ABL90 FLEX SP90 944-157Recalls & reported problems
Radiometer Medical ApS · ABL90 FLEX Solution pack · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 05700699441577), received between Dec 22, 2016 and Feb 3, 2023 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak4
- Mechanical Problem2
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- No Patient Involvement2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 3, 2023Fluid/Blood Leak; Fluid/Blood Leak
ACCORDING TO THE COMPLAINT, ON (B)(6) 2022, WHEN THE USER WAS USING THE ABL90 FLEX ANALYZER, IT WAS DISCOVERED THAT THE ABL90 FLEX SOLUTION PACK (LOT NUMBER: NA18) WAS FOUND TO BE LEAKING. NO HARM TO THE USER OR PATIENT WAS REPORTED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16289634 ↗MalfunctionOct 30, 2020Leak/Splash; Leak/Splash
ACCORDING TO THE COMPLAINT THE USER FOUND A LOT OF LIQUID LEAKAGE FROM THE SOLUTION PACK(SP) SEVERAL DAYS AFTER THE 90SP WAS INSTALLED. NOBODY TOUCHED THE LEAKED LIQUID DIRECTLY. THE SP CONTAINED CONTAMINATED BLOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10761727 ↗MalfunctionFeb 2, 2018Fluid/Blood Leak; Fluid/Blood Leak
ACCORDING TO THE COMPLAINT, ON (B)(4) 2017, A CUSTOMER GOT THE ERROR MESSAGES BELOW ON ABL90 DISPLAY, AND THEY COULD NOT MEASURE A SAMPLE. ANALYZER NEED MAINTENANCE. LEAK CURRENT DETECTED (#(B)(4)). THE CUSTOMER REMOVED THE SOLUTION PACK FROM ABL90 AND RE-INSTALLED IT BECAUSE THE SOLUTION PACK WAS FRESH. THEY INSTALLED A NEW 90SP WITHIN A DAY. THEN THEY FOUND PINK LIQUID LEAKAGE FROM THE BOTTOM OF THE SOLUTION PACK. IT WAS REPORTED THAT THE SOLUTION PACK HAD A SLIGHT LEAK AT THE CUSTOMER SITE, BUT IT WAS LEAKING A LOT AT RADIOMETER (B)(4) SITE. THE CUSTOMER DID NOT GET IN CONTACT WITH THE LEAKED LIQUID, BUT AN RADIOMETER (B)(4) EMPLOYEE GOT IN CONTACT WITH A BIT OF LEAKING LIQUID WHEN RECEIVING THE SOLUTION PACK.
Patient problems: No Patient Involvement
MAUDE report 7236658 ↗MalfunctionOct 11, 2017Mechanical Problem; Mechanical Problem
THE CUSTOMER EXPERIENCED LIQUID LEAKAGE FROM ABL90 SOLUTION PACK DURING ABL90 WORKING. A MEDICAL WORKER GOT IN TOUCH WITH THE LEAKED LIQUID, WHICH MAY CONTAIN BLOOD. HE GOT LIQUID ON HIS HAND, HIS SUITS, HIS SOCKS AND HIS BAG.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6939631 ↗MalfunctionJul 20, 2017Mechanical Problem; Mechanical Problem
A MEDICAL WORKER TOUCHED THE LEAKED LIQUID THAT MAY CONTAIN BLOOD. SHE WORE GLOVES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6727373 ↗MalfunctionFeb 14, 2017Fluid/Blood Leak; Fluid/Blood Leak
ACCORDING TO THE CUSTOMER THE SOLUTION PACK HAD LEAKED ALL OVER THE ANALYZER AND BENCH AND THE ANALYZER WAS REQUESTING A REPLACEMENT. THE SOLUTION PACK WAS TAKEN TO THE RADIOMETER OFFICE AND OPENED. IT COULD BE SEEN THAT THE WASTE POUCH HAD LEAKED OUT AND WAS EMPTY. MOST OF THE LIQUID WAS CONTAINED WITHIN THE OUTER CASE OF THE SOLUTION PACK BUT THERE WAS SOME ON THE OUTSIDE AS WELL. THE ANALYSER IS USED FOR ANIMAL BLOOD. THE OPERATOR WORE GLOVE AND THE INCIDENT DID NOT CAUSE HARM TO ANYONE.
Patient problems: No Patient Involvement
MAUDE report 6328681 ↗MalfunctionDec 22, 2016Fluid/Blood Leak; Fluid/Blood Leak
THE SOLUTION PACK GAVE A LEAKING PACK ERROR. WHEN THE USER REMOVED THE SOLUTION PACK FROM THE ANALYZER THERE WAS A POOL OF PINK LIQUID ON THE BOTTOM OF THE ANALYZER. WHEN THE USER LIFTED THE SOLUTION PACK, PINK LIQUID POURED OUT EVERYWHERE ON THE FLOOR. INSPECTION OF THE SOLUTION PACK SHOWED THAT THE PINK SOLUTION CAME FROM THE WASTE POUCH WHICH CONTAINS WASTE LIQUID INCLUDING BLOOD. NOBODY CAME INTO CONTACT WITH THE SOLUTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6197961 ↗
Other ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH devices
- GEM Premier 5000 00024019255Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph5 recalls5 deaths7 reports
- ABL800 FLEX 393-800Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls81 reports
- ABL90 FLEX Analyzer 393-090Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall112 reports
- i-STAT 03P88-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH105 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 300 TEST 00055430010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls3 deaths3 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 450 TEST 00055445010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls2 reports
- Stat Profile Prime PlusNova Biomedical Corporation · Electrode measurement, blood-gases (pco2, po2) and blood ph1 recall1 death50 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 150 TEST 00055415010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page