St. Jude Medical Infinity™ DBS System (REF 6173)Recalls & reported problems

St. Jude Medical, Inc. · 8CH Directional Lead, 40 cm, 1.5 · Class 3 · Implant

All models of St. Jude Medical Infinity (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Medical Infinity DBS system is an implanted brain stimulator used for Parkinsonian tremor. Its 182 reports include 5 deaths and 177 injuries. Many describe ineffective therapy, often followed by repeat surgery to move or replace a lead, and some describe infections or wound problems. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
178
Deaths reported
5
Injuries reported
173
Recalls naming it
0

Summary

The FDA's MAUDE database contains 178 adverse event reports that list this exact device (DI 05415067030337), received between Apr 11, 2019 and May 15, 2026 — classified as 5 deaths, 173 injuries.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (48 reports).

5 other recalls cover devices cleared under the same FDA submission (P140009); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 2 reportsJun 2024: 6 reportsJul 2024: 1 reportsAug 2024: 3 reportsSep 2024: 1 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 2 reportsJan 2025: 4 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 2 reportsMay 2025: 4 reportsJun 2025: 1 reportsJul 2025: 6 reportsAug 2025: 3 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 1 reportsFeb 2026: 3 reportsMar 2026: 3 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem92
  • Therapeutic or Diagnostic Output Failure48
  • High impedance10
  • Migration10
  • Use of Device Problem10
  • Insufficient Device Problem Information4
  • No Apparent Adverse Event2
  • Therapy Delivered to Incorrect Body Area2

Patient problems

  • Inadequate Pain Relief44
  • Movement Disorder35
  • Unspecified Infection28
  • Erosion21
  • Post Operative Wound Infection18
  • No Clinical Signs, Symptoms or Conditions14
  • Failure of Implant4
  • Insufficient Information4

Recent adverse event reports

The 25 most recent of 178 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 15, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. REPROGRAMMING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Movement Disorder

    MAUDE report 25179649 ↗
  • InjuryApr 3, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED A RETURN OF SYMPTOMS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEAD TO WAS REPLACED ADDRESS THE ISSUE. THERAPY WAS RESTORED POST-OPERATIVELY.

    Patient problems: Movement Disorder

    MAUDE report 24779128 ↗
  • InjuryMar 30, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. SURGICAL INTERVENTION TOOK PLACE ON (B)(6) 2026 WHEREIN THE LEAD WAS REPOSITIONED TO ADDRESS THE ISSUE.

    Patient problems: Movement Disorder

    MAUDE report 24722247 ↗
  • InjuryMar 30, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. SURGICAL INTERVENTION TOOK PLACE ON (B)(6) 2026 WHEREIN A LEAD WAS ADDED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS ESTABLISHED.

    Patient problems: Movement Disorder

    MAUDE report 24717544 ↗
  • InjuryMar 18, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Movement Disorder

    MAUDE report 24630441 ↗
  • InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT'S SYSTEM WAS INFECTED AT THE BURR HOLE AND IPG POCKET SITES. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM EXPLANTED. PATIENT WAS PRESCRIBED ANTIBIOTICS TO ADDRESS THE INFECTION.

    Patient problems: Post Operative Wound Infection

    MAUDE report 24381553 ↗
  • InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT'S SYSTEM WAS INFECTED AT THE BURR HOLE AND IPG POCKET SITES. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM EXPLANTED. PATIENT WAS PRESCRIBED ANTIBIOTICS TO ADDRESS THE INFECTION.

    Patient problems: Post Operative Wound Infection

    MAUDE report 24381531 ↗
  • InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT EXPERIENCED AN INFECTION AT THE BURR HOLE CAP SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE BURR HOLE CAP AND RIGHT LEAD WERE EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Unspecified Infection

    MAUDE report 24377086 ↗
  • InjuryJan 26, 2026Use of Device Problem

    IT WAS REPORTED THE PHYSICIAN ACCIDENTALLY MOVED THE PATIENT'S LEAD AND THE PATIENT NO LONGER HAD EFFECTIVE THERAPY. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS REPLACED RESTORING THERAPY.

    Patient problems: Movement Disorder

    MAUDE report 24184441 ↗
  • InjuryDec 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT EXPERIENCED INFECTION AT THE LEAD/EXTENSION SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025, WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. PATIENT WAS ADMINISTERED IV AND ORAL ANTIBIOTICS, AND THE INFECTION HAS RESOLVED.

    Patient problems: Post Operative Wound Infection

    MAUDE report 23854379 ↗
  • InjuryDec 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT EXPERIENCED INFECTION AT THE LEAD/EXTENSION SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025, WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. PATIENT WAS ADMINISTERED IV AND ORAL ANTIBIOTICS AND THE INFECTION HAS RESOLVED.

    Patient problems: Post Operative Wound Infection

    MAUDE report 23854375 ↗
  • InjuryDec 2, 2025High impedance; No Apparent Adverse Event; High impedance; No Apparent Adverse Event

    ADDITIONAL INFORMATION RECEIVED INDICATES THERE WAS NO ALLEGATION AGAINST THE LEAD. AS SUCH, THE LEAD IS NO LONGER DEEMED REPORTABLE.

