineffective therapy. X-ray imaging reveled migration of leads. As a result, the entire system was explanted to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50756995 ↗
  • InjurySep 14, 2026Material Fragmentation

    It was reported that during a lead revision on 21AUG2026 (Related Manufacturer Reference Number:1627487-2026-01395) a portion of lead was left implanted. As such, surgical intervention may take place at a later date to address the issue.Investigation was unable to determine which of the leads is the left lead.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755119 ↗
  • InjuryAug 20, 2026Therapeutic or Diagnostic Output Failure

    It was reported that the patient experienced ineffective therapy. As a result, surgical intervention may be undertaken to address the issue. It is unknown which lead is liable.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50358570 ↗
  • InjuryJul 30, 2026High impedance

    It was reported that the patient experienced ineffective therapy following a fall. Troubleshooting revealed high impedances on all leads. As a result, surgical intervention was undertaken on 16JUL2026 wherein leads were explanted and replaced to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50089819 ↗
  • InjuryJul 30, 2026High impedance

    It was reported that the patient experienced ineffective therapy following a fall. Troubleshooting revealed high impedances on all leads. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein leads were explanted and replaced to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50089818 ↗
  • InjuryJul 27, 2026Migration

    It was reported that the patient experienced ineffective therapy. Diagnostic testing revealed high impedances on three leads, and X-ray imaging showed migration of the left L5 lead. As a result, surgical intervention was performed on (b)(6) 2026. During the procedure, it was observed that the leads had become disconnected from the IPG because they had not been properly tightened into the device. Consequently, the IPG and the migrated lead were replaced, while the remaining original leads were tightened into the new IPG to address the issue. Following the procedure, the patient continued to experience ineffective therapy because the leads that were not replaced were not providing stimulation. Therefore, additional surgical intervention may be undertaken to address the remaining issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50047162 ↗
  • InjuryJul 27, 2026Therapeutic or Diagnostic Output Failure

    It was reported that the patient experienced ineffective therapy. Diagnostic testing revealed high impedances on three leads, and X-ray imaging showed migration of the left L5 lead. As a result, surgical intervention was performed on 06JUL2026. During the procedure, it was observed that the leads had become disconnected from the IPG because they had not been properly tightened into the device. Consequently, the IPG and the migrated lead were replaced, while the remaining original leads were tightened into the new IPG to address the issue. Following the procedure, the patient continued to experience ineffective therapy because the leads that were not replaced were not providing stimulation. Therefore, additional surgical intervention may be undertaken to address the remaining issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50047077 ↗
  • InjuryJun 29, 2026Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. X-RAY IMAGING REVEALED MIGRATION OF THE L4 AND L5 LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25652075 ↗
  • InjuryJun 29, 2026Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. X-RAY IMAGING REVEALED MIGRATION OF THE L4 AND L5 LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25652018 ↗
  • InjuryMay 27, 2026Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. X-RAY IMAGING REVEALED MIGRATION OF BOTH L5 AND S1 LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN BOTH LEADS WERE EXPLANTED AND THE S1 LEAD WAS REPLACED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25310620 ↗
  • InjuryMay 12, 2026Use of Device Problem

    IT WAS REPORTED THAT DURING A DRG SYSTEM EXPLANT SURGERY, A PORTION OF A LEAD WAS LEFT IMPLANTED IN THE PATIENT. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH LEAD WAS LEFT IMPLANTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 25149061 ↗
  • InjuryApr 30, 2026High impedance

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. DIAGNOSTICS REVEALED HIGH IMPEDANCES ON ONE OF THE PATIENT'S LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN AN ADDITIONAL LEAD WAS ADDED TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH LEAD HAD HIGH IMPEDANCES.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25053448 ↗
  • InjuryApr 23, 2026Use of Device Problem

    IT WAS REPORTED THAT DURING AN EXPLANT PROCEDURE ON (B)(6) 2026 [RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-01991] THE L5 LEAD WAS PARTIALLY REMOVED, AND A PORTION OF THE LEAD WAS LEFT IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 24966765 ↗
  • InjuryApr 23, 2026High impedance

    IT WAS REPORTED THAT PATIENT'S SYSTEM WAS UNABLE TO ENTER MRI MODE. TROUBLESHOOTING REVEALED HIGH IMPEDANCES ON THE L5 LEAD. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEAD WAS PARTIALLY REMOVED, AND A PORTION OF THE LEAD WAS LEFT IN THE PATIENT [RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-01992].

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24966763 ↗
  • InjuryApr 6, 2026Fracture

    IT WAS REPORTED THAT IMPEDANCE CHECK REVEALED HIGH IMPEDANCE ON THE LEAD. AS SUCH, SURGICAL INTERVENTION TOOK PLACE WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. AFTER THE EXPLANT, THE LEAD APPEARED TO BE FRACTURED. REPORTEDLY, EFFECTIVE THERAPY WAS CONFIRMED POST OP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24794002 ↗
  • InjuryMar 26, 2026Migration

    IT WAS REPORTED THAT X-RAY IMAGING REVEALED MIGRATION OF THREE LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE S1 AND L5 LEADS WERE EXPLANTED AND REPLACED, AND THE L3 LEAD WAS EXPLANTED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698384 ↗
  • InjuryMar 26, 2026Migration

    IT WAS REPORTED THAT X-RAY IMAGING REVEALED MIGRATION OF THREE LEADS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE S1 AND L5 LEADS WERE EXPLANTED AND REPLACED, AND THE L3 LEAD WAS EXPLANTED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698376 ↗
  • InjuryMar 24, 2026High impedance

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. IMPEDANCE CHECK REVEALED HIGH IMPEDANCES ON ALL CONTACTS OF THE LEFT LEAD. AS SUCH, SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE TO ADDRESS THE ISSUE. INVESTIGATION WAS UNABLE TO DETERMINE WHICH OF THE LEADS IS THE LEFT LEAD.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24681415 ↗
  • InjuryMar 20, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN IN 2016 WHEREIN THE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. THE INVESTIGATION WAS UNABLE TO DETERMINE THE LEAD THAT ASSOCIATED WITH THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24654231 ↗
  • InjuryMar 10, 2026Fracture

    IT WAS REPORTED THAT THE PATIENT'S L5 LEAD WAS FRACTURED, AND THE L3 LEAD HAD LOW IMPENDENCES. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN BOTH LEADS WERE EXPLANTED AND REPLACED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24560700 ↗
  • InjuryMar 6, 2026Fracture

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY AFTER TAKING A FALL. X-RAY IMAGING REVEALED THAT ALL THE LEADS MIGRATED AND WERE FRACTURED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED POST PROCEDURE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24536427 ↗

Other Dorsal root ganglion stimulator for pain relief devices

All Dorsal root ganglion stimulator for pain relief devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page