TCM 4 series ETX base unit 391-880Recalls & reported problems
Radiometer Medical ApS · TCM 4 series ETX base unit · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05700693918808), received between Apr 2, 2025 and Dec 10, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Computer Software Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Computer Software Problem2
- Electrical /Electronic Property Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 10, 2025Computer Software Problem; Computer Software Problem
RADIOMETER COMPLAINT REFERENCE: (B)(4). ACCORDING TO THE COMPLAINT, THE TCM 4 SERIES ETX BASE UNIT (SERIAL NUMBER (B)(6) COMPLETES CALIBRATION SUCCESSFULLY AND DISPLAYS "READY." ELECTRODES ARE APPLIED TO THE PATIENT'S SKIN FOR MONITORING, AND MEASUREMENTS APPEAR ON THE SCREEN AS EXPECTED. HOWEVER, AFTER CONTINUOUS MONITORING FOR A PERIOD (AS SHORT AS A FEW MINUTES OR AS LONG AS 10-20 MINUTES), THE SCREEN SUDDENLY SHOWS "GENERAL ERROR," AND THE MACHINE STOPS MONITORING. PREVIOUSLY RECORDED DATA IS NOT LOST; IT IS SAVED IN THE PATIENT DATABASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23766424 ↗MalfunctionDec 10, 2025Computer Software Problem; Computer Software Problem
RADIOMETER COMPLAINT REFERENCE (B)(4). ACCORDING TO THE COMPLAINT, THE TCM 4 SERIES ETX BASE UNIT (SERIAL NUMBER (B)(6)) COMPLETES CALIBRATION SUCCESSFULLY AND DISPLAYS "READY." ELECTRODES ARE APPLIED TO THE PATIENT'S SKIN FOR MONITORING, AND MEASUREMENTS APPEAR ON THE SCREEN AS EXPECTED. HOWEVER, AFTER CONTINUOUS MONITORING FOR A PERIOD (AS SHORT AS A FEW MINUTES OR AS LONG AS 10-20 MINUTES), THE SCREEN SUDDENLY SHOWS "GENERAL ERROR," AND THE MACHINE STOPS MONITORING. PREVIOUSLY RECORDED DATA IS NOT LOST; IT IS SAVED IN THE PATIENT DATABASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23766343 ↗MalfunctionApr 2, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
ACCORDING TO THE COMPLAINT, ON (B)(6) 2025 DURING THE MONITORING OF THE PATIENT TCPO2 AND TCPCO2, THE TCM MONITOR (SERIAL NUMBER: (B)(6) DISPLAY PANEL SUDDENLY SHUT DOWN AND COULD NOT BE RESTARTED. THIS RESULTED IN THE INABILITY TO CONTINUE THE MONITORING OF THE PATIENT, LEADING TO AN ARTERIAL BLOOD SAMPLE DRAWN INSTEAD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21750030 ↗
Other Oximeter, ear devices
- MasimoSET LNCS® TC-I SpO2 Reusable Ear Sensor 1895Masimo Corporation · Oximeter, ear1 recall3 reports
- MasimoSET® M-LNCS™ TC-I SpO2 Reusable Ear Sensor 2503Masimo Corporation · Oximeter, ear1 recall1 report
- Portrait VSM MVC101Ge Medical Systems Information Technologies, Inc. · Oximeter, ear1 recall0 reports
- TruSignal SpO2 TS-SE-3GE Healthcare Finland Oy · Oximeter, ear1 recall0 reports
- V-Sign™ Sensor 2 VS-A/P/NSenTec AG · Oximeter, Ear1 recall0 reports
- MasimoSET® LNOP® TC-I SpO2 Reusable Ear Sensor 1794Masimo Corporation · Oximeter, ear1 recall0 reports
- Membrane Changer (single-use) / single unit MCSenTec AG · Oximeter, Ear1 recall0 reports
- TruSignal SpO2 TS-APGE Healthcare Finland Oy · Oximeter, ear1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page