safePICO 956-623Recalls & reported problems
Radiometer Medical ApS · safePICO 25Gx16mm (nsd) Arterial blood sampler with 25Gx16mm needle, with needle shield device (nsd) and vented TIPCAP, 1.5 mL, box of 100 pcs. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05700699566232), received between May 23, 2019 and Apr 21, 2026 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Fail-Safe Did Not Operate2
- Crack1
- Leak/Splash1
- Use of Device Problem1
Patient problems
- Needle Stick/Puncture3
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2026Crack; Leak/Splash
RADIOMETER REFERENCE NUMBER: (B)(4). ACCORDING TO THE COMPLAINT, THE HOSPITAL REPORTED THAT ON (B)(6) 2026 DURING A BLOOD GAS SAMPLE COLLECTION AS PART OF A ROUTINE CHECK-UP, THE SAMPLING SETUP WITH A RADIOMETER ARTERIAL BLOOD SAMPLER SAFEPICO (LOT NUMBER UNKNOWN) HAD BEEN PREPARED IN ADVANCE BY THE NIGHT NURSE. DURING THE PROCEDURE, BLOOD WAS OBSERVED ON THE NURSE'S FINGER, THOUGH ITS SOURCE COULD NOT BE IDENTIFIED. DURING THE PROCESS OF PURGING THE BLOOD GAS SYRINGE USING THE SPECIFIC STOPPER, A SPLASH OF BLOOD OCCURRED DUE TO A CRACK IN THE SYRINGE. NO BLOOD REACHED THE FACE OF THE NURSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24937364 ↗InjuryJul 12, 2023Fail-Safe Did Not Operate; Fail-Safe Did Not Operate
ON (B)(6) 2023, THE NEEDLE GUARD WAS BLOCKED ON A SAFEPICO 956-623 (LOT: NL51) WHEN THE USER (FEMALE, AGE 24) WAS HANDLING IT. THE USER TOOK THE BLOOD SAMPLE AND AFTERWARDS PRICKED HER FINGER WITH THE USED NEEDLE. THE USER WAS CHECKED AND CONTINUED WORKING AFTER THE INCIDENT.
Patient problems: Needle Stick/Puncture
MAUDE report 17304838 ↗InjuryNov 22, 2022Use of Device Problem; Use of Device Problem
ACCORDING TO THE COMPLAINT, THE NURSE WAS NOT ABLE TO ACTIVATE THE NEEDLE SHIELD DEVICE ON THE SAFEPICO AND WAS PRICKED BY THE NEEDLE AT THE END OF THE ARTERIAL SAMPLING.
Patient problems: Needle Stick/Puncture
MAUDE report 15842065 ↗InjuryOct 22, 2019Detachment of Device or Device Component; Fail-Safe Did Not Operate; Detachment of Device or Device Component; Fail-Safe Did Not Operate
A RESPIRATORY THERAPIST HAD A NEEDLE STICK FROM SAFEPICO BLOOD SAMPLER. AFTER NORMAL BLOOD SAMPLING PROCEDURE THE THERAPIST COULD NOT ACTIVATE THE SAFETY SHIELD OF THE SAMPLER, WHICH ULTIMATELY LED TO A FINGER STICK. THE THERAPIST STATED THAT SHE HAS EXPERIENCED SEVERAL BLOOD SAMPLERS WHERE THIS SAFETY MECHANISMS FAILS TO WORK. THE THERAPIST WAS FOLLOWING TESTED IN THE EMERGENCY DEPARTMENT AFTER THE NEEDLE STICK.
Patient problems: Needle Stick/Puncture; No Consequences Or Impact To Patient
MAUDE report 9217757 ↗MalfunctionMay 23, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
THE PHARMACY OF TOULON HOSPITAL COMPLAINED THAT IT WAS NOT POSSIBLE TO ACTIVATE THE NSD ON THREE CONSECUTIVE SYRINGES TAKEN FROM BATCH HD-53. THE USER WAS NOT ABLE TO ACTIVATE THE NSD WITH ONE HAND. IT FELT LIKE AS THE NEEDLE SHIELD WAS SEALED TO THE SYRINGE. THEY HAD TO USE BOTH HANDS TO MAKE IT MOVE. NO PERSONS WERE EXPOSED TO BLOOD AND NO IMPACT PATIENT MANAGEMENT. THE PERSON WHO DREW THE SAMPLE HAS BEEN CORRECTLY TRAINED TO USE OUR SAMPLER. ACCORDING TO INFORMATION FROM THE CUSTOMER THE AFFECTED SYRINGES WAS DISCARDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8635949 ↗
Other Tubes, vials, systems, serum separators, blood collection devices
- BD Vacutainer® SST™ 367986Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls2,711 reports
- BD Vacutainer® Buffered Sodium Citrate (9NC) Blood Collection Tubes 363083Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall1,744 reports
- BD Vacutainer® SST™ II Advance 367955Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection1,478 reports
- BD Vacutainer® Eclipse™ Blood Collection Needle 368607Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,220 reports
- BD Vacutainer® SST™ 367983Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection2 recalls1,075 reports
- BD Vacutainer® K2 EDTA (K2E) 5.4mg Blood Collection Tubes 367856Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1,010 reports
- BD Vacutainer® PST™ Gel and Lithium HeparinN 56 Units 367960Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection1 recall747 reports
- BD Vacutainer® SST™ II Advance 367957Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection767 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page