Skip to content

Quick-set® MMT-392Recalls & reported problems

Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05705244006294), received between Feb 3, 2019 and Jun 3, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Appropriate Device Problem Term/Code Not Available1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 3, 2024Leak/Splash; Leak/Splash

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN DENMARK. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKAGE FROM SITE/PIPE/KANYLA ON (B)(6) 2024. INFUSION SET HAS BEEN USED FOR LESS THAN A DAY. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19447427 ↗
  • MalfunctionJul 13, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. ON (B)(6) 2023, THE PATIENT'S MOTHER REPORTED THAT, A FEW DAYS AGO, THERE WAS A SMALL CRACK AND SUBSEQUENT LEAKAGE IN ONE OF THE NEEDLES OF THE INFUSION SET. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17314584 ↗
  • MalfunctionMar 31, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. THE PATIENT REPORTED THAT WHILE SLEEPING, THE TUBING SNAPPED OFF AND WAS NO LONGER ATTACHED. HER BLOOD GLUCOSE LEVEL WAS 467 MG/DL, WHICH SHE TREATED WITH A SHOT. SHE CHANGED THE TUBING AND EVERYTHING WAS FINE. SHE ALSO HAD ISSUES FROM TIME TO TIME WITH GETTING THE QUICK RELEASE BACK ON AFTER TAKING A SHOWER. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8467977 ↗
  • MalfunctionFeb 3, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    (B)(4). EVENT OCCURRED IN (B)(6). THE PATIENT REPORTED THAT TUBING GOT DETACHED AT THE QUICK RELEASE. THE PATIENT OBSERVED THAT INSULIN CAME OUT OF THE TUBING, 5 HOURS AFTER THE SET WAS CHANGED AND THIS ISSUE OCCURRED WITH ONE SET. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8300431 ↗

Other Set, Administration, Intravascular devices

All Set, Administration, Intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page