Silhouette® Paradigm® MMT-381Recalls & reported problems
Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2
All models of Silhouette Paradigm (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Unomedical Silhouette Paradigm MMT-381 is a single-use infusion set placed under the skin to deliver insulin or similar medicine. The 43 reports mostly describe the set coming apart, the soft cannula bending or kinking after insertion, and leaks or partial blockage that interrupted flow; several patients reported high blood glucose, pain at the insertion site, or ketones. One report describes a death and 8 describe injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 45
- Deaths reported
- 1
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 05705244008588), received between Dec 18, 2017 and Sep 26, 2026 — classified as 1 death, 10 injuries, 34 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component18
- Material Twisted/Bent7
- Leak/Splash6
- Adverse Event Without Identified Device or Use Problem5
- Partial Blockage4
- Infusion or Flow Problem2
- Complete Blockage1
- Device Handling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
- Hyperglycemia10
- No Known Impact Or Consequence To Patient7
- Elevated Ketones/ Diabetic Ketoacidosis5
- Pain4
- Discomfort2
- Hypoglycemia2
- Coma1
Recent adverse event reports
The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.
InjurySep 26, 2026Device Handling Problem
Patient problems: Hyperglycemia
MAUDE report 50907573 ↗InjurySep 5, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Hyperglycemia
MAUDE report 50629994 ↗InjuryAug 17, 2026Adverse Event Without Identified Device or Use Problem
It was reported that patient experienced high blood glucose event on (b)(6) 2026 and was treated with insulin pen.
Patient problems: Hyperglycemia
MAUDE report 50269860 ↗InjuryAug 10, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient had high blood glucose and treated with insulin pump on (b)(6)2026
Patient problems: Hyperglycemia
MAUDE report 50189858 ↗DeathJun 13, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED DUE TO LOW BLOOD GLUCOSE AND WAS LATER DECEASED DUE TO HEART CONDITION AND HYPOTENSION ADRENAL INSUFFICIENCY ON (B)(6) 2020.
Patient problems: Hypoglycemia
MAUDE report 25495546 ↗InjuryApr 13, 2026Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2026, IT WAS REPORTED THAT THE PATIENT FACED HIGH BLOOD GLUCOSE WHICH GOT TREATED WITH BOLUS. THE BLOOD GLUCOSE WAS BEEN HIGH FOR 4 HOURS.
Patient problems: Hyperglycemia
MAUDE report 24855941 ↗MalfunctionMar 30, 2026Material Twisted/Bent
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN SWITZERLAND. IT WAS REPORTED THAT THE PATIENT FACED CANNULA ISSUE ON (B)(6) 2026. THE INFUSION SET CANNULA WAS BENT AND THE PATIENT EXPERIENCED PAIN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Discomfort
MAUDE report 24731211 ↗MalfunctionMar 30, 2026Material Twisted/Bent
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN SWITZERLAND. IT WAS REPORTED THAT THE PATIENT FACED CANNULA ISSUE ON (B)(6) 2026. THE INFUSION SET CANNULA WAS BENT AND THE PATIENT EXPERIENCED PAIN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Discomfort
MAUDE report 24730266 ↗MalfunctionMar 30, 2026Partial Blockage
REFERENCE NUMBER (B)(4) EVENT COUNTRY IN SWITZERLAND. IT WAS REPORTED THAT THE THE PATIENT EXPERIENCED PAIN AT THE INSERTION SITE BECAUSE THE INFUSION SET CANNULA BECAME BENT WITHIN A FEW MINUTES OF USE ON (B)(6) 2026. NO FURTHER INFORMATION AVAILABLE
Patient problems: Pain
MAUDE report 24728111 ↗MalfunctionMar 30, 2026Partial Blockage
REFERENCE NUMBER (B)(4) EVENT COUNTRY IN SWITZERLAND. IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE INSERTION SITE BECAUSE THE INFUSION SET CANNULA BECAME BENT WITHIN A FEW MINUTES OF USE ON (B)(6) 2026. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Pain
MAUDE report 24726215 ↗MalfunctionMar 30, 2026Partial Blockage
REFERENCE NUMBER (B)(4). EVENT COUNTRY IN THE SWITZERLAND. IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE INSERTION SITE BECAUSE THE INFUSION SET CANNULA BECAME BENT WITHIN A FEW MINUTES OF USE ON (B)(6) 2026. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Pain
