Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

All models of Autosoft90 (21)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Unomedical Autosoft90 infusion set has 517 adverse event reports, with no deaths and 43 injuries. Many reports describe tubing that is bent or twisted, parts that do not stay bonded, sets falling off or separating during use, and leaks. Some describe high blood sugar or diabetic ketoacidosis, and others describe blocked tubing. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
511
Deaths reported
0
Injuries reported
50
Recalls naming it
0

Summary

The FDA's MAUDE database contains 511 adverse event reports that list this exact device (DI 05705244018242), received between Feb 1, 2021 and Sep 29, 2026 — classified as 50 injuries, 461 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (266 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 49 reportsJun 2024: 55 reportsJul 2024: 55 reportsAug 2024: 29 reportsSep 2024: 15 reportsOct 2024: 31 reportsNov 2024: 6 reportsDec 2024: 10 reportsJan 2025: 30 reportsFeb 2025: 30 reportsMar 2025: 17 reportsApr 2025: 32 reportsMay 2025: 10 reportsJun 2025: 2 reportsJul 2025: 5 reportsAug 2025: 6 reportsSep 2025: 5 reportsOct 2025: 0 reportsNov 2025: 7 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 6 reportsApr 2026: 16 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 53 reportsSep 2026: 18 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202655
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent266
  • Loss of or Failure to Bond108
  • Premature Separation61
  • Leak/Splash44
  • Infusion or Flow Problem32
  • Detachment of Device or Device Component29
  • Device Handling Problem12
  • Complete Blockage6

Patient problems

  • No Clinical Signs, Symptoms or Conditions351
  • Hyperglycemia107
  • Elevated Ketones/ Diabetic Ketoacidosis32
  • Hemorrhage/Blood Loss/Bleeding9
  • Skin Inflammation/ Irritation7
  • Pain3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Nausea1

Recent adverse event reports

The 25 most recent of 511 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Use of Incorrect Control/Treatment Settings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932096 ↗
  • InjurySep 27, 2026Complete Blockage; Material Twisted/Bent

    It was reported that patient faced an issue with kinked cannula event on (b)(6)2026 and also had high blood glucose levels corrected via pump.

    Patient problems: Hyperglycemia

    MAUDE report 50909877 ↗
  • MalfunctionSep 24, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885920 ↗
  • MalfunctionSep 24, 2026Loss of or Failure to Bond; Premature Separation

    The patient reported that infusion set that fell off during use on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885916 ↗
  • MalfunctionSep 24, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885912 ↗
  • MalfunctionSep 24, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885911 ↗
  • MalfunctionSep 24, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885903 ↗
  • MalfunctionSep 23, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient faced an event where the infusion set which was in use for one day fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863615 ↗
  • MalfunctionSep 23, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient faced an event where the infusion set which was in use for one day fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863612 ↗
  • MalfunctionSep 23, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient faced an event where the infusion set which was in use for one day fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863610 ↗
  • MalfunctionSep 23, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient faced an event where the infusion set which was in use for one day fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863607 ↗
  • MalfunctionSep 23, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient faced an event where the infusion set which was in use for one day fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863588 ↗
  • MalfunctionSep 20, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50834581 ↗
  • MalfunctionSep 20, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50834566 ↗
  • MalfunctionSep 20, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50834562 ↗
  • MalfunctionSep 4, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50619636 ↗
  • MalfunctionSep 4, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50619617 ↗
  • MalfunctionSep 2, 2026Partial Blockage; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50557502 ↗
  • MalfunctionAug 31, 2026Leak/Splash

    It was reported that patient faced infusion set leakage event on (b)(6) 2025. The leakage was at the site. The infusion set was in use for two to three hours.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50518046 ↗
  • MalfunctionAug 20, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient experienced seven infusion sets came off during use event on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50336584 ↗
  • MalfunctionAug 17, 2026Leak/Splash

    Patient problems: Skin Inflammation/ Irritation

    MAUDE report 50280547 ↗
  • MalfunctionAug 17, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50280545 ↗
  • MalfunctionAug 17, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50280544 ↗
  • MalfunctionAug 17, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50280542 ↗
  • MalfunctionAug 17, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50280540 ↗

FDA clearance and classification

The FDA regulates product code FPA (set, administration, intravascular) under 21 CFR 880.5440 as a Class II device. 21 CFR 880.5440 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the Autosoft90™ 1002821

Has the Autosoft90™ 1002821 been recalled?

No FDA recall record in our data names this exact device identifier (DI 05705244018242), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Autosoft90™ 1002821?

The FDA's MAUDE database lists 511 adverse event reports for this device, received between Feb 1, 2021 and Sep 29, 2026. In those reports, the most common specific device problems are “Material Twisted/Bent” (266), “Loss of or Failure to Bond” (108) and “Premature Separation” (61); the most common patient problems are “Hyperglycemia” (107), “Elevated Ketones/ Diabetic Ketoacidosis” (32) and “Hemorrhage/Blood Loss/Bleeding” (9). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Autosoft90™ 1002821?

None of the reports are classified as deaths, 50 as injuries and 461 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Autosoft90™ 1002821, and is it still sold?

The FDA's device database lists Unomedical A/S as the company responsible for it (brand Autosoft90™, model 1002821). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Set, Administration, Intravascular devices

All Set, Administration, Intravascular devices →