    Patient problems: Movement Disorder; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23690376 ↗
  • InjuryNov 17, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT HAD EXCESSIVE TREMORS IN THE RIGHT SIDE OF THE BODY. SURGICAL INTERVENTION WAS UNDERTAKEN, AND THE LEFT LEAD WAS EXPLANTED AND REPLACED.

    Patient problems: Movement Disorder

    MAUDE report 23575966 ↗
  • InjuryNov 13, 2025Therapeutic or Diagnostic Output Failure; Insufficient Device Problem Information; Therapeutic or Diagnostic Output Failure; Insufficient Device Problem Information

    IT WAS REPORTED PATIENT IS SCHEDULED FOR A LEAD REVISION FOR AN UNKNOWN REASON.

    Patient problems: Inadequate Pain Relief; Insufficient Information

    MAUDE report 23543383 ↗
  • InjuryNov 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT'S RIGHT LEAD WAS EXPOSED. IN TURN, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE SURGEON OPENED A NEW SKIN FLAP TO TUCK THE LEAD UNDER TO ADDRESS THE ISSUE. THERE WAS NO INFECTION.

    Patient problems: Erosion

    MAUDE report 23523910 ↗
  • InjurySep 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT SUFFERED A STROKE AND WAS HOSPITALIZED IN THE INTENSIVE CARE UNIT. THE INVESTIGATION WAS UNABLE TO DETERMINE THE LEAD THAT ASSOCIATED WITH THE ISSUE.

    Patient problems: Stroke/CVA

    MAUDE report 22946574 ↗
  • InjuryAug 28, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED PATIENT'S RIGHT LEAD HAD HIGH IMPEDANCES AND WAS BENT AT THE TIP WHICH WAS FOUND DURING PERM IMPLANT ON (B)(6) 2025. THE PHYSICIAN BENT THE LEAD BACK MANUALLY. THE PROCEDURE WAS EXTENDED AS A RESULT. ISSUE WAS RESOLVED AND THERAPY WAS ESTABLISHED POST-OP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22916660 ↗
  • InjuryAug 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION AT THE LEFT BURR HOLE CAP AND WAS HOSPITALIZED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. THE PATIENT WAS ADMINISTERED IV ANTIBIOTICS TO ADDRESS THE ISSUE. THE INFECTION HAS RESOLVED.

    Patient problems: Unspecified Infection

    MAUDE report 22791223 ↗
  • InjuryAug 6, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE STIMULATION. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Movement Disorder

    MAUDE report 22730083 ↗
  • InjuryJul 28, 2025Use of Device Problem; Malposition of Device; Use of Device Problem; Malposition of Device

    IT WAS REPORTED THAT THE PATIENT'S LEAD WAS ORIGINALLY IMPLANTED TO ANTERIOR. SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 IN WHICH THE LEAD WAS REPOSITIONED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22628332 ↗
  • InjuryJul 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT HAD AN INFECTION AT THE LEAD SITE. SURGICAL INTERVENTION WAS UNDERTAKEN AT AN UNKNOWN TIME WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED. PATIENT WAS PRESCRIBED ANTIBIOTICS TO ADDRESS THE ISSUE.

    Patient problems: Post Operative Wound Infection

    MAUDE report 22570262 ↗
  • InjuryJul 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PATIENT HAD AN INFECTION AT THE LEAD SITE. SURGICAL INTERVENTION WAS UNDERTAKEN AT AN UNKNOWN TIME WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED. PATIENT WAS PRESCRIBED ANTIBIOTICS TO ADDRESS THE ISSUE.

    Patient problems: Post Operative Wound Infection

    MAUDE report 22570162 ↗
  • InjuryJul 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN EROSION ON THE RIGHT LEAD SITE AND BURR HOLE CAP. SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ENTIRE RIGHT SIDED SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Erosion

    MAUDE report 22542247 ↗
  • InjuryJul 10, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION IN MANAGING HER TREMORS. AS SUCH, SURGICAL INTERVENTION MAY OCCUR TO ADDRESS THE ISSUE. THE INVESTIGATION DID NOT DETERMINE WHICH LEAD RESULTED IN INEFFECTIVE STIMULATION.

    Patient problems: Movement Disorder

    MAUDE report 22472194 ↗
  • InjuryJul 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE LEFT LEAD AND LEFT EXTENSION HAD ERODED THROUGH THE SKIN. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE LEAD AND EXTENSION WERE EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Skin Erosion

    MAUDE report 22412790 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140009. One submission often covers many models, so check the product and lot details in each record.

  • Z-0459-2026 · Oct 6, 2025 · Open, Classified
    Liberta RC DBS IPG (Implantable Pulse Generator), Model Number 62400
  • Z-2696-2017 · Jun 2, 2017 · Terminated
    Infinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
  • Z-2693-2017 · Jun 2, 2017 · Terminated
    Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions.
  • Z-2695-2017 · Jun 2, 2017 · Terminated
    Infinity 7, Implantable Pulse Generator, REF 6662, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
  • Z-2694-2017 · Jun 2, 2017 · Terminated
    Infinity 5, Implantable Pulse Generator, REF 6661, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I

Other STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR devices

All STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR devices →