Skip to content

SlimTip™ MN10450-50ARecalls & reported problems

Advanced Neuromodulation Systems, Inc. · DRG Implant Lead Kit · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This SlimTip dorsal root ganglion stimulator lead kit is an implanted device used for pain relief. Reports mostly describe leads moving out of place, high electrical resistance, and inadequate pain relief, with some patients needing surgery. There are 8,232 reports, including 7 deaths and 8,189 injuries. Reports fell to 749 in the last 12 months from 1,211 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
8,308
Deaths reported
7
Injuries reported
8,266
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8,308 adverse event reports that list this exact device (DI 05415067027153), received between Jun 18, 2018 and Sep 30, 2026 — classified as 7 deaths, 8,266 injuries, 35 malfunctions.

The most frequently reported specific device problem is “Migration” (3,207 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 106 reportsDec 2023: 106 reportsJan 2024: 110 reportsFeb 2024: 125 reportsMar 2024: 124 reportsApr 2024: 109 reportsMay 2024: 102 reportsJun 2024: 81 reportsJul 2024: 73 reportsAug 2024: 103 reportsSep 2024: 100 reportsOct 2024: 122 reportsNov 2024: 110 reportsDec 2024: 92 reportsJan 2025: 115 reportsFeb 2025: 117 reportsMar 2025: 99 reportsApr 2025: 66 reportsMay 2025: 131 reportsJun 2025: 113 reportsJul 2025: 122 reportsAug 2025: 107 reportsSep 2025: 58 reportsOct 2025: 81 reportsNov 2025: 97 reportsDec 2025: 88 reportsJan 2026: 99 reportsFeb 2026: 85 reportsMar 2026: 95 reportsApr 2026: 87 reportsMay 2026: 66 reportsJun 2026: 41 reportsJul 2026: 36 reportsAug 2026: 29 reportsSep 2026: 102 reportsOct 2026: 0 reports202420252026131
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration3,207
  • High impedance1,367
  • Therapeutic or Diagnostic Output Failure1,272
  • Adverse Event Without Identified Device or Use Problem1,065
  • Fracture925
  • Use of Device Problem268
  • Material Fragmentation173
  • Impedance Problem134

Patient problems

  • Inadequate Pain Relief6,588
  • Failure of Implant576
  • No Clinical Signs, Symptoms or Conditions527
  • Undesired Nerve Stimulation175
  • No Consequences Or Impact To Patient170
  • Foreign Body In Patient147
  • Cerebrospinal Fluid Leakage129
  • Device Embedded In Tissue or Plaque113

Recent adverse event reports

The 25 most recent of 8,308 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50954942 ↗
  • InjurySep 30, 2026Use of Device Problem

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 50950484 ↗
  • InjurySep 29, 2026Migration

    It was reported that the patient experienced ineffective therapy. Imaging revealed migration of one lead. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the lead was replaced to address the issue. Therapy was restored post-operatively.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50937993 ↗
  • InjurySep 29, 2026Migration

    It was reported that following a fall the patient experienced ineffective therapy due to one lead fracturing and the other lead migrating. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the leads were replaced to address the issue. Therapy was restored post-operatively.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50937766 ↗
  • InjurySep 29, 2026Fracture

    It was reported that following a fall the patient experienced ineffective therapy due to one lead fracturing and the other lead migrating. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the leads were replaced to address the issue. Therapy was restored post-operatively.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50937762 ↗
  • InjurySep 29, 2026Fracture

    It was reported that the patient experienced ineffective therapy following a fall. X-ray imaging revealed fracture of the left T12 lead and migration of the left L2 lead. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the leads were explanted and replaced to address the issue. It is unknown which lead fractured, and which lead migrated.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50937328 ↗
  • InjurySep 29, 2026High impedance

    It was reported that the patient experienced ineffective therapy due to high impedances on the leads. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the leads were replaced to address the issue. Therapy was restored post-operatively.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50936464 ↗
  • InjurySep 29, 2026High impedance

