safePICO70 956-613Recalls & reported problems
Radiometer Medical ApS · safePICO 22Gx25mm Arterial blood sampler with 22Gx25mm needle, with vented TIPCAP, 1.5 mL, box of 100 pcs. · Class 2
All models of safePICO70 (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 05700699566133), received between Oct 14, 2022 and Nov 18, 2025 — classified as 1 injury, 9 malfunctions.
The most frequently reported specific device problem is “Corroded” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Corroded4
- Contamination /Decontamination Problem2
- Manufacturing, Packaging or Shipping Problem2
- Material Integrity Problem2
- Break1
- Component Missing1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Needle Stick/Puncture1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 18, 2025Corroded; Corroded
RADIOMETER REFERENCE NUMBER: (B)(4). ACCORDING TO COMPLAINT, THE CUSTOMER REPORTED THAT ON 04NOV2025, WHEN THE NURSE OPENED THE PACKAGING OF A SAFEPICO70 (LOT NUMBER CJ05) TO COLLECT AN ARTERIAL BLOOD SAMPLE, SHE FOUND THAT THE NEEDLE WAS RUSTY. SHE IMMEDIATELY REPORTED IT TO THE HEAD NURSE AND DID NOT USE THE SAMPLER. A NEW SAFEPICO WAS USED TO COLLECT THE SAMPLE FROM THE PATIENT. NO HARM WAS CAUSED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23581489 ↗InjuryAug 6, 2025Component Missing; Component Missing
RADIOMETER COMPLAINT REFERENCE: (B)(4). ACCORDING TO THE COMPLAINT, ON (B)(6) 2025 A CUSTOMER WAS OPENING THE PACKAGING OF THE SAFEPICO SAMPLER (LOT NUMBER: DF11) AND WAS PRICKED BY THE NEEDLE, AS IT WAS MISSING THE NEEDLE PROTECTIVE COVER. THE SAMPLER WAS AN UNUSED SAMPLER WHEN THE INCIDENT OCCURRED. THIS ISSUE IS COVERED BY RISK ID 74 WITH A SEVERITY OF 4 (PERMANENT/SERIOUS) ACCORDING TO THE (B)(4) - PRODUCT RISK ANALYSIS FOR SAMPLERS, (REV. 21). NO REPORTS OF DEATH.
Patient problems: Needle Stick/Puncture
MAUDE report 22724979 ↗MalfunctionFeb 21, 2025Break; Material Integrity Problem; Break; Material Integrity Problem
ACCORDING TO THE COMPLAINT, AFTER SUCCESSFULLY COLLECTING A BLOOD SAMPLE USING THE SAFEPICO70 ARTERIAL BLOOD SAMPLER, THE USER PREPARED TO REMOVE THE METAL NEEDLE AND THEN PLACE THE VENT CAP. DURING THE REMOVAL OF THE NEEDLE, THE LUER CONNECTOR BROKE, CAUSING BLOOD TO SPURT OUT. NEITHER THE OPERATOR NOR THE PATIENT WAS HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431454 ↗MalfunctionNov 7, 2024Corroded; Corroded
ACCORDING TO THE COMPLAINT, THE USER WAS PREPARING TO PERFORM AN ARTERIAL BLOOD SAMPLING ON THE PATIENT USING SAFEPICO70 SAMPLER (LOT NUMBER: BV06). UPON OPENING THE OUTER PACKAGING, AND INSPECTING THE SAMPLER, THE USER NOTICED THE NEEDLE HAD CORRODED WITH FOREIGN OBJECTS INSIDE THE NEEDLE HUB. THE USER IMMEDIATELY REPLACED IT WITH A NEW SAMPLER AND WAS NOT USED FOR TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20629824 ↗MalfunctionAug 22, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
ON 2024-08-01, DURING BLOOD GAS ANALYSIS USING THE SAFEPICO ARTERIAL BLOOD SAMPLER (LOT NO BI01), IT WAS FOUND THAT THE NEEDLE CUBE WAS MISSING, HENCE THE USER HAD NO NEEDLE STICK PROTECTION. IMMEDIATELY, THE DISPOSABLE ARTERIAL BLOOD SAMPLE COLLECTOR WAS REPLACED, AND THE INCIDENT DID NOT CAUSE ANY ADVERSE EFFECTS ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20027962 ↗MalfunctionAug 20, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
