HeartMate™ 107754Recalls & reported problems

Thoratec Corporation · Mobile Power Unit · Class 3

All models of HeartMate (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This HeartMate mobile power unit supplies portable power to a HeartMate heart pump. Its 460 reports mostly describe complete loss of power and unexpected disconnections. The reports include no deaths and 2 injuries, and fell to 68 in the last 12 months from 268 in the 12 months before. There are 2 open recalls, one for power cords that can disconnect and one for an electrical fault that can stop it turning on or shut it down.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
454
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 454 adverse event reports that list this exact device (DI 05415067038234), received between Jun 13, 2023 and Sep 30, 2026 — classified as 2 injuries, 452 malfunctions.

The most frequently reported specific device problem is “Complete Loss of Power” (202 reports).

2 FDA recall records name this device's identifier, 2 still open.

30 other recalls cover devices cleared under the same FDA submission (P060040, P160054); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 8 reportsJan 2024: 10 reportsFeb 2024: 5 reportsMar 2024: 7 reportsApr 2024: 11 reportsMay 2024: 7 reportsJun 2024: 13 reportsJul 2024: 11 reportsAug 2024: 10 reportsSep 2024: 14 reportsOct 2024: 16 reportsNov 2024: 13 reportsDec 2024: 19 reportsJan 2025: 39 reportsFeb 2025: 17 reportsMar 2025: 13 reportsApr 2025: 24 reportsMay 2025: 32 reportsJun 2025: 33 reportsJul 2025: 29 reportsAug 2025: 19 reportsSep 2025: 10 reportsOct 2025: 20 reportsNov 2025: 17 reportsDec 2025: 20 reportsJan 2026: 5 reportsFeb 2026: 6 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 2 reportsAug 2026: 4 reportsSep 2026: 5 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202639
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 23, 2025Class IOpen, Classified

    Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.

    Z-2291-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

  2. Mar 13, 2025Class IOpen, Classified

    Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.

    Z-1570-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Complete Loss of Power202
  • Use of Device Problem113
  • Electrical Power Problem80
  • Insufficient Device Problem Information66
  • Device Difficult to Setup or Prepare46
  • Disconnection42
  • Mechanical Problem38
  • Material Deformation15

Patient problems

  • No Clinical Signs, Symptoms or Conditions452
  • Electric Shock2
  • Increased Respiratory Rate1
  • Syncope/Fainting1
  • Weight Changes1

Recent adverse event reports

The 25 most recent of 454 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Mechanical Problem

    It was reported that the Mobile Power Unit (MPU) patient cord fell into water, and the clinic advised the device should not be used. A warranty replacement was requested.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948278 ↗
  • MalfunctionSep 25, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902002 ↗
  • MalfunctionSep 9, 2026Complete Loss of Power

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50696364 ↗
  • MalfunctionSep 2, 2026Complete Loss of Power

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50572531 ↗
  • MalfunctionSep 2, 2026Electrical Power Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50565853 ↗
  • MalfunctionAug 28, 2026Complete Loss of Power

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50503713 ↗
  • MalfunctionAug 25, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50438637 ↗
  • MalfunctionAug 24, 2026Disconnection; Electrical Power Problem; Complete Loss of Power

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50419759 ↗
  • MalfunctionAug 4, 2026Use of Device Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50133567 ↗
  • MalfunctionJul 28, 2026Use of Device Problem; Complete Loss of Power

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50060507 ↗
  • MalfunctionJul 27, 2026Insufficient Device Problem Information

    It was reported that the Mobile Power Unit (MPU) was malfunctioning. The MPU was alarming a continuous audio alarm without anything illuminated other than the green power icon. The patient was given a new MPU.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50000808 ↗
  • MalfunctionJun 11, 2026Failure to Power Up

    IT WAS REPORTED THAT THE MOBILE POWER UNIT (MPU) STOPPED WORKING AND WAS NOT TURNING ON AT ALL. THE MPU WOULD BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25477671 ↗
  • MalfunctionMay 10, 2026Use of Device Problem

