MiniMed™ Sure T™ MMT-886ARecalls & reported problems
Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2
All models of MiniMed Infusion Sets (58) · MiniMed Sure T (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This MiniMed Sure T infusion set delivers insulin under the skin for diabetes patients. There are 23 adverse event reports naming its identifier, with no deaths and 6 injuries. Reports mostly describe insulin flow blockages, the set detaching at the connector, and leaks, along with cases of high blood sugar and one hospitalization for low blood sugar. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 05705244019447), received between Sep 24, 2020 and Sep 30, 2026 — classified as 6 injuries, 25 malfunctions.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of or Failure to Bond8
- Premature Separation8
- Detachment of Device or Device Component7
- Infusion or Flow Problem7
- Leak/Splash4
- Adverse Event Without Identified Device or Use Problem3
- Crack1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- Hyperglycemia4
- Hypoglycemia2
- Abscess1
- Elevated Ketones/ Diabetic Ketoacidosis1
- Hemorrhage/Blood Loss/Bleeding1
- Tissue Breakdown1
- Unspecified Infection1
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation
It was reported by the patient that tape was not sticking on (b)(6) 2026.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948216 ↗MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation
It was reported that patient face issue with tape not sticking on (b)(6)2026.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943227 ↗MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation
It was reported that patient face issue with tape not sticking on (b)(6) 2026.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943095 ↗MalfunctionSep 18, 2026Loss of or Failure to Bond; Premature Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50827239 ↗MalfunctionSep 18, 2026Loss of or Failure to Bond; Premature Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50822392 ↗MalfunctionSep 16, 2026Loss of or Failure to Bond; Premature Separation
It was reported that the patient experienced an event where an Infusion set tape not sticking due to perspiration on 15 Aug 2026.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50783449 ↗InjurySep 13, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Hypoglycemia
MAUDE report 50747796 ↗MalfunctionSep 1, 2026Loss of or Failure to Bond; Premature Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50537923 ↗MalfunctionSep 1, 2026Loss of or Failure to Bond; Premature Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50536343 ↗InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem
It was reported that patient experienced high blood glucose event (300 mg/dL) on (b)(6) 2026 and it was addressed by manual bolus via pump.
Patient problems: Hyperglycemia
MAUDE report 50017348 ↗InjuryMar 23, 2026Adverse Event Without Identified Device or Use Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED DUE TO HYPOGLYCEMIA ON (B)(6) 2026. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Hypoglycemia
MAUDE report 24659045 ↗MalfunctionNov 17, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
REFERENCE NUMBER (B)(4) EVENT OCCURRED IN GERMANY. IT WAS REPORTED THAT PATIENT FACED AN INFUSION SET CANNULA NEEDLE BREAK EVENT ON (B)(6) 2025. THE EVENT OCCURRED DURING INSERTION. THE INSERTION SITE WAS ABDOMEN. THE INFUSION SET WAS IN USE FOR LESS THAN ONE DAY. THE PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23568105 ↗MalfunctionNov 13, 2025Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED AN INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. THE BLOCKAGE IS IN THE TUBING. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23540439 ↗InjuryNov 11, 2025Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025 DUE TO WHICH PATIENT HAD ISSUE WITH SITE AND GOT AN ABSCESS AND HAD TO DRAIN. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Abscess; Unspecified Infection
MAUDE report 23519477 ↗MalfunctionSep 12, 2025Infusion or Flow Problem; Infusion or Flow Problem
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23035338 ↗MalfunctionJun 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
REFERENCE NUMBER: (B)(4). EVENT OCCURRED IN THE UNITED KINGDOM. IT WAS REPORTED THAT PATIENT FACED INFUSION SET DETACHMENT EVENT ON 27 MAY-2025. THE SITE OF DETACHMENT WAS CONNECTOR. THE INFUSION SET WAS IN USE FOR 1 DAY. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22173207 ↗MalfunctionJun 4, 2025Crack; Crack
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN HUNGARY. IT WAS REPORTED THAT PATIENT FACED A CRIMPED INFUSION SET TUBING EVENT ON (B)(6) 2025 AT SET CONNECTOR. THE BLOOD GLUCOSE LEVEL WAS HIGH AND THE PATIENT WAS TREATED WITH SUBCUTANEOUS INSULIN INJECTION AFTER INFUSION SET CHANGE. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Hyperglycemia
MAUDE report 22139759 ↗MalfunctionMar 19, 2025Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN GERMANY. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKAGE EVENT ON (B)(6) 2025 DUE TO BLOCKAGE IN TUBING. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21640651 ↗MalfunctionDec 5, 2024Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN POLAND. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2024. THE BLOCKAGE WAS IN THE TUBING. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20852081 ↗MalfunctionSep 28, 2024Material Twisted/Bent; Material Twisted/Bent
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INFUSION SET INSULIN FLOW BLOCKED DUE TO BENT NEEDLE ON (B)(6) 2024.PATIENT ALSO EXPERIENCED BLEEDING AFTER REMOVING THE NEEDLE FROM THE BODY. THE INFUSION SET HAS BEEN USED FOR 1-2 DAYS. THE INSERTION SITE WAS ABDOMEN. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20328801 ↗MalfunctionSep 24, 2024Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT, THE PATIENT EXPERIENCED INSULIN FLOW BLOCK ALARM EVENT ON (B)(6) 2024.THE BLOCKAGE WAS LOCATED AT THE TUBING. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20290450 ↗MalfunctionAug 31, 2024Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES, IT WAS REPORTED THAT ON (B)(6) 2024 PATIENT REPORTED LEAKING FOR INFUSION SET AND RESERVOIR. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20113940 ↗MalfunctionAug 8, 2024Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE POLAND. IT WAS REPORTED THAT PATIENT FACED INFUSION SET BLOCKAGE IN TUBING EVENT ON 08-JUL-2024. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19936114 ↗MalfunctionJul 8, 2024Leak/Splash; Leak/Splash
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRIA. ON 07-JUN-2024, IT WAS REPORTED THAT THE PATIENT FACED INFUSION SET LEAKED FROM THE ADHESIVE. THE INFUSION SET WAS IN USE FOR ONE DAY. NO FURTHER INFORMATION AVAILABLE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19689161 ↗MalfunctionJun 26, 2024Leak/Splash; Leak/Splash
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRIA. IT WAS REPORTED THAT PATIENT EXPERIENCED THAT THE INFUSION SET WAS LEAKED AT INSERTION SITE. THE INFUSION SET WAS USED FOR ONE DAY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19613718 ↗
Other Set, Administration, Intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page