Skip to content

safePICO Aspirator 956-622Recalls & reported problems

Radiometer Medical ApS · safePICO Aspirator Arterial blood sampler. Aspirator only with vented TIPCAP, 1.5 mL, box of 100 pcs. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05700699566225), received between Aug 17, 2017 and Dec 23, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak1
  • High Test Results1
  • Insufficient Device Problem Information1
  • Manufacturing, Packaging or Shipping Problem1
  • Material Integrity Problem1
  • Mechanical Problem1
  • Protective Measures Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Exposure to Body Fluids1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 23, 2025Fluid/Blood Leak; Insufficient Device Problem Information; Fluid/Blood Leak; Insufficient Device Problem Information

    RADIOMETER COMPLAINT REFERENCE: (B)(4). THE USER TOOK THE ARTERIAL BLOOD SAMPLE PREVIOUSLY COLLECTED WITH A SAFEPICO ASPIRATOR (LOT NUMBER: RE11) BY HER COLLEAGUE. WHEN THE USER INSERTED THE SAMPLER INTO THE MACHINE, IT STARTED SPRAYING EVERYWHERE.

    Patient problems: Exposure to Body Fluids

    MAUDE report 23875535 ↗
  • MalfunctionAug 16, 2023Protective Measures Problem; Protective Measures Problem

    ACCORDING TO THE COMPLAINT, ON B)(6) 2023, WHEN THE OPERATOR TRIED TO VENT AIR FROM BLOOD SAMPLE IN THE SAFEPICO ASPIRATOR USING VTC, HE FELT THE PLUNGER WAS STIFF, BUT PUSHED THE PLUNGER ANYWAY. THEN BLOOD SPOUTED FROM THE TIP OF THE VTC, AND THE OPERATOR WAS EXPOSED TO THE BLOOD ON HIS FACE. THE OPERATOR UNDERWENT A BLOOD TEST, AND FORTUNATELY NO INFECTION HAS BEEN CONFIRMED AT THIS TIME. IT WAS CONFIRMED THAT THE OPERATOR WAS EXPOSED TO THE BLOOD ON ONLY HIS SKIN AND CLOTHES, NOT ON MUCOUS MEMBRANES SUCH AS EYES AND MOUTH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17555913 ↗
  • MalfunctionSep 30, 2022Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    ACCORDING TO THE COMPLAINT, ON (B)(6) 2022, WHEN THE OPERATOR TRIED TO VENTILATE AIR FROM THE SAMPLER, THE TIP CAP (VTC) GOT OUT OF THE SAMPLER, AND THE BLOOD INSIDE THE SAMPLER SPURTED UP TO THE CEILING. ACCORDING TO A RADIOMETER SERVICE ENGINEER WHO VISITED THE HOSPITAL, THE OPERATOR WAS TRAINED, AND THIS EVENT OCCURRED WHEN THE OPERATOR TRIED TO VENTILATE AIR AS USUAL (WITH THE SAME FORCE AS USUAL). NO ONE WAS EXPOSED TO THE BLOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15518596 ↗
  • MalfunctionMay 23, 2022Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    ACCORDING TO THE COMPLAINT WHEN THE CUSTOMER WAS TO ASPIRATE A BLOOD SAMPLE FROM A CATHETER, THEY EXPERIENCED THAT THERE WAS HARDLY ANY STOPPER FOR THE PLUNGER OF THE ASPIRATOR, WHICH CAUSED THE USER TO PULL OUT THE PLUNGER OF THE SAMPLER WHICH SPLASHED THEM WITH BLOOD ALL OVER. NO REPORTS OF DEATH OR SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14468831 ↗
  • MalfunctionAug 17, 2017High Test Results; High Test Results

    ACCORDING TO THE CUSTOMER, FALSE HIGH NA RESULTS WERE OBTAINED FROM THE ABL800 ANALYZER WHEN USING THE SAFEPICO ASPIRATOR, LOT YM-01. WHEN USING A DIFFERENT LOT OF THE SAMPLER, THE PROBLEM DIDN'T APPEAR AT ALL. NO PATIENTS WERE INVOLVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6800066 ↗

Other Tubes, vials, systems, serum separators, blood collection devices

All Tubes, vials, systems, serum separators, blood collection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page