    IT WAS REPORTED THAT THE MOBILE POWER UNIT (MPU) WAS CONSTANTLY, CONTENTLY, AND AUDIBLY ALARMING. THE BATTERIES WERE REPLACED, BUT THE ALARMS PERSISTED. THE MPU WAS REPLACED AND INTENDED TO BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25133439 ↗
  • MalfunctionApr 28, 2026Mechanical Problem

    IT WAS REPORTED THAT REPLACEMENT WAS REQUESTED DUE TO DEFECTIVE THREAD ON THE CONNECTION OF THE MOBILE POWER UNIT (MPU) CABLE. WHEN CONNECTING THE DEVICE TO THE SYSTEM CONTROLLER, THERE WAS AN ISSUE WITH THE SCREW THREADS, CAUSING RESISTANCE AND STIFFNESS WHEN TURNING THE SCREW. IN ADDITION, IT WAS NOTED THAT IT WAS DIFFICULT TO VISUALLY IDENTIFY ANY DIFFERENCES. NO PATIENT CONSEQUENCE WAS NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25021507 ↗
  • MalfunctionFeb 24, 2026Complete Loss of Power

    IT WAS REPORTED THAT FURTHER REVIEW OF THE LOG FILES CAPTURED NO EXTERNAL POWER ALARMS WHILE CONNECTED TO THE MOBILE POWER UNIT (MPU). THE MPU HAD A V-LOCK CONNECTOR ON THE AC POWER CORD. THE CAUSE OF THE NO EXTERNAL POWER ALARMS WAS NOT IDENTIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24434062 ↗
  • MalfunctionFeb 10, 2026Use of Device Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED A MOBILE POWER UNIT (MPU) ALARM. THE PATIENT REPORTED THAT THEY HAD IT PLUGGED INTO A SURGE PROTECTOR VERSUS A DIRECT WALL OUTLET. NO LOG FILES WERE AVAILABLE. THE PATIENT WAS TOLD TO WEAR BATTERIES AT NIGHT UNTIL MPU WAS EVALUATED AND REPAIRED. THE PATIENT WAS PROVIDED A LOANER MPU.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24306465 ↗
  • MalfunctionFeb 5, 2026Failure to Power Up

    IT WAS REPORTED THAT THE PATIENT'S MOBILE POWER UNIT (MPU) WOULD NOT POWER UP ON THREE DIFFERENT OUTLETS AT HOME AND AFTER CHECKING POWER CABLES. THE PATIENT WAS ABLE TO GET IT TO TURN ON AND BROUGHT IT IN FOR REPLACEMENT AS THE PATIENT WAS AFRAID TO CONTINUE TO USE IT. THE MPU WAS TESTED AND WAS WORKING FINE, PASSED SELF-TESTS AND LOOKED FINE BUT THE PATIENT DID NOT WANT IT BACK. THE MPU HAD A V-LOCK CONNECTOR ON THE AC POWER CORD. A NEW MPU WAS GIVEN TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24270781 ↗
  • MalfunctionFeb 3, 2026Device Difficult to Setup or Prepare; Complete Loss of Power

    IT WAS REPORTED THAT DURING ROUTINE LOG FILE ANALYSIS, THE LOG FILES CAPTURED A LOSS OF POWER TO THE MOBILE POWER UNIT ON (B)(6) 2026. THE SYSTEM CONTINUED TO OPERATE AT THE SET SPEED AND WAS SUPPORTED BY THE SYSTEM CONTROLLER¿S BACKUP BATTERY UNTIL EXTERNAL POWER WAS RESTORED. THE MPU DID HAVE A V-LOCK CONNECTOR ON THE AC POWER CORD. THE AC POWER CORD CAME LOOSE AT THE WALL OUTLET. THE MPU WAS NOT REMOVED FROM SERVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251713 ↗
  • MalfunctionFeb 2, 2026Complete Loss of Power

