Minimed™ Mio™ MMT-941A600Recalls & reported problems

Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

All models of MiniMed Infusion Sets (58) · Minimed Mio (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a MiniMed Mio MMT-941A600 infusion set, a single-use set placed under the skin to deliver insulin. The 30 reports mostly describe the set leaking at the insertion site or where it clips to the pump, coming loose early, or adhesive not holding (some attributed to sleeping or showering). All 30 describe malfunctions with no injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 05705244020306), received between May 18, 2024 and Apr 17, 2026 — classified as 30 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 13 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 6 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 6 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash20
  • Premature Separation6
  • Loss of or Failure to Bond3
  • Infusion or Flow Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions30

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2026Premature Separation

    IT WAS REPORTED THAT PATIENT EXPERIENCED INFUSION SET DETACHMENT ON (B)(6) 2026, ATTRIBUTED TO SLEEPING AND SHOWERING AFFECTING ADHESIVE INTEGRITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912654 ↗
  • MalfunctionApr 17, 2026Premature Separation

    IT WAS REPORTED THAT PATIENT EXPERIENCED INFUSION SET DETACHMENT ON (B)(6) 2026, ATTRIBUTED TO SLEEPING AND SHOWERING AFFECTING ADHESIVE INTEGRITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912544 ↗
  • MalfunctionApr 17, 2026Premature Separation

    IT WAS REPORTED THAT PATIENT EXPERIENCED INFUSION SET DETACHMENT ON (B)(6) 2026, ATTRIBUTED TO SLEEPING AND SHOWERING AFFECTING ADHESIVE INTEGRITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912440 ↗
  • MalfunctionApr 17, 2026Premature Separation

    IT WAS REPORTED THAT PATIENT EXPERIENCED INFUSION SET DETACHMENT ON (B)(6) 2026, ATTRIBUTED TO SLEEPING AND SHOWERING AFFECTING ADHESIVE INTEGRITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912437 ↗
  • MalfunctionApr 17, 2026Premature Separation

    IT WAS REPORTED THAT PATIENT EXPERIENCED INFUSION SET DETACHMENT ON (B)(6) 2026, ATTRIBUTED TO SLEEPING AND SHOWERING AFFECTING ADHESIVE INTEGRITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912348 ↗
  • MalfunctionApr 17, 2026Premature Separation

    IT WAS REPORTED THAT PATIENT EXPERIENCED INFUSION SET DETACHMENT ON (B)(6) 2026, ATTRIBUTED TO SLEEPING AND SHOWERING AFFECTING ADHESIVE INTEGRITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912148 ↗
  • MalfunctionMar 26, 2026Leak/Splash

    REFERENCE NUMBER: (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKAGE EVENT ON 09-NOV-2023. THE LEAKAGE WAS AT THE SITE. THE INFUSION SET WAS IN USE FOR ONE AND A HALF DAY. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24701467 ↗
  • MalfunctionJun 13, 2025Leak/Splash; Leak/Splash

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKAGE FROM 6 OUT OF 10 INFUSION SETS ON (B)(6) 2024. THE LEAKAGE WAS FROM THE BOTTOM THAT CLIPS TO THE PUMP. THE INFUSION SET WAS IN USE FOR NOT A COMPLETE DAY. NO FURTHER INFORMATION IS AVALIABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22200511 ↗
  • MalfunctionJun 13, 2025Leak/Splash; Leak/Splash

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKAGE FROM 6 OUT OF 10 INFUSION SETS ON (B)(6) 2024. THE LEAKAGE WAS FROM THE BOTTOM THAT CLIPS TO THE PUMP. THE INFUSION SET WAS IN USE FOR NOT A COMPLETE DAY NO FURTHER INFORMATION IS AVALIABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22200509 ↗
  • MalfunctionJun 13, 2025Leak/Splash; Leak/Splash

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKAGE FROM 6 OUT OF 10 INFUSION SETS ON (B)(6) 2024. THE LEAKAGE WAS FROM THE BOTTOM THAT CLIPS TO THE PUMP. THE INFUSION SET WAS IN USE FOR NOT A COMPLETE DAY. NO FURTHER INFORMATION IS AVALIABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22200506 ↗
  • MalfunctionJun 13, 2025Leak/Splash; Leak/Splash

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKAGE FROM 6 OUT OF 10 INFUSION SETS ON (B)(6) 2024. THE LEAKAGE WAS FROM THE BOTTOM THAT CLIPS TO THE PUMP. THE INFUSION SET WAS IN USE FOR NOT A COMPLETE DAY. NO FURTHER INFORMATION IS AVALIABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22200505 ↗
  • MalfunctionJun 13, 2025Leak/Splash; Leak/Splash

