Minimed™ Mio™ MMT-941ARecalls & reported problems
Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2
All models of MiniMed Infusion Sets (58) · Minimed Mio (14)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 05705244019980), received between May 10, 2024 and Oct 8, 2024 — classified as 2 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Infusion or Flow Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem3
- Leak/Splash3
- Material Deformation2
- Loss of or Failure to Bond1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Elevated Ketones/ Diabetic Ketoacidosis2
- Hyperglycemia1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 8, 2024Infusion or Flow Problem; Infusion or Flow Problem
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. ON (B)(6) 2024, IT WAS REPORTED BY THE PATIENT THAT HE RECEIVE AN ALARM. IN TROUBLESHOOTING BLOCKAGE WAS FOUND IN THE TUBE. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20401130 ↗MalfunctionSep 26, 2024Leak/Splash; Leak/Splash
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. ON (B)(6) 2024, IT WAS REPORTED THAT THE PATIENT ENCOUNTERED INFUSION SET LEAKS IN THE TUBING SITE. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20308644 ↗MalfunctionJul 21, 2024Material Deformation; Material Deformation
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET TUBING BENT WITH WHITE MARKS EVENT ON 13-JUN-2024. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19788898 ↗MalfunctionJul 20, 2024Material Deformation; Material Deformation
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA IT WAS REPORTED THAT THE PATIENT FACED KINKED TUBING. THE SITE WAS PATIENT'S STOMACH. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19786622 ↗InjuryJul 14, 2024Material Twisted/Bent; Material Twisted/Bent
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA ON 13-JUNE-2024, IT WAS REPORTED BY THE PATIENT THAT INFUSION SET CANNULA WAS KINKED DUE TO WHICH HOSPITALIZATION OCCURS DUE TO HYPERGLYCEMIA AND HIGH KETONES. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis
MAUDE report 19737512 ↗MalfunctionJul 1, 2024Infusion or Flow Problem; Infusion or Flow Problem
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED AUSTRALIA. ON (B)(6) 2024, IT WAS REPORTED THAT THERE WAS OCCLUSION IN INFUSION SET WHICH PREVENTS THE FLOW OF INSULIN. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19641563 ↗MalfunctionJun 26, 2024Loss of or Failure to Bond; Loss of or Failure to Bond
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN EGYPT. ON (B)(6) 2024, IT WAS REPORTED BY THE PATIENT THAT INFUSION SET TAPE NOT STICKING WITHIN TWO DAYS OF USE. INFUSION SET FELL OFF AT THE TIME OF SWIMMING. INFUSION SET WAS PLACED IN ABDOMEN. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19611729 ↗InjuryMay 22, 2024Infusion or Flow Problem; Infusion or Flow Problem
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA ON (B)(6) 2024, IT WAS REPORTED BY THE PATIENT THAT HE WAS HOSPITALIZED FOR 7 HOURS DUE TO HYPERGLYCEMIA EPISODE AND HIGH KETONES. HE RECEIVED AN INFUSION FLOW BLOCK ALARM FROM TWO SETS. PATIENT ALSO REPORTING THAT HER PUMP IS NOT DELIVERING AUTO CORRECTION. HIGH BLOOD GLUCOSE LEVEL WAS 27 MMOL/L. IT WAS CORRECTED BY THE HOSPITALIZATION TREATMENT AND VIA BOLUS PUMP. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis
MAUDE report 19369275 ↗MalfunctionMay 13, 2024Leak/Splash; Leak/Splash
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. ON (B)(6) 2024, THE PATIENT REPORTED THAT THE INFUSION SET HAD BEEN LEAKED AND LOCATION OF LEAKAGE WAS AT SITE. PATIENT'S BLOOD GLUCOSE AT TIME OF INCIDENT WAS 20 MMOL/L. PATIENT WAS TREATED WITH MANUAL INJECTION. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Hyperglycemia
MAUDE report 19298970 ↗MalfunctionMay 10, 2024Leak/Splash; Leak/Splash
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. ON (B)(6) 2024, IT WAS REPORTED THAT THE PATIENT FACED A LEAKAGE OF INFUSION SET. THE INFUSION SET WAS USED FOR 1-2 DAYS AND SITE WAS PATIENT'S TUMMY. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19283981 ↗
Other Set, Administration, Intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page