Autosoft90™ 1002822Recalls & reported problems
Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Autosoft90 infusion set delivers insulin under the skin. Reports describe its material twisting or bending, leaks, patches failing to stick, and insulin flow problems, with high blood sugar reported in some cases. There are 100 reports naming its identifier: 2 injuries and no deaths. Reports fell sharply to 14 in the last 12 months from 47 the year before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 100
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 100 adverse event reports that list this exact device (DI 05705244018273), received between Dec 17, 2021 and Aug 6, 2026 — classified as 2 injuries, 98 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (49 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent49
- Leak/Splash16
- Loss of or Failure to Bond13
- Infusion or Flow Problem9
- Detachment of Device or Device Component6
- Premature Separation5
- Deformation Due to Compressive Stress3
- Material Deformation3
Patient problems
- No Clinical Signs, Symptoms or Conditions57
- Hyperglycemia42
- Elevated Ketones/ Diabetic Ketoacidosis1
Recent adverse event reports
The 25 most recent of 100 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50152265 ↗MalfunctionAug 4, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50125197 ↗MalfunctionAug 4, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50125187 ↗InjuryApr 1, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT PATIENT FACED ADHESIVE PATCH AND CANNULA HOUSING DETACHED FROM SPRING PRIOR TO INSERTION ON (B)(6) 2026. THE PATIENT EXPERIENCED HIGH BLOOD GLUCOSE LEVELS AND RECEIVED TREATMENT WITH CORRECTION BOLUS VIA PUMP. THE PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY.
Patient problems: Hyperglycemia
MAUDE report 24753275 ↗MalfunctionMar 30, 2026Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT HAS ISSUE WITH AN INFUSION SET LEAKING AT THE SITE AFTER ABOUT TWENTY HOURS OF USE ON (B)(6) 2024. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24713271 ↗MalfunctionMar 21, 2026Loss of or Failure to Bond; Premature Separation
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFUSION SET ADHESIVE ISSUE EVENT ON (B)(6) 2026. THE INFUSION SET FELL OFF DURING USE WITHIN 10 MINUTES. THE ISSUE OCCURRED WITH TWO INFUSION SETS. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24657996 ↗MalfunctionMar 21, 2026Loss of or Failure to Bond; Premature Separation
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFUSION SET ADHESIVE ISSUE EVENT ON (B)(6) 2026. THE INFUSION SET FELL OFF DURING USE WITHIN 10 MINUTES. THE ISSUE OCCURRED WITH A TOTAL OF FOUR INFUSION SETS. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24657994 ↗MalfunctionMar 17, 2026Loss of or Failure to Bond; Premature Separation
REFERENCE NUMBER (B)(4) EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED ISSUE WITH THREE INFUSION SETS ON (B)(6) 2026 AS THE INFUSION SET PATCH DID NOT STICK TO THE SKIN UPON INSERTION. THE PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24615494 ↗MalfunctionMar 17, 2026Loss of or Failure to Bond; Premature Separation
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED ISSUE WITH THREE INFUSION SETS ON (B)(6) 2026 AS THE INFUSION SET PATCH DID NOT STICK TO THE SKIN UPON INSERTION. THE PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24615491 ↗MalfunctionMar 17, 2026Loss of or Failure to Bond; Premature Separation
REFERENCE NUMBER (B)(4) EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED ISSUE WITH THREE INFUSION SETS ON (B)(6) 2026 AS THE INFUSION SET PATCH DID NOT STICK TO THE SKIN UPON INSERTION. THE PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24615484 ↗MalfunctionFeb 11, 2026Deformation Due to Compressive Stress
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Patient problems: Hyperglycemia
MAUDE report 24311709 ↗MalfunctionFeb 11, 2026Deformation Due to Compressive Stress
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Patient problems: Hyperglycemia
MAUDE report 24311697 ↗MalfunctionFeb 11, 2026Deformation Due to Compressive Stress
