NA 16000Recalls & reported problems
Thoratec Corporation · Charger · Class 3
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 05415067040756), received between Jan 31, 2024 and Sep 16, 2026 — classified as 4 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Melted” (13 reports).
19 other recalls cover devices cleared under the same FDA submission (P010032, P140009); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Melted13
- Smoking1
- Temperature Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
InjurySep 16, 2026Melted
It was reported that during troubleshooting with the patient for charging difficulties the Field Representative observed marks on the charger near the port that appear melted or deformed. A replacement charger has been provided to the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50796544 ↗MalfunctionFeb 19, 2026Melted
IT WAS REPORTED PATIENT WAS UNABLE TO CHARGE THE SYSTEM BECAUSE THE CHARGER HAD A WIRE MELTED. A REPLACEMENT CHARGER WAS SENT TO PATIENT TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24394962 ↗InjuryDec 2, 2025Melted; Melted
IT WAS REPORTED THAT THE CHARGING PORT OF THE PATIENT'S CHARGER MELTED. AS A RESULT, A REPLACEMENT CHARGER WAS SENT, AND PATIENT WAS ABLE TO RECHARGE THE IPG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23690113 ↗InjuryNov 4, 2025Melted; Melted
IT WAS REPORTED THAT THE PATIENT'S IPG CHARGER HAD A SMELL OF "BURNING PLASTIC" WHEN PLUGGED INTO THE WALL AND THE CHARGING ADAPTER HEATED UP AND MELTED. A REPLACEMENT IPG CHARGER WAS SENT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23468335 ↗MalfunctionSep 11, 2025Melted; Melted
IT WAS REPORTED THAT THE PATIENT'S GENERATOR CHARGER MELTED. AS A RESULT, A REPLACEMENT CHARGER MAY BE SENT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23034705 ↗MalfunctionAug 29, 2025Melted; Melted
IT WAS REPORTED THAT THE CHARGER GOT HOT WHILE CHARGING, AND THE CHARGER PORT MELTED WITH PART OF THE CHARGING CABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22929137 ↗MalfunctionJul 9, 2025Melted; Melted
IT WAS REPORTED THAT THE PATIENT'S CHARGER PORT MELTED. THE PATIENT WAS SHIPPED A REPLACEMENT CHARGER UNDER WARRANTY TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22460426 ↗MalfunctionMay 1, 2025Melted; Melted
IT WAS REPORTED THAT THE EXTERNAL DEVICE CHARGING PORT MELTED. THERE WAS NO REPORTED HARM TO THE PATIENT. THE DEVICE WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21943218 ↗InjuryApr 21, 2025Melted; Melted
IT WAS REPORTED THAT THE PATIENT'S CHARGER STARTED MELTING WHEN PLUGGED IN TO THE WALL TO CHARGE. A REPLACEMENT CHARGER WAS DELIVERED TO ADDRESS THE ISSUE. THE PATIENT WAS NOT HARMED IN ANY WAY BY THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21874884 ↗MalfunctionApr 17, 2025Melted; Melted
IT WAS REPORTED THAT THE PATIENT'S CHARGER AND CHARGING CABLE MELTED TOGETHER. THE PATIENT WAS PROVIDED A REPLACEMENT AND EXPERIENCED NO ADVERSE EFFECTS RELATED TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21853730 ↗MalfunctionMar 25, 2025Temperature Problem; Temperature Problem
IT WAS REPORTED PATIENT NOTICED THAT WHEN CHARGER WAS PLUGGED IN CHARGER GETS HOT QUICKLY AND SMELLS OF SOMETHING BURNING. THEY HAVE ALSO NOTED BLACK DOTS APPEARING ON THE BLUE LABEL. AS SUCH THE CHARGER WAS REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21687700 ↗MalfunctionJan 31, 2025Melted; Melted
IT WAS REPORTED THAT PATIENTS CHARGING CABLE MELTED. PATIENT EXPERIENCED NO HARM. PRODUCT WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21288927 ↗MalfunctionJan 31, 2025Melted; Melted
IT WAS REPORTED THAT WHEN ATTEMPTING TO CHARGE CHARGER, PATIENTS CHARGING CABLE AND A PORT ON PATIENT'S CHARGER MELTED. PATIENT EXPERIENCED NO HARM. PRODUCT WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21288924 ↗MalfunctionJan 31, 2025Melted; Melted
IT WAS REPORTED THAT WHEN ATTEMPTING TO CHARGE CHARGER, PATIENTS CHARGING CABLE AND A PORT ON PATIENT'S CHARGER MELTED. PATIENT EXPERIENCED NO HARM. PRODUCT WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21288921 ↗MalfunctionJan 31, 2024Smoking; Smoking
IT WAS REPORTED THE PATIENTS CHARGER EMITTED SMOKE DURING THE CHARGING SEQUENCE. NO ADVERSE CONSEQUENCES TO THE PATIENT. THE ISSUE WAS RESOLVED BY REPLACING THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18616007 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P140009 on Jun 12, 2015. Submitted by Abbott Medical as “Brio Neurostimulation System”. The FDA review took 403 days. The FDA has since approved 139 supplements (changes to the device or its labeling), most recently on Oct 30, 2026. FDA record ↗
Approved through premarket approval (PMA) P010032 on Nov 21, 2001. Submitted by Abbott Medical as “Genesis And Eon Family Neurostimulation (Ipg) Systems”. The FDA review took 176 days. The FDA has since approved 248 supplements (changes to the device or its labeling), most recently on Oct 30, 2026. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Recalls under the same FDA submission
These recalls involve devices cleared under P010032, P140009. One submission often covers many models, so check the product and lot details in each record.
