Skip to content

safePICO 956-623Recalls & reported problems

Radiometer Medical ApS · safePICO 25Gx16mm (nsd) Arterial blood sampler with 25Gx16mm needle, with needle shield device (nsd) and vented TIPCAP, 1.5 mL, box of 100 pcs. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05700699566232), received between May 23, 2019 and Apr 21, 2026 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Fail-Safe Did Not Operate2
  • Crack1
  • Leak/Splash1
  • Use of Device Problem1

Patient problems

  • Needle Stick/Puncture3
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 21, 2026Crack; Leak/Splash

    RADIOMETER REFERENCE NUMBER: (B)(4). ACCORDING TO THE COMPLAINT, THE HOSPITAL REPORTED THAT ON (B)(6) 2026 DURING A BLOOD GAS SAMPLE COLLECTION AS PART OF A ROUTINE CHECK-UP, THE SAMPLING SETUP WITH A RADIOMETER ARTERIAL BLOOD SAMPLER SAFEPICO (LOT NUMBER UNKNOWN) HAD BEEN PREPARED IN ADVANCE BY THE NIGHT NURSE. DURING THE PROCEDURE, BLOOD WAS OBSERVED ON THE NURSE'S FINGER, THOUGH ITS SOURCE COULD NOT BE IDENTIFIED. DURING THE PROCESS OF PURGING THE BLOOD GAS SYRINGE USING THE SPECIFIC STOPPER, A SPLASH OF BLOOD OCCURRED DUE TO A CRACK IN THE SYRINGE. NO BLOOD REACHED THE FACE OF THE NURSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24937364 ↗
  • InjuryJul 12, 2023Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    ON (B)(6) 2023, THE NEEDLE GUARD WAS BLOCKED ON A SAFEPICO 956-623 (LOT: NL51) WHEN THE USER (FEMALE, AGE 24) WAS HANDLING IT. THE USER TOOK THE BLOOD SAMPLE AND AFTERWARDS PRICKED HER FINGER WITH THE USED NEEDLE. THE USER WAS CHECKED AND CONTINUED WORKING AFTER THE INCIDENT.

    Patient problems: Needle Stick/Puncture

    MAUDE report 17304838 ↗
  • InjuryNov 22, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE COMPLAINT, THE NURSE WAS NOT ABLE TO ACTIVATE THE NEEDLE SHIELD DEVICE ON THE SAFEPICO AND WAS PRICKED BY THE NEEDLE AT THE END OF THE ARTERIAL SAMPLING.

    Patient problems: Needle Stick/Puncture

    MAUDE report 15842065 ↗
  • InjuryOct 22, 2019Detachment of Device or Device Component; Fail-Safe Did Not Operate; Detachment of Device or Device Component; Fail-Safe Did Not Operate

    A RESPIRATORY THERAPIST HAD A NEEDLE STICK FROM SAFEPICO BLOOD SAMPLER. AFTER NORMAL BLOOD SAMPLING PROCEDURE THE THERAPIST COULD NOT ACTIVATE THE SAFETY SHIELD OF THE SAMPLER, WHICH ULTIMATELY LED TO A FINGER STICK. THE THERAPIST STATED THAT SHE HAS EXPERIENCED SEVERAL BLOOD SAMPLERS WHERE THIS SAFETY MECHANISMS FAILS TO WORK. THE THERAPIST WAS FOLLOWING TESTED IN THE EMERGENCY DEPARTMENT AFTER THE NEEDLE STICK.

    Patient problems: Needle Stick/Puncture; No Consequences Or Impact To Patient

    MAUDE report 9217757 ↗
  • MalfunctionMay 23, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE PHARMACY OF TOULON HOSPITAL COMPLAINED THAT IT WAS NOT POSSIBLE TO ACTIVATE THE NSD ON THREE CONSECUTIVE SYRINGES TAKEN FROM BATCH HD-53. THE USER WAS NOT ABLE TO ACTIVATE THE NSD WITH ONE HAND. IT FELT LIKE AS THE NEEDLE SHIELD WAS SEALED TO THE SYRINGE. THEY HAD TO USE BOTH HANDS TO MAKE IT MOVE. NO PERSONS WERE EXPOSED TO BLOOD AND NO IMPACT PATIENT MANAGEMENT. THE PERSON WHO DREW THE SAMPLE HAS BEEN CORRECTLY TRAINED TO USE OUR SAMPLER. ACCORDING TO INFORMATION FROM THE CUSTOMER THE AFFECTED SYRINGES WAS DISCARDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8635949 ↗

Other Tubes, vials, systems, serum separators, blood collection devices

All Tubes, vials, systems, serum separators, blood collection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page