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Patient problems: Hyperglycemia
MAUDE report 24311696 ↗MalfunctionNov 3, 2025Device Handling Problem; Device Handling Problem
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Patient problems: Hyperglycemia
MAUDE report 23449432 ↗MalfunctionSep 17, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN GERMANY. IT WAS REPORTED THAT THE PATIENT FACED AN INFUSION SET TUBING DETACHEMENT EVENT ON (B)(6) 2025. THE TUBING WAS DETACHED AT SITE CONNECTOR. INFUSION SET WAS USED FOR 3 TO 7 HOURS. BLOOD GLUCOSE LEVEL WAS 500 MG/DL AT THE TIME OF THE EVENT. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Hyperglycemia
MAUDE report 23071932 ↗MalfunctionAug 8, 2025Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INFUSION SETS TUBING BLOCKAGE EVENTS BETWEEN (B)(6) 2025. THE INFUSION SETS WERE IN USE FOR SOMETIMES FOR ONE DAY AND SOMETIMES FOR A COUPLE HOURS FOR EVERY EVENTS. THE BLOOD GLUCOSE LEVEL WAS HIGH AT THE TIME OF EVENTS AND THE PATIENT GOT TREATED WITH CORRECTION BOLUS VIA PUMP. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Hyperglycemia
MAUDE report 22746687 ↗MalfunctionAug 8, 2025Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. THIS EMDR IS BEING SUBMITTED FOR UNKNOWN NUMBER QUANTITY. IT WAS REPORTED THAT THE PATIENT FACED INFUSION SETS TUBING BLOCKAGE EVENTS BETWEEN (B)(6) 2025.THE INFUSION SETS WERE IN USE FOR SOMETIMES FOR ONE DAY AND SOMETIMES FOR A COUPLE HOURS FOR EVERY EVENTS. THE BLOOD GLUCOSE LEVEL WAS HIGH AT THE TIME OF EVENTS AND THE PATIENT GOT TREATED WITH CORRECTION BOLUS VIA PUMP. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Hyperglycemia
MAUDE report 22746686 ↗MalfunctionAug 5, 2025Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED AN INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. THE BLOCKAGE WAS IN THE TUBING. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22711071 ↗MalfunctionMay 23, 2025Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED SEVEN EVENTS WHERE THE INFUSION SET TUBING WAS LEAKING AT COUPLING HOUSING FROM (B)(6) 2025 TO (B)(6) 2025. THE INFUSION SET WAS IN USE FOR LESS THAN 48 HOURS. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22076142 ↗MalfunctionMay 23, 2025Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED SEVEN EVENTS WHERE THE INFUSION SET TUBING WAS LEAKING AT COUPLING HOUSING FROM (B)(6) 2025 TO (B)(6) 2025. THE INFUSION SET WAS IN USE FOR LESS THAN 48 HOURS. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22076140 ↗MalfunctionMay 23, 2025Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED SEVEN EVENTS WHERE THE INFUSION SET TUBING WAS LEAKING AT COUPLING HOUSING FROM (B)(6) 2025. THE INFUSION SET WAS IN USE FOR LESS THAN 48 HOURS. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22076139 ↗MalfunctionMay 23, 2025Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED SEVEN EVENTS WHERE THE INFUSION SET TUBING WAS LEAKING AT COUPLING HOUSING FROM (B)(6) 2025. THE INFUSION SET WAS IN USE FOR LESS THAN 48 HOURS. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22076138 ↗MalfunctionMay 23, 2025Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED SEVEN EVENTS WHERE THE INFUSION SET TUBING WAS LEAKING AT COUPLING HOUSING FROM (B)(6) 2025. THE INFUSION SET WAS IN USE FOR LESS THAN 48 HOURS. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22076137 ↗MalfunctionMay 23, 2025Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED SEVEN EVENTS WHERE THE INFUSION SET TUBING WAS LEAKING AT COUPLING HOUSING FROM (B)(6) 2025 TO (B)(6) 2025. THE INFUSION SET WAS IN USE FOR LESS THAN 48 HOURS. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22076136 ↗MalfunctionMay 23, 2025Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT EXPERIENCED SEVEN EVENTS WHERE THE INFUSION SET TUBING WAS LEAKING AT COUPLING HOUSING FROM (B)(6) 2025. THE INFUSION SET WAS IN USE FOR LESS THAN 48 HOURS. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22076135 ↗
Other Set, Administration, Intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.