MiniMed™ Sure T™ MMT-874ARecalls & reported problems

Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

All models of MiniMed Infusion Sets (58) · MiniMed Sure T (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This MiniMed Sure-T infusion set is a single-use tube set for subcutaneous insulin infusion. Its 98 adverse event reports include no deaths, 34 injuries and 64 malfunctions. The reports mostly describe the set separating early or failing to stick to the skin, and some patients report high blood sugar. Reports in the last 12 months number 43, up from 26 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
133
Deaths reported
0
Injuries reported
48
Recalls naming it
0

Summary

The FDA's MAUDE database contains 133 adverse event reports that list this exact device (DI 05705244019355), received between Nov 17, 2020 and Sep 30, 2026 — classified as 48 injuries, 85 malfunctions.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (44 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 16 reportsJul 2024: 4 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 7 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 10 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 16 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 23 reportsSep 2026: 33 reportsOct 2026: 0 reports20242025202633
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of or Failure to Bond44
  • Adverse Event Without Identified Device or Use Problem38
  • Premature Separation29
  • Detachment of Device or Device Component16
  • Infusion or Flow Problem11
  • Appropriate Device Problem Term/Code Not Available4
  • Complete Blockage4
  • Fluid/Blood Leak4

Patient problems

  • No Clinical Signs, Symptoms or Conditions79
  • Hyperglycemia31
  • Hypoglycemia14
  • Elevated Ketones/ Diabetic Ketoacidosis9
  • Diaphoresis1
  • Unspecified Infection1
  • Urinary Frequency / Polyuria1

Recent adverse event reports

The 25 most recent of 133 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951641 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950459 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient infusion set tape not sticking due to sweating working caused tape to fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942816 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient infusion set sweating working caused tape to fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942813 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient infusion set tape not sticking due to sweating working caused tape to fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942642 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient infusion set tape not sticking due to swimming on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942639 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient infusion set due to sweating working caused tape to fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942610 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient infusion set tape not sticking due to sweating working caused tape to fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942608 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient infusion set tape not sticking due to sweating working caused tape to fell off on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942607 ↗
  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that patient experienced hyperglycemia event on (b)(6)2026 which was treated with pen injection.

    Patient problems: Hyperglycemia

    MAUDE report 50942064 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the Patient faced an event where Tape Not Sticking and fell off due to sweating & working on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941787 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931098 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931095 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50929739 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50929738 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927691 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient experienced an infusion set tape fell off due to sweat on (B)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927684 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that patient experienced low blood glucose event on (b)(6) 2026 which was addressed by intake of apple juice along with dextrose energy clementines.

    Patient problems: Hypoglycemia

    MAUDE report 50911920 ↗
  • InjurySep 27, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Hyperglycemia

    MAUDE report 50910082 ↗
  • InjurySep 26, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Diaphoresis; Hyperglycemia; Urinary Frequency / Polyuria

    MAUDE report 50907195 ↗
  • InjurySep 26, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Hypoglycemia

    MAUDE report 50907193 ↗
  • InjurySep 26, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 50906851 ↗
  • MalfunctionSep 24, 2026Fluid/Blood Leak

    It was reported that the patient faced an event where leakage at site on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883268 ↗
  • MalfunctionSep 24, 2026Fluid/Blood Leak

    It was reported that the patient faced an event where leakage at site on (b)(6)2026. The set was used for one day.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883264 ↗
  • MalfunctionSep 24, 2026Fluid/Blood Leak

    It was reported that infusion set leaked at the insertion site after one day of use with no reported stress or pull on the tubing (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50877520 ↗

Other Set, Administration, Intravascular devices

All Set, Administration, Intravascular devices →