Skip to content

Minimed™ Mio™ MMT-945A600Recalls & reported problems

Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

All models of MiniMed Infusion Sets (58) · Minimed Mio (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05705244020368), received between Dec 8, 2021 and Nov 14, 2024 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Material Twisted/Bent2
  • Infusion or Flow Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Hyperglycemia2
  • Elevated Ketones/ Diabetic Ketoacidosis1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN ITALY. ON (B)(6)2024, PATIENT FACED INFUSION SET TUBING DETACHMENT. SITE OF INSERTION WAS ABDOMEN. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20682512 ↗
  • InjuryOct 8, 2024Material Twisted/Bent; Material Twisted/Bent

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. THIS MDR IS BEING SUBMITTED FOR AN UNKNOWN NUMBER OF DEVICES IN WHICH THE ISSUE WAS EXPERIENCED. ON (B)(6) 2024, IT WAS REPORTED THAT PATIENT FACED INFUSION SET CANNULA BENT. PATIENT WAS ADMITTED TO HOSPITAL DUE TO ANY BLOOD GLUCOSE VALUE, DIABETIC KETOACIDOSIS SEIZURE AND DIABETIC COMA. LENGTH OF HOSPITALIZATION WAS FOR TWO NIGHTS. DATE OF HOSPITALIZATION WAS (B)(6) 2024.THE BLOOD GLUCOSE LEVEL WAS AT TIME OF HOSPITALIZATION WAS 22.2 MMOL/L. NAUSEA, DEHYDRATION HEADACHES SYMPTOMS WERE REPORTED AT TIME OF HOSPITALIZATION. THEREFORE, PATIENT WAS TREATED WITH INTRAVENOUS. COMPANY DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 20396552 ↗
  • MalfunctionSep 24, 2024Infusion or Flow Problem; Infusion or Flow Problem

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. ON (B)(6) 2024, IT WAS REPORTED THAT THE PATIENT ENCOUNTERED INFUSION SET OCCLUSION EVENT. THE BLOOD GLUCOSE WAS REPORTED 20 MMOL/L AND TREATED WITH BOLUS VIA PUMP. NO FURTHER INFORMATION.

    Patient problems: Hyperglycemia

    MAUDE report 20292576 ↗
  • MalfunctionSep 12, 2024Material Twisted/Bent; Material Twisted/Bent

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. IT WAS REPORTED ON (B)(6) 2024 THAT THE PATIENT ENCOUNTERED INFUSION SET CANNULA WAS KINKED WHICH RESULTS INTO HIGH BLOOD GLUCOSE LEVEL OF 27MMOL/L. DURATION OF INFUSION SET USED WAS WITHIN 48 HOURS. THE ISSUE GOT RESOLVED BY REPLACING THE INFUSION SET. COMPANY DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Hyperglycemia

    MAUDE report 20209121 ↗
  • MalfunctionJun 1, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA ON (B)(6) 2023, IT WAS REPORTED THAT THE PATIENT INSERTED THE INFUSION SET ON (B)(6) 2023 AND IT CAME APART ON (B)(6) 2023 AT THE SITE CONNECTOR WHILE HE WAS SLEEPING. THE SITE LOCATION WAS PATIENT'S TUMMY. THE INFUSION HAD BEEN USED FOR THREE DAYS. REPORTEDLY, THE PATIENT WAS UNSURE WHETHER THERE WAS STRESS OR PULL ON THE TUBING. FURTHER, THE PUMP WAS NOT DROPPED WITH THE SET CONNECTED TO PATIENT'S BODY. IN THE MORNING ((B)(6) 2023), HIS BLOOD GLUCOSE LEVEL WAS 11 MMOL/L. NO FURTHER INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17036853 ↗
  • MalfunctionDec 8, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN (B)(6). ON (B)(6) 2021, THE PATIENT'S FATHER REPORTED THAT HIS SON HAD ANOTHER INSTANCE OF TUBING DETACHMENT LAST NIGHT WHILE SLEEPING. THE PUMP WAS IN A SOFT TOY WHICH HE CUDDLES. THE INFUSION SET HAD BEEN USED LESS THAN A DAY AND THEY ARE IN A CLOSED CUPBOARD IN HIS BEDROOM. FURTHER, THE PUMP WAS NOT DROPPED AND THERE WAS NO STRESS ON THE TUBING. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12959066 ↗

Other Set, Administration, Intravascular devices

All Set, Administration, Intravascular devices →

Sources and updates