ON (B)(6) 2024, A NEWLY ADMITTED PATIENT UNDERWENT ARTERIAL BLOOD GAS ANALYSIS IN ACCORDANCE WITH MEDICAL ORDERS. HOWEVER, UPON OPENING THE PACKAGING OF THE SAFEPICO ARTERIAL BLOOD SAMPLER (LOT NO NY02), IT WAS FOUND THAT THE NEEDLE CUBE WAS MISSING, HENCE THE USER HAD NO NEEDLE STICK PROTECTION. IMMEDIATELY, THE DISPOSABLE ARTERIAL BLOOD SAMPLE COLLECTOR WAS REPLACED, AND THE INCIDENT DID NOT CAUSE ANY ADVERSE EFFECTS ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20001566 ↗MalfunctionJun 27, 2024Corroded; Corroded
ACCORDING TO THE COMPLAINT, ON (B)(6) 2024, AS THE NURSE WAS PREPARING TO PERFORM A BLOOD GAS ANALYSIS TEST ON A PATIENT, SHE CHECKED THE SAFEPICO (LOT NUMBER: NA01) AND CONFIRMED THAT IT WAS AIRTIGHT AND WITHIN ITS EXPIRATION DATE. HOWEVER, UPON OPENING THE PACKAGE AND RECHECKING THE DEVICE, THE NURSE FOUND THAT THE NEEDLE OF THE SAFEPICO WAS RUSTY. THE RUSTY NEEDLE WAS NOT USED FOR TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19622599 ↗MalfunctionMar 25, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
ACCORDING TO THE COMPLAINT ON (B)(6) 2024, THE NURSE OPENED THE BLISTER, AND FOUND THERE WAS A LINT IN THE SYRINGE (SAFEPICO). THE USER REPLACED WITH A NEW SAMPLER. THERE WAS NO HARM REPORTED. FROM THE DATA PROVIDED IT IS UNKNOWN IF THE LINT WAS COMING THROUGH THE BLISTER OR IF THE LINT WAS COMPLETELY INSIDE THE BLISTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18967055 ↗MalfunctionJan 3, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
ACCORDING TO THE COMPLAINT, ON 2023-11-10 THE USER OPENED THE BLISTER OF THE SAFEPICO (LOT NUMBER: NL50) AND FOUND A BLACK FOREIGN MATTER IN THE SYRINGE. THE USER REPLACED THE SAFEPICO, AND THERE WAS NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18434409 ↗MalfunctionOct 14, 2022Corroded; Material Integrity Problem; Corroded; Material Integrity Problem
ACCORDING TO THE COMPLAINT THE CUSTOMER DISCOVERED A RUSTY NEEDLE ON A SAFEPICO SAMPLER. THE CUSTOMER RETURNED THE SAMPLER BOX (100 PIECES) BACK TO RADIOMETER ON (B)(6) 2022. IT WAS CONFIRMED THAT THE RUSTY NEEDLE WAS DISCARDED AND NOT USED ON A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15605183 ↗
Other Tubes, vials, systems, serum separators, blood collection devices
- BD Vacutainer® SST™ 367986Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls2,711 reports
- BD Vacutainer® Buffered Sodium Citrate (9NC) Blood Collection Tubes 363083Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall1,744 reports
- BD Vacutainer® SST™ II Advance 367955Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection1,478 reports
- BD Vacutainer® Eclipse™ Blood Collection Needle 368607Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,220 reports
- BD Vacutainer® SST™ 367983Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls1,075 reports
- BD Vacutainer® K2 EDTA (K2E) 5.4mg Blood Collection Tubes 367856Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,010 reports
- BD Vacutainer® PST™ Gel and Lithium HeparinN 56 Units 367960Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall747 reports
- BD Vacutainer® SST™ II Advance 367957Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection767 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page