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKAGE FROM 6 OUT OF 10 INFUSION SETS ON (B)(6) 2024. THE LEAKAGE WAS FROM THE BOTTOM THAT CLIPS TO THE PUMP. THE INFUSION SET WAS IN USE FOR NOT A COMPLETE DAY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22200503 ↗
  • MalfunctionJun 13, 2025Leak/Splash; Leak/Splash

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKAGE FROM 6 OUT OF 10 INFUSION SETS ON (B)(6) 2024. THE LEAKAGE WAS FROM THE BOTTOM THAT CLIPS TO THE PUMP. THE INFUSION SET WAS IN USE FOR NOT A COMPLETE DAY. NO FURTHER INFORMATION IS AVALIABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22200502 ↗
  • MalfunctionOct 7, 2024Infusion or Flow Problem; Infusion or Flow Problem

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA, IT WAS REPORTED THAT ON (B)(6) 2024 OCCLUSION ALARM OCCURRED AND INSULIN SLOWLY EXIT THROUGH TUBING AFTER TROUBLESHOOTING. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20381395 ↗
  • MalfunctionJun 21, 2024Loss of or Failure to Bond; Loss of or Failure to Bond

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE CANADA. IT WAS REPORTED ISSUES WITH 3 INFUSION SETS WHERE 2 INFUSION SETS FELL OFF AFTER 2 HOURS AND THE THIRD ONE FELL OFF AFTER 24 HOUR. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19590414 ↗
  • MalfunctionJun 21, 2024Loss of or Failure to Bond; Loss of or Failure to Bond

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE CANADA. IT WAS REPORTED ISSUES WITH 3 INFUSION SETS WHERE 2 INFUSION SETS FELL OFF AFTER 2 HOURS AND THE THIRD ONE FELL OFF AFTER 24 HOUR. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19590410 ↗
  • MalfunctionJun 21, 2024Loss of or Failure to Bond; Loss of or Failure to Bond

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE CANADA. IT WAS REPORTED ISSUES WITH 3 INFUSION SETS WHERE 2 INFUSION SETS FELL OFF AFTER 2 HOURS AND THE THIRD ONE FELL OFF AFTER 24 HOUR. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19590405 ↗
  • MalfunctionMay 27, 2024Leak/Splash; Leak/Splash

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE CANADA ON 26-APR-2024, IT WAS REPORTED BY THE PATIENT THAT FOUR INFUSION SET WAS LEAKING AT 2AM THIS MORNING FROM THE SITE. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19402436 ↗
  • MalfunctionMay 27, 2024Leak/Splash; Leak/Splash

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE CANADA. ON (B)(6) 2024, IT WAS REPORTED BY THE PATIENT THAT FOUR INFUSION SET WAS LEAKING AT 2AM THIS MORNING FROM THE SITE. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19402435 ↗
  • MalfunctionMay 27, 2024Leak/Splash; Leak/Splash

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE CANADA. ON (B)(6) 2024, IT WAS REPORTED BY THE PATIENT THAT FOUR INFUSION SET WAS LEAKING AT 2AM THIS MORNING FROM THE SITE. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19402434 ↗
  • MalfunctionMay 27, 2024Leak/Splash; Leak/Splash

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE CANADA ON (B)(6) 2024, IT WAS REPORTED BY THE PATIENT THAT FOUR INFUSION SET WAS LEAKING AT 2AM THIS MORNING FROM THE SITE. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19402433 ↗
  • MalfunctionMay 23, 2024Leak/Splash; Leak/Splash

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT FACED THREE INFUSION SET LEAKING FROM SITE ON (B)(6) 2024. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19379104 ↗
  • MalfunctionMay 23, 2024Leak/Splash; Leak/Splash

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA IT WAS REPORTED THAT PATIENT FACED THREE INFUSION SET LEAKING FROM SITE ON (B)(6) 2024 NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19379103 ↗
  • MalfunctionMay 23, 2024Leak/Splash; Leak/Splash

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT PATIENT FACED THREE INFUSION SET LEAKING FROM SITE ON (B)(6) 2024 NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19379102 ↗
  • MalfunctionMay 19, 2024Leak/Splash; Leak/Splash

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE CANADA. ON 17 APR 2024, IT WAS REPORTED THAT SIX OUT OF TEN INFUSION SETS HAVE FAILED TO LAST LESS THAN A DAY DUE TO LEAKAGE. INFUSION SETS HAVE BEEN IN USE FOR HOURS NOT FOR A COMPLETE DAY. THE SETS HAVE BEEN LEAKING OUT FROM THE BOTTOM THAT CLIPS TO THE PUMP AND THE PATIENT STATES THE TUBING DID NOT COME APART. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19348403 ↗

Other Set, Administration, Intravascular devices

All Set, Administration, Intravascular devices →