Skip to content

Mio® MMT-965600Recalls & reported problems

Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

All models of Mio (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05705244007819), received between Aug 26, 2019 and Nov 14, 2024 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent2
  • Detachment of Device or Device Component1
  • Improper or Incorrect Procedure or Method1
  • Infusion or Flow Problem1

Patient problems

  • Elevated Ketones/ Diabetic Ketoacidosis3
  • Hyperglycemia1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 14, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    REFERENCE NUMBER: (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED ON 13-OCT DUE TO HIGH BLOOD GLUCISE LEVEL AND REPORTED ISSUE WITH THE PUMP. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 20682539 ↗
  • MalfunctionJul 6, 2022Infusion or Flow Problem; Infusion or Flow Problem

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA ON (B)(6) 2022, THE PATIENT'S MOTHER REPORTED THAT HER DAUGHTER WAS NOT ABLE TO USE THE PUMP FOR THREE DAYS DUE TO INSULIN FLOW BLOCKED ALARM. THIS ISSUE STARTED THREE DAYS AGO, SINCE THEN THE PATIENT HAS BEEN USING NEEDLES TO MANAGE HER BLOOD GLUCOSE LEVELS. MOREOVER, SHE CHANGED THE INFUSION SET AND RESERVOIR FOUR TIMES, BUT STILL FACED THE SAME ISSUE. REPORTEDLY, THE PATIENT WAS WAITING TO SEE HER HEALTHCARE PROFESSIONAL. SHE RECEIVED THE ALARMS DURING BASAL. AT THE TIME OF THE REPORT, HER BLOOD GLUCOSE LEVEL WAS 29 MMOL/L, AND THE PATIENT WAS VOMITING. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Hyperglycemia

    MAUDE report 14927367 ↗
  • InjuryMay 16, 2021Material Twisted/Bent; Material Twisted/Bent

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN (B)(6). ON (B)(6) 2021, IT WAS REPORTED THAT THE PATIENT EXPERIENCED HIGH BLOOD GLUCOSE LEVEL OF 30 MMOL/L. THEREFORE, IN THE MID-DAY OF (B)(6) 2021, THE PATIENT WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS. AT THE HOSPITAL, THE PATIENT REMOVED THE INFUSION SET AND NOTICED THAT CANNULA WAS BENT ON THE FIRST DAY OF USE. REPORTEDLY, THE SITE LOCATION WAS PATIENT'S ABDOMEN AND THE INFUSION SET HAD BEEN USED OVERNIGHT. MOREOVER, THE PATIENT HAD OCCLUSION AND THE PUMP DID NOT ALARM WITH INSULIN FLOW BLOCK. ON (B)(6) 2021, THE PATIENT WAS RELEASED FROM THE HOSPITAL. CURRENTLY, THE PATIENT'S BLOOD GLUCOSE LEVEL WAS 7.2 MMOL/L. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 11830261 ↗
  • MalfunctionOct 25, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN (B)(6). IT WAS REPORTED THAT THE PATIENT'S INFUSION SET'S TUBING HAD DETACHED AT THE QUICK RELEASE, WHILE THE PATIENT WAS UNCOILING THE TUBING TO INSERT. REPORTEDLY, AS HE UNCOILED THE INFUSION SET, IT WAS NOTICED THAT IT WAS NOT CONNECTED TO THE CANNULA AND IT WAS ALL STICKY. THE PATIENT'S SERUM GLUCOSE LEVEL WAS 13.9. THE INFUSIONS WERE STORED WERE STORED IN A CLOSET IN THE BEDROOM, AT A COOL DRY PLACE, AT ROOM TEMPERATURE. FURTHER, THERE WAS NO STRESS OR PULL ON THE TUBING AND THE PUMP WAS NOT DROPPED WITH THE SET CONNECTED TO THE PATIENT'S BODY. UPON INVESTIGATION OF RETURNED USED DEVICE (1 SET), IT WAS FOUND THAT THE TUBING WAS DETACHED FROM THE TUBING CONNECTOR (MISSING GLUE). NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9236713 ↗
  • InjuryAug 26, 2019Material Twisted/Bent; Material Twisted/Bent

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN (B)(6). IT WAS REPORTED THAT THE PATIENT EXPERIENCED NAUSEA, VOMITING, ABDOMINAL PAIN, HAD DIFFICULTY IN BREATHING AND WAS TESTED POSITIVE FOR KETONES. THE PATIENT ALSO HAD HIGH BLOOD GLUCOSE LEVEL REPORTED TO BE 28 MMOL/L AT THE TIME OF INCIDENT. ON (B)(6) 2019, SHE WAS SUBSEQUENTLY ADMITTED TO THE HOSPITAL. WHEN HER INFUSION SET WAS REMOVED, IT WAS FOUND THAT THERE WAS A LITTLE CURVE IN THE CANNULA. THE HOSPITAL TOLD THE PATIENT THAT THERE WERE MORE ISSUES THAN THE SLIGHTLY BENT CANNULA. ON (B)(6) 2019, SHE HAD COMPUTED TOMOGRAPHY SCAN AND WAS STILL IN THE HOSPITAL. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 8929257 ↗

Other Set, Administration, Intravascular devices

All Set, Administration, Intravascular devices →

Sources and updates