Skip to content

HeartMateRecalls & reported problems across all models

Thoratec Corporation · 7 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

4 open recalls name at least one HeartMate model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
525
Deaths reported
1
Injuries reported
7
Recalls
4

We found 7 models of HeartMate by Thoratec Corporation in the FDA's device registry (GUDID). Together they are named in 525 adverse event reports in the FDA's MAUDE database, received between Aug 14, 2018 and Sep 30, 2026 — classified as 1 death, 7 injuries, 517 malfunctions.

The most frequently reported specific device problem is “Complete Loss of Power” (201 reports).

4 FDA recall records name at least one of these models by its device identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 3 reportsDec 2023: 9 reportsJan 2024: 13 reportsFeb 2024: 6 reportsMar 2024: 8 reportsApr 2024: 11 reportsMay 2024: 9 reportsJun 2024: 13 reportsJul 2024: 13 reportsAug 2024: 10 reportsSep 2024: 15 reportsOct 2024: 18 reportsNov 2024: 14 reportsDec 2024: 22 reportsJan 2025: 39 reportsFeb 2025: 17 reportsMar 2025: 13 reportsApr 2025: 25 reportsMay 2025: 33 reportsJun 2025: 35 reportsJul 2025: 31 reportsAug 2025: 21 reportsSep 2025: 10 reportsOct 2025: 20 reportsNov 2025: 18 reportsDec 2025: 21 reportsJan 2026: 5 reportsFeb 2026: 6 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports20242025202639
Last 36 months. The most recent month may be incomplete.

HeartMate recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. May 13, 2026Open, Classified

    Due to 11 volt Backup Battery failures.

    Z-2520-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jun 23, 2025Open, Classified

    Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.

    Z-2291-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

  3. Jun 23, 2025Open, Classified

    Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.

    Z-2290-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

  4. Mar 13, 2025Open, Classified

    Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.

    Z-1570-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported HeartMate problems

Device problems

  • Complete Loss of Power201
  • Use of Device Problem113
  • Electrical Power Problem81
  • Insufficient Device Problem Information66
  • Mechanical Problem60
  • Device Difficult to Setup or Prepare48
  • Disconnection43
  • Overheating of Device29

Patient problems

  • No Clinical Signs, Symptoms or Conditions509
  • No Consequences Or Impact To Patient7
  • No Known Impact Or Consequence To Patient5
  • Electric Shock2
  • No Patient Involvement2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Increased Respiratory Rate1
  • Syncope/Fainting1

HeartMate models

7 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent HeartMate adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Thoratec Corporation products

All Thoratec Corporation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page