contact detach 100-905-00Recalls & reported problems
Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 05705244005716), received between Mar 6, 2018 and Apr 22, 2019 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Break1
- Defective Device1
- Infusion or Flow Problem1
- Material Deformation1
Patient problems
- No Known Impact Or Consequence To Patient4
- Device Embedded In Tissue or Plaque1
- Elevated Ketones/ Diabetic Ketoacidosis1
- Hyperglycemia1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryApr 22, 2019Infusion or Flow Problem; Infusion or Flow Problem
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. THE PATIENT'S DAUGHTER REPORTED AN OCCLUSION ISSUE AS PATIENT WAS TAKEN TO THE HOSPITAL. PATIENT WAS ADMITTED TO HOSPITAL WITH BLOOD GLUCOSE LEVEL OF 878 MG/DL AND DIAGNOSED WITH DIABETES KETOACIDOSIS. PATIENT HAD SYMPTOMS OF LETHARGY, INCOHERENCY AND VOMITING. THE PATIENT RECEIVED INSULIN VIA INJECTION, FLUIDS VIA INTRAVENOUS ROUTE AS CORRECTIVE TREATMENT AND THE INSERTION SITE WAS CHANGED. THE PATIENT WAS HOSPITALIZED FOR MORE THAN THREE DAYS. CURRENTLY, THE PATIENT WAS IN THE INTENSIVE CARE UNIT AND WAS BEING TREATED WITH AN INSULIN DRIP. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis
MAUDE report 8535340 ↗MalfunctionApr 8, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. THE PATIENT REPORTED THAT THE TUBING ON THE INFUSION SET DISCONNECTED FROM THE PLASTIC CLIP. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8490449 ↗MalfunctionApr 7, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. THE PATIENT'S MOTHER REPORTED THAT TWO OF THE INFUSION SETS BROKE AT THE NEEDLE. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8489932 ↗MalfunctionMar 29, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
(B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE QUICK RELEASE PORTION OF THE CONTACT DETACH FELL OFF PRIOR TO USE. PATIENT'S MOTHER STATED THAT IT LOOKED LIKE THERE WAS NO GLUE HOLDING THE QUICK RELEASE TO THE TUBING. UPON INVESTIGATION, IT WAS FOUND THAT THE TUBING WAS DETACHED FROM THE TUBING CONNECTOR DUE TO MISSING GLUE. FURTHERMORE, A DROP OF GLUE ON TUBING CONNECTOR WAS FOUND. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8463103 ↗MalfunctionJan 25, 2019Material Deformation; Material Deformation
(B)(4). EVENT OCCURRED IN (B)(6). THE PATIENT REPORTED CRACKED/BROKEN TUBING. THE HOSE HAD GONE OFF AT THE BRACKET AS WELL. IN THE COMPLAINT INVESTIGATION, IT WAS FOUND THAT THE TUBING WAS DETACHED FROM THE TUBING CONNECTOR DUE TO MISSING GLUE. FURTHERMORE, THERE WAS EXCESS OF GLUE PRESENT WHERE THE TUBING WAS JOINED TO THE TUBING CONNECTOR. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8276824 ↗MalfunctionJan 7, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN (B)(6). THE PATIENT STATED THAT TUBING WAS BROKEN. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Hyperglycemia
MAUDE report 8221046 ↗InjuryMar 6, 2018Break; Defective Device
PROBLEM: THREE DIFFERENT NEEDLE PORTIONS OF CONTACT DETACH BROKE OFF IN OUR DAUGHTER. WHAT HAPPENED: OUR TYPE 1 DIABETIC DAUGHTER USES AN ANIMAS INSULIN PUMP, WITH CONTACT DETACH, AND A DEXCOM CGM. DURING THE FIRST INSTANCE, WE NOTICED BLOOD SUGAR LEVELS GREATER THAN 400 SUSTAINED FOR MULTIPLE HOURS. AFTER MULTIPLE ATTEMPTS TO ADMINISTER INSULIN AND NOT SEEING ANY CHANGE IN BLOOD SUGAR LEVELS, WE ENDED UP REPLACING AND REINSERTING ALL OF THE PUMP STUFF. IT WAS THEN THAT WE DISCOVERED THAT THE NEEDLE THAT WAS INSERTED IN HER TO DELIVER INSULIN HAD BROKEN OFF RESULTING IN NO OR MINIMAL INSULIN DELIVERY. WE CONTACTED HER ENDOCRINOLOGIST, AND ANIMAS. SINCE ANIMAS IS CLOSING, THE REPRESENTATIVE THAT WE SPOKE SAID THEY WERE FROM MEDTRONIC. WE ENDED UP NEEDING A SURGICAL CONSULT, RESULTING IN THE DECISION TO LEAVE THE EMBEDDED NEEDLE UNTIL IT CAUSES INFECTION. THE SECOND INSTANCE OF THE NEEDLE BREAKING HAPPENED DURING A RESET OF HER CGM. SHE PRESENTED WITH BLOOD SUGAR LEVELS OF OVER 550 FOR OVER 15 HOURS, FROM AFTER DINNER TO WHEN WE WOKE HER UP IN THE MORNING. WE AGAIN HAD MULTIPLE ATTEMPTS TO BRING HER BLOOD SUGAR DOWN, BUT WITHOUT THE AID OF A CGM WE ASSUMED THAT THE AMOUNT OF INSULIN
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7320606 ↗
Other Set, Administration, Intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page