    IT WAS REPORTED THAT A REVIEW OF LOG FILES SHOWED POWER CABLE DISCONNECT AND NO EXTERNAL POWER WHILE CONNECTED TO MOBILE POWER UNIT (MPU) ON 15JAN2026, 17JAN2026 AND 21JAN2026. THIS WAS CAUSED BY POWER TO THE MPU BEING BRIEFLY INTERRUPTED. THE MPU DID NOT COME LOOSE AT THE WALL OUTLET OR FROM THE BACK OF THE UNIT. THE UNIT WAS EQUIPPED WITH A V-LOCK CONNECTOR ON THE AC POWER CORD. THERE WERE NOT ANY ENVIRONMENTAL CIRCUMSTANCES THAT WOULD HAVE AFFECTED AC POWER AT THE TIME OF THE EVENT. THE MPU REMAINED IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24241622 ↗
  • MalfunctionFeb 2, 2026Improper or Incorrect Procedure or Method; Complete Loss of Power

    IT WAS REPORTED THAT PATIENT ADMITTED ACCIDENTALLY DISCONNECTING MOBILE POWER UNIT (MPU) FROM OUTLET FORGETTING THEY WERE CONNECTED. THE PATIENT IMMEDIATELY SWITCHED TO BATTERY POWER BEFORE REALIZING THEY UNPLUGGED ACCIDENTLY. LOG FILES CAPTURED SEVERAL POWER LOSSES TO THE MOBILE POWER UNIT ON (B)(6) 2026. THE SYSTEM CONTINUED TO OPERATE AT THE SET SPEED AND WAS SUPPORTED BY BACKUP BATTERY UNTIL EXTERNAL POWER WAS RESTORED. THERE WERE NO OTHER UNUSUAL EVENTS RECORDED IN THE LOG FILE EVENT HISTORY. THE MPU WAS EQUIPPED WITH A V-LOCK CONNECTOR ON THE AC POWER CORD. THE AC POWER CORD COME LOOSE AT THE WALL OUTLET OR FROM THE BACK OF THE UNIT. THE MPU REMAINED IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24240727 ↗
  • MalfunctionJan 20, 2026Complete Loss of Power

    IT WAS REPORTED THAT THE PATIENT WAS AT HOME UTILIZING THE MOBILE POWER UNIT (MPU) AND THE UNIT TURNED OFF. IT WAS BROUGHT TO THE HOSPITAL FOR EVALUATION AND COULD NOT BE CORRECTED. A NEW POWER CORD WAS TRIED ALONG WITH SEVERAL OUTLETS. THE MPU HAD A V-LOCK CONNECTOR ON THE AC POWER CORD. THE PATIENT WAS GIVEN A NEW MPU. THE MPU HAD A V-LOCK CONNECTOR AND THE AC POWER CORD DID NOT COME LOOSE. THE OLD MPU WAS NOT RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24120922 ↗
  • MalfunctionJan 19, 2026Disconnection; Complete Loss of Power

    IT WAS REPORTED THAT THE PATIENT'S MOBILE POWER UNIT (MPU) HAD BEEN ALARMING THE PAST FEW NIGHTS. LOG FILES WERE SENT FOR REVIEW. THE LOG FILES CAPTURED NO EXTERNAL POWER ALARMS AND MULTIPLE OTHER ASSOCIATED ALARM TYPES ON 26DEC2025 AND 27DEC2025. THIS WAS CAUSED BY INTERRUPTION OF POWER TO THE MPU. THERE WERE RANDOM POWER CABLE DISCONNECTS ON 29DEC2025 WHILE CONNECTED TO THE MPU ASSOCIATED WITH THE BLACK LEAD. THE ALARM SILENCE BUTTON WAS REPEATEDLY ACTIVATED ON 31DEC2025 BETWEEN 11:07 AND 11:16 BUT NO ALARMS AT THESE TIMES. THESE OCCURRED WHILE THE PATIENT WAS CONNECTED TO THE MPU. IT WAS COMMUNICATED ON 09JAN2026 THAT THERE WERE NO ENVIRONMENTAL CIRCUMSTANCES THAT WOULD HAVE AFFECTED AC POWER AT THE TIME OF THE EVENT. THE NO EXTERNAL POWER ALARMS HAD RESOLVED AFTER THE PATIENT SWITCHED OUTLETS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24110100 ↗
  • MalfunctionJan 12, 2026Disconnection; Mechanical Problem; Electrical Power Problem; Complete Loss of Power