MAUDE report 24725958 ↗MalfunctionMar 30, 2026Partial Blockage
REFERENCE NUMBER (B)(4). EVENT COUNTRY IN THE SWITZERLAND. IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE INSERTION SITE BECAUSE THE INFUSION SET CANNULA BECAME BENT WITHIN A FEW MINUTES OF USE ON (B)(6) 2026. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Pain
MAUDE report 24725938 ↗MalfunctionMar 26, 2026Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED LEAKAGE AT THE CANNULA SITE DURING PHYSICAL ACTIVITY ON (B)(6) 2023. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24701339 ↗MalfunctionMar 26, 2026Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CZECH REPUBLIC. IT WAS REPORTED THAT THE PATIENT EXPERIENCED LEAKAGE AT THE SITE ON (B)(6) 2023, AFTER THE INFUSION SET HAD BEEN IN USE FOR 3 DAYS. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24701337 ↗MalfunctionFeb 10, 2026Complete Blockage; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN ITALY. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2026. THE BLOCKAGE WAS IN THE TUBING. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24301311 ↗MalfunctionSep 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN COLOMBIA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET TUBING DETACHMENT EVENT ON 13-AUG-2025. THE SITE OF DETACHMENT WAS CLOSE TO SITE LOCATION. THE INFUSION SET WAS IN USE FOR ONE DAY NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23034882 ↗MalfunctionSep 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN COLOMBIA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET TUBING DETACHMENT EVENT ON 13-AUG-2025. THE SITE OF DETACHMENT WAS CLOSE TO SITE LOCATION. THE INFUSION SET WAS IN USE FOR ONE DAY NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23023838 ↗MalfunctionApr 18, 2025Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN COLOMBIA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED THAT INFUSION SET WAS LEAKING IN THE CANNULA SIDE AT QUICK RELEASE SITE ON (B)(6) 2025. THE INFUSION SET WAS IN USE FOR ONE DAY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21864068 ↗MalfunctionApr 17, 2025Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN COLOMBIA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED THAT INFUSION SET WAS LEAKING IN THE CANNULA SIDE AT QUICK RELEASE SITE ON (B)(6) 2025. THE INFUSION SET WAS IN USE FOR ONE DAY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21855535 ↗MalfunctionMar 7, 2025Material Twisted/Bent; Material Twisted/Bent
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED AN INFUSION SET CANNULA KINKED EVENTS ON (B)(6) 2025. COMPANY DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21544749 ↗MalfunctionDec 6, 2024Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED OCCLUSION AT THE TUBING EVENT ON (B)(6) 2024. PATIENT CHANGED SUPPLIES AS NECESSARY AND RESUMED INSULIN THERAPY. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20862391 ↗MalfunctionNov 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN COLOMBIA. IT WAS REPORTED THAT THE PATIENT FACED INFUSION SET CANNULA TUBING WAS DETATCHED CLOSE TO THE CANNULA EVENT ON (B)(6) 2024. SITE OF INSERTION WAS ABDOMEN. INFUSION SET WAS IN USE FOR 12 HOURS. NO FURTHER INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20703595 ↗MalfunctionOct 9, 2024Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN COLOMBIA. IT WAS REPORTED THAT THE PATIENT FACED INFUSION SET LEAKING AT THE TUBING EVENT ON (B)(6) 2024. THE INFUSION SET HAS BEEN USED FOR 2 DAYS. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20407953 ↗MalfunctionOct 9, 2024Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN COLOMBIA. IT WAS REPORTED THAT THE PATIENT FACED INFUSION SET LEAKING AT THE TUBING EVENT ON (B)(6) 2024.THE INFUSION SET HAS BEEN USED FOR 2 DAYS. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20407846 ↗MalfunctionOct 4, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN COLOMBIA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET TUBING DETACHMENT EVENT ON 04-SEP-2024. THE LOCATION OF DETACHMENT WAS AT SITE. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20368404 ↗
Other Set, Administration, Intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page