    It was reported that the patient experienced ineffective therapy due to high impedances on the leads. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the leads were replaced to address the issue. Therapy was restored post-operatively.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50936455 ↗
  • InjurySep 29, 2026Material Fragmentation

    It was reported the patient experienced ineffective stimulation due to both leads migrating. Surgical intervention took place on (b)(6) 2026 wherein one lead was explanted and the second lead broke and partial remains implanted. Two new leads were implanted, and effective stimulation was established post-op. Surgical intervention may take place at a later date to address the partial lead that remains implanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933048 ↗
  • InjurySep 29, 2026Migration

    It was reported the patient experienced ineffective stimulation due to both leads migrating. Surgical intervention took place on (b)(6) 2026 wherein one lead was explanted and the second lead broke and partial remains implanted. Two new leads were implanted, and effective stimulation was established post-op. Surgical intervention may take place at a later date to address the partial lead that remains implanted.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50933025 ↗
  • InjurySep 29, 2026Migration

    It was reported the patient experienced ineffective stimulation due to both leads migrating. Surgical intervention took place on (b)(6) 2026 wherein one lead was explanted and the second lead broke and partial remains implanted. Two new leads were implanted, and effective stimulation was established post-op. Surgical intervention may take place at a later date to address the partial lead that remains implanted.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50932992 ↗
  • InjurySep 29, 2026Migration

    It was reported that patient experienced ineffective therapy, patients leads had high and low impedances, imaging revealed patients 3rd lead migrated. Patient also experienced sharp sensation and heating at the IPG site. As a result, surgical intervention surgical intervention was undertaken on 21 September wherein IPG and leads were explanted and replaced to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50931798 ↗
  • InjurySep 28, 2026Migration

    Patient problems: Inadequate Pain Relief

    MAUDE report 50926572 ↗
  • InjurySep 28, 2026High impedance

    Patient problems: Inadequate Pain Relief

    MAUDE report 50918918 ↗
  • InjurySep 28, 2026Migration

    It was reported that the patient experienced ineffective therapy following a fall. X-ray imaging revealed migration of the S1-S2 lead. Trouble shooting revealed high impedances on the L5-S1 lead. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the system was explanted to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50918912 ↗
  • InjurySep 27, 2026Migration

    Patient problems: Inadequate Pain Relief

    MAUDE report 50909802 ↗
  • InjurySep 27, 2026Migration

    Patient problems: Inadequate Pain Relief

    MAUDE report 50909801 ↗
  • InjurySep 25, 2026Therapeutic or Diagnostic Output Failure

    Patient problems: Inadequate Pain Relief

    MAUDE report 50903429 ↗
  • InjurySep 25, 2026Inappropriate/Inadequate Shock/Stimulation

    Patient problems: Burning Sensation

    MAUDE report 50902954 ↗
  • InjurySep 25, 2026Migration

    It was reported that following a fall the patient experienced ineffective therapy due to impedances on one lead and migration of the other. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50899834 ↗
  • InjurySep 25, 2026Impedance Problem

    It was reported that following a fall the patient experienced ineffective therapy due to impedances on one lead and migration of the other. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50899828 ↗
  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced an infection at the lead site. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the entire system was explanted to address the issue. The patient was treated with oral antibiotics, IV antibiotics, and hospitalization to address the issue. It is unknown which lead is related to this issue.

    Patient problems: Unspecified Infection

    MAUDE report 50897612 ↗
  • InjurySep 25, 2026High impedance

    It was reported that the patient experienced ineffective therapy following falls. Diagnostics revealed high impedances on the leads. It was also reported that the patient is scheduled for surgery for the IPG for an unknown reason. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50893374 ↗
  • InjurySep 25, 2026High impedance

    It was reported that the patient experienced ineffective therapy following falls. Diagnostics revealed high impedances on the leads. It was also reported that the patient is scheduled for surgery for the IPG for an unknown reason. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50893372 ↗
  • InjurySep 24, 2026Fracture

    Patient problems: Inadequate Pain Relief

    MAUDE report 50892116 ↗

Other Dorsal root ganglion stimulator for pain relief devices

All Dorsal root ganglion stimulator for pain relief devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page