    IT WAS REPORTED THAT THE PATIENT HAD A NURSE VISIT WITH INCREASED SHORTNESS OF BREATH AND WEIGHT GAIN OF 25LBS IN 4 DAYS. THE LOG FILE WAS SENT FOR ANALYSIS. THE LOG FILE CAPTURED POWER CABLE DISCONNECT, LOW POWER HAZARD AND NO EXTERNAL POWER ALARMS WHILE CONNECTED TO THE MOBILE POWER UNIT (MPU) ON (B)(6) 2025 AT 23:22. THERE WERE NO PUMP INTERRUPTION DURING THE EVENTS AS THE SYSTEM CONTROLLER'S EMERGENCY BACKUP BATTERY (EBB) FUNCTIONED AS INTENDED. THE ALARMS RESOLVED AFTER THE MPU REGAINED POWER. BASED ON THE LOG FILE, THE MECHANICAL CIRCULATORY SYSTEM (MCS) EQUIPMENT OPERATED AS INTENDED ELECTRONICALLY AND MECHANICALLY. IT WAS LATER REPORTED ON 02JAN2026, THAT THE MPU HAD A V-LOCK CONNECTOR ON THE ALTERNATING CURRENT (AC) POWER CORD. THE PATIENT DID NOT RECALL IF THE AC POWER CORD CAME LOOSE FROM THE WALL OUTLET OR BACK OF THE MPU. THE PATIENT WAS UNSURE BUT BELIEVED THAT A BRIEF LOSS OF POWER TO THE HOUSE MAY HAVE AFFECTED THE AC POWER AT THE TIME. THE MPU WAS NOT EXCHANGED.

    Patient problems: Increased Respiratory Rate; No Clinical Signs, Symptoms or Conditions; Weight Changes

    MAUDE report 24041260 ↗
  • MalfunctionJan 7, 2026Disconnection; Mechanical Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED POWER CABLE DISCONNECTS WHEN CONNECTED TO THE MOBILE POWER UNIT (MPU). THE MPU WAS EXCHANGED. UPON CLOSER INSPECTION, THERE APPEARS TO BE BITE MARKS ON THE PATIENT CABLE AND AC POWER CORD. PUMP FUNCTIONED AS INTENDED. IT WAS SAID NO ADDITIONAL INFORMATION WOULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23999744 ↗
  • MalfunctionJan 2, 2026Device Difficult to Setup or Prepare; Complete Loss of Power

    IT WAS REPORTED THAT THERE WAS LOSS OF AC POWER WHILE CONNECTED TO MOBILE POWER UNIT (MPU) POWER AND LOG FILES WERE SUBMITTED FOR EVALUATION. IT APPEARED THAT THERE WAS NO EXTERNAL POWER EVENTS WERE RECORDED ON 14DEC2025, 15DEC2025, AND 16DEC2025 WHILE CONNECTED TO MPU POWER. THE PATIENT WAS USING THE LOCKING VERSION OF THE MPU AC POWER CORD. THE PATIENT WAS RE-EDUCATED THE PATIENT ON PROPER USE OF THE MPU. THE AC POWER CORD CAME LOOSE AT THE WALL OUTLET. THE MPU WAS NOT EXCHANGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23944871 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P160054 on Aug 23, 2017. Submitted by Abbott Medical as “HeartMate 3™ Left Ventricular Assist System”. The FDA review took 261 days. The FDA has since approved 132 supplements (changes to the device or its labeling), most recently on Oct 19, 2026. FDA record ↗

Approved through premarket approval (PMA) P060040 on Apr 21, 2008. Submitted by Abbott Medical as “Thoratec Heartmate Ii Left Ventricular Assist System”. The FDA review took 486 days. The FDA has since approved 125 supplements (changes to the device or its labeling), most recently on Oct 1, 2026. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Recalls under the same FDA submission

These recalls involve devices cleared under P060040, P160054. One submission often covers many models, so check the product and lot details in each record.

