Skip to content

Minimed™ Mio™ MMT-921A600Recalls & reported problems

Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

All models of MiniMed Infusion Sets (58) · Minimed Mio (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05705244020214), received between Sep 23, 2021 and Oct 26, 2023 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Leak/Splash1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Elevated Ketones/ Diabetic Ketoacidosis1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 26, 2023Material Twisted/Bent; Material Twisted/Bent

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. ON 11-OCT-2023, IT WAS REPORTED BY PATIENT'S MOTHER THAT HER DAUGHTER FACED A BENT CANNULA DUE TO WHICH SHE EXPERIENCED HIGH BLOOD GLUCOSE LEVEL. THEREFORE, ON (B)(6) 2023, SHE WENT TO EMERGENCY ROOM DUE TO HIGH BLOOD GLUCOSE LEVEL. DURING HOSPITALIZATION, THE PATIENT RECEIVED FLUIDS OF SALINE (UNKNOWN) AND INSULIN. CURRENTLY, THE PATIENT IS ON MULTIPLE DAILY INJECTIONS. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 18009774 ↗
  • MalfunctionJan 23, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    UNOMEDICAL REFERENCE NUMBER: (B)(4). EVENT OCCURRED IN CANADA. ON (B)(6) 2022, IT WAS REPORTED THAT WHILE THE PATIENT WAS SLEEPING, THE INFUSION SET'S TUBING DETACHED AT THE QUICK RELEASE. THE SITE LOCATION WAS PATIENT'S LEFT THIGH AND THE PUMP WAS CLIPPED TO PATIENT'S UNDERWEAR. THE INFUSION HAD BEEN INSERTED ON SATURDAY. REPORTEDLY, THE INFUSIONS WERE STORED IN THE KITCHEN CUPBOARD. THERE WAS NO STRESS OR PULL ON THE TUBING AND THE PUMP WAS NOT DROPPED WITH THE SET CONNECTED TO PATIENT'S BODY. ACCORDING TO THE COMPLAINT INVESTIGATION PERFORMED ON THE RETURNED USED DEVICE (1 SET), IT WAS FOUND THAT THE TUBING WAS DETACHED FROM THE TUBING CONNECTOR. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16222533 ↗
  • MalfunctionDec 26, 2022Leak/Splash; Leak/Splash

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. ON (B)(6) 2022, THE PATIENT REPORTED THAT SHE IS PREGNANT, AND HER INFUSION SET'S SITE WAS NOT STICKING. THEREFORE, THE INFUSION SET'S SITE WAS LEAKING, AND THE SITE LOCATION WAS PATIENT'S LOWER BACK. HER BLOOD GLUCOSE LEVEL WAS WITHIN RANGE. THIS ISSUE OCCURRED WITH EIGHT INFUSION SETS. MOREOVER, THE INFUSION SET HAD BEEN USED FOR FOUR DAYS. REPORTEDLY, THERE WAS NO STRESS OR PULL ON THE INFUSION SET. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16047294 ↗
  • MalfunctionSep 12, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. ON (B)(6) 2022, THE PATIENT REPORTED THAT THE INFUSION SET'S TUBING CAME APART AT THE TUBING CONNECTOR WHILE HE WAS INSERTING THE INFUSION SET. THE SITE LOCATION WAS PATIENT'S ABDOMEN, AND THE PUMP WAS LOCATED ON THE OTHER SIDE OF THE ABDOMEN. MOREOVER, THE INFUSION HAD BEEN USED FOR ONE DAY. REPORTEDLY, THE INFUSIONS WERE NOT STORED OR USED IN A PLACE WHERE THEY MIGHT HAVE BEEN EXPOSED TO EXTREME TEMPERATURES AND HUMIDITY. THERE WAS NO STRESS OR PULL ON THE TUBING AND THE PUMP WAS NOT DROPPED WITH SET CONNECTED TO PATIENT'S BODY. NO FURTHER INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15400892 ↗
  • MalfunctionSep 23, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN (B)(6). ON (B)(6) 2021, IT WAS REPORTED THAT THE PATIENT'S INFUSION SET'S TUBING CAME APART. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12517765 ↗

Other Set, Administration, Intravascular devices

All Set, Administration, Intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page