TruMatch SD980.003Recalls & reported problems

Materialise NV · TRUMATCH Orthognathics - Basic Bimaxillary Surgical Kit · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05420060380037), received between Apr 26, 2019 and May 28, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Fracture” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K170272); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Fracture1

Patient problems

  • Difficulty Chewing1
  • No Code Available1
  • Obstruction/Occlusion1
  • Physical Asymmetry1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 28, 2025Fracture; Fracture

    SURGERY OF THIS CASE WENT WELL, MAXILLA WAS IN CORRECT POSITION RIGHT AFTER SURGERY. DURING FOLLOW UP APPOINTMENT SOME WEEKS LATER IT WAS DISCOVERED THAT THE MAXILLA MOVED POSTERIORLY AND UPWARDS WHICH LEAD TO A BAD OCCLUSION. IN THE SCAN IT WAS NOTICED THAT THE MAXILLA PLATE HAD MOVED POSTERIOR INTO THE SINUS AND THE MEDIAL LINK OF THE PLATE ON THE LEFT SIDE BROKE. ALSO ALMOST ALL SCREWS IN THE UPPER SKULL DISLOCATED FROM THE PLATE. REVISION SURGERY PLANNED FOR 10 JUNE 2025.

    Patient problems: Physical Asymmetry

    MAUDE report 22095134 ↗
  • InjuryOct 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    WHEN THE PATIENT WAS RELEASED FROM IMF, THE MANDIBLE AND THE OCCLUSION SAGGED BACK SO THE PATIENT ENDED UP WITH A BIGGER OVER JET THAN WE PLANNED, AND PREMATURE CONTACT CAUSED OPEN BITE.

    Patient problems: Obstruction/Occlusion

    MAUDE report 17929850 ↗
  • InjuryOct 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT HAS ANTERIOR OPEN BITE AND IS UNABLE TO CHEW. NEEDS REVISION SURGERY.

    Patient problems: Difficulty Chewing

    MAUDE report 15585852 ↗
  • InjuryApr 26, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT UNDERWENT A BI-SAGITTAL SPLIT OSTEOTOMY OF THE MANDIBLE. POST-OPERATIVE IT APPEARED THE OCCLUSION DID NOT MATCH THE PLANNED OUTCOME, WHILE DURING SURGERY NO OCCLUSION DISCREPANCIES WERE DETECTED. THE PATIENT NEEDS REVISION SURGERY, SURGERY NOT YET PLANNED.

    Patient problems: No Code Available

    MAUDE report 8554420 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170272. One submission often covers many models, so check the product and lot details in each record.

  • Z-1484-2020 · Feb 3, 2020 · Terminated
    TRUMATCH Orthognathic - Full Bimaxillary Surgical Kit, Catalog Number SD980.001
  • Z-1485-2020 · Feb 3, 2020 · Terminated
    TRUMATCH Orthognathics - Genioplasty Surgical Kit, Add-on, Catalog Number: SD980.013

Other Plate, Bone devices

All Plate, Bone devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.