Skip to content

Quick-set® MMT-391Recalls & reported problems

Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05705244006638), received between Jan 19, 2019 and Jul 5, 2019 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Material Twisted/Bent1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • Elevated Ketones/ Diabetic Ketoacidosis1
  • Hyperglycemia1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 5, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    (B)(4). EVENT OCCURRED IN (B)(6). IT WAS REPORTED THAT THE INFUSION SET'S TUBING HAD DETACHED AT THE QUICK RELEASE AND THIS ISSUE OCCURRED WITH TEN SETS. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8763170 ↗
  • InjuryApr 25, 2019Material Twisted/Bent; Material Twisted/Bent

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED BENT CANNULA ISSUE. HIS BLOOD GLUCOSE LEVEL WAS 470 MG/DL AND HE SUFFERED FROM VOMITING AND DEHYDRATION. THE EMERGENCY MEDICAL SERVICE WAS CALLED, AND THE PATIENT WAS ADMITTED TO THE HOSPITAL. HE WAS GIVEN INSULIN DRIP AS CORRECTIVE TREATMENT. THE PATIENT WAS HOSPITALIZED FOR TWO AND A HALF DAYS. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 8548023 ↗
  • MalfunctionApr 1, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    (B)(4). EVENT OCCURRED IN (B)(6). IT WAS REPORTED THAT THE TUBING CAME OUT OF THE CANNULA. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8468278 ↗
  • MalfunctionJan 19, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN (B)(6). THE PATIENT REPORTED THAT WHEN SHE WOKE UP IN THE MORNING, SHE HAD HIGH BLOOD GLUCOSE LEVEL (27 MMOL/L) AS SHE DID NOT KNOW WHETHER SHE HAD INSULIN AT NIGHT OR NOT. REPORTEDLY, THE TUBING CAME APART FROM WHERE IT WAS ATTACHED TO HER SITE. THE PATIENT TREATED HIGH BLOOD GLUCOSE LEVEL WITH HER PUMP AND IT WAS STILL COMING DOWN. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Hyperglycemia

    MAUDE report 8263039 ↗

Other Set, Administration, Intravascular devices

All Set, Administration, Intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page