  • Z-2518-2026 · May 13, 2026 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US Software Version: N/A Product Description: The HeartMate 3" Left
  • Z-2520-2026 · May 13, 2026 · Open, Classified
    Brand Name: HeartMate Product Name: HM 11V Li-ion Backup Battery, ABT Model/Catalog Number: 106128 Software Version: N/A Product Description: The HeartMate 11 Volt Lithium Ion Stan
  • Z-2519-2026 · May 13, 2026 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N/A Product Description: The HeartMate 3" System Controller acts
  • Z-0506-2026 · Oct 9, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531LF
  • Z-0508-2026 · Oct 9, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate II Controller (standalone) Model/Catalog Number: 106017, HeartMate II Pocket Controller, US; 106762, HeartMate II Pocket Controller, U
  • Z-0505-2026 · Oct 9, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS;
  • Z-0507-2026 · Oct 9, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: HeartMate II LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II LVAS Kit, with Pocket Controller, Sealed, US; 106016, HeartMate II LVAS
  • Z-2290-2025 · Jun 23, 2025 · Open, Classified
    Brand Name: HeartMate Product Name: Mobile Power Unit, AC Power Cord Model/Catalog Number: 107760 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assis
  • Z-2583-2024 · Jun 25, 2024 · Open, Classified
    HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately: HeartMate 3 LVAS Implant Ki
  • Z-1921-2024 · May 8, 2024 · Open, Classified
    Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device
  • Z-1692-2024 · Mar 1, 2024 · Open, Classified
    Thoratec HeartMate 3, LVAS Implant Kit, REF:106524US, Sterile EO, RxOnly
  • Z-1693-2024 · Mar 1, 2024 · Open, Classified
    Thoratec HeartMate 3, LVAS Implant Kit, REF:106524INT, Sterile EO, RxOnly
  • Z-1300-2024 · Feb 19, 2024 · Open, Classified
    HeartMate II Left Ventricular Assist System (LVAS): 1. HeartMate II Sealed Outflow Graft with Bend Relief REF 103393 2. HeartMate II LVAS Implant Kit, US REF 106015 3. HeartMate
  • Z-1301-2024 · Feb 19, 2024 · Open, Classified
    HeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief, Investigational Device Exemption (IDE) 2. HeartMate 3 Sealed Outflow Graf
  • Z-0842-2024 · Jan 3, 2024 · Open, Classified
    1. HeartMate Touch Communication System, REF HMT1100 2. HeartMate Touch Communication System, Rental REF HMT1100-R 3. HeartMate Touch Communication System, HMT1150 The HeartMate T

…and 15 more.

Common questions about the HeartMate™ 107754

Has the HeartMate™ 107754 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2291-2025 (Class I), began on Jun 23, 2025: Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the HeartMate™ 107754?

The FDA's MAUDE database lists 454 adverse event reports for this device, received between Jun 13, 2023 and Sep 30, 2026. In those reports, the most common specific device problems are “Complete Loss of Power” (202), “Use of Device Problem” (113) and “Electrical Power Problem” (80); the most common patient problems are “Electric Shock” (2), “Increased Respiratory Rate” (1) and “Syncope/Fainting” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HeartMate™ 107754?

None of the reports are classified as deaths, 2 as injuries and 452 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HeartMate™ 107754 FDA approved?

It was approved through premarket approval (PMA) P060040 on Apr 21, 2008; its record lists 2 FDA submissions in total.

Who makes the HeartMate™ 107754, and is it still sold?

The FDA's device database lists Thoratec Corporation as the company responsible for it (brand HeartMate™, model 107754). Its record shows it as in commercial distribution.