Silhouette® Paradigm® MMT-381600Recalls & reported problems
Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05705244008564), received between Jan 11, 2019 and Jan 30, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- No Apparent Adverse Event1
Patient problems
- Coma1
- Hypoglycemia1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 30, 2019No Apparent Adverse Event; No Apparent Adverse Event
UNOMEDICAL REFERENCE NUMBER:(B)(4). A MALE DIABETIC PATIENT WAS HOSPITALIZED ON SATURDAY (B)(6) 2018 AT 3:30 AM DUE TO HYPOGLYCEMIA AND DIABETIC COMA. LOW BLOOD GLUCOSE LEVEL OF 1.7 MMOL/L AT TIME OF EVENT. THE PATIENT REPORTED THAT HE WAS IN THE HOSPITAL FOR FEW HOURS AND WAS DISCHARGED ON THE SAME DAY. IT WAS STATED THAT THE PATIENT DID NOT HAVE A SENSOR AND HE FELL IN TO DIABETIC COMA IN THE NIGHT AND WENT TO THE HOSPITAL. THE PATIENT REPORTED THAT BEFORE GOING TO SLEEP HE NOTICED A BUMP ON HIS SKIN AND HE THOUGHT THAT THE BUMP WAS DUE TO INSULIN STACKING. IT WAS REPORTED THAT THE PATIENT CHANGED THE INFUSION SET AND SITE OF INSERTION AND GAVE HIMSELF A CORRECTION. THE PATIENT REPORTED THAT HE WENT TO SLEEP AND CHECKED HIS BLOOD GLUCOSE JUST BEFORE GOING TO BED. IT WAS STATED THAT HIS WIFE NOTICED THAT HE WAS SWEATY AND GAVE HIM ORANGE JUICE AND SUGAR. IT WAS REPORTED THAT THE AMBULANCE WAS CALLED OUT AND HE WAS GIVEN GLUCAGON INJECTIONS TO BRING HIS BLOOD GLUCOSE BACK TO NORMAL. IT WAS REPORTED THAT THE CAUSE OF HOSPITALIZATION PER HEALTH CARE PROFESSIONAL WAS LOW BLOOD GLUCOSE. IT WAS REPORTED THAT THE PATIENT WAS NOT WEARING THE INSULIN PUMP AT THE TIME OF HOSPITALIZATION AND
Patient problems: Coma; Hypoglycemia
MAUDE report 8288254 ↗MalfunctionJan 11, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
(B)(4). EVENT OCCURRED IN (B)(6). THE PATIENT REPORTED THAT THE INFUSION SET TUBING GOT DETACHED WHILE SHOPPING. REPORTEDLY, THE LOCATION OF DETACHMENT WAS AT THE CONNECTOR. THE SITE LOCATION WAS ABDOMEN AND THE PUMP WAS WORN ON BELT. THE INFUSIONS WERE NOT STORED, OR USED, IN A PLACE WHERE THEY MIGHT HAVE BEEN EXPOSED TO EXTREME TEMPERATURES AND HUMIDITY. THE INFUSION HAD BEEN IN USE FOR FOUR DAYS. ALSO, THERE WAS NO STRESS OR PULL ON THE TUBING AND THE PUMP WASN'T DROPPED WITH THE SET CONNECTED TO THE BODY. THE PATIENT HAD FACED THE SAME ISSUE EARLIER AS WELL, TWO TIMES BUT REALIZED IT TWO HOURS LATER. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8238776 ↗
Other Set, Administration, Intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page