- Z-0459-2026 · Oct 6, 2025 · Open, ClassifiedLiberta RC DBS IPG (Implantable Pulse Generator), Model Number 62400
- Z-0458-2026 · Oct 6, 2025 · Open, ClassifiedEterna SCS IPG (Implantable Pulse Generator), Model Number 32400
- Z-2509-2023 · Jul 18, 2023 · Open, ClassifiedProclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
- Z-2510-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2505-2023 · Jul 18, 2023 · Open, ClassifiedProclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
- Z-2511-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2508-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2507-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2506-2023 · Jul 18, 2023 · Open, ClassifiedProclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
- Z-2694-2017 · Jun 2, 2017 · TerminatedInfinity 5, Implantable Pulse Generator, REF 6661, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
- Z-2695-2017 · Jun 2, 2017 · TerminatedInfinity 7, Implantable Pulse Generator, REF 6662, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
- Z-2696-2017 · Jun 2, 2017 · TerminatedInfinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
- Z-2691-2017 · Jun 2, 2017 · TerminatedProclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-2692-2017 · Jun 2, 2017 · TerminatedProclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-2693-2017 · Jun 2, 2017 · TerminatedInfinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions.
…and 4 more.
Common questions about the NA 16000
Has the NA 16000 been recalled?
No FDA recall record in our data names this exact device identifier (DI 05415067040756), based on the FDA recall file dated Nov 27, 2026. However, 19 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the NA 16000?
The FDA's MAUDE database lists 15 adverse event reports for this device, received between Jan 31, 2024 and Sep 16, 2026. In those reports, the most common specific device problems are “Melted” (13), “Smoking” (1) and “Temperature Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 16000?
None of the reports are classified as deaths, 4 as injuries and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 16000 FDA approved?
It was approved through premarket approval (PMA) P010032 on Nov 21, 2001; its record lists 2 FDA submissions in total.
Is the NA 16000 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the NA 16000, and is it still sold?
The FDA's device database lists Thoratec Corporation as the company responsible for it (brand NA, model 16000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Stimulator, electrical, implanted, for essential tremor devices
- Vercise Genus™ DB-1216Boston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for essential tremor2 recalls23 deaths572 reports
- Liberta RC™ 62400Advanced Neuromodulation Systems, Inc. · Stimulator, electrical, implanted, for essential tremor2 recalls1 death104 reports
- SureTek™ DB-4600-CBoston Scientific Neuromodulation Corporation · Stimulator, electrical, implanted, for essential tremor1 recall137 reports
- Activa® SC 37603Medtronic, Inc. · Stimulator, electrical, implanted, for essential tremor1 recall107 reports
- Activa® SC 37602Medtronic, Inc. · Stimulator, electrical, implanted, for essential tremor1 recall85 reports
- SenSight™ B3301542Medtronic, Inc. · Stimulator, electrical, implanted, for essential tremor3 deaths76 reports
- N/A 3708660Medtronic, Inc. · Stimulator, electrical, implanted, for essential tremor43 reports
- SenSight™ B3301542MMedtronic, Inc. · Stimulator, electrical, implanted, for essential tremor3 deaths41 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page