Quick-set® Paradigm® MMT-399T (UDI 05705244020436)Recalls & reported problems

Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

All models of Quick-set Paradigm (23)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05705244020436), received between Aug 14, 2024 and Jun 26, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Infusion or Flow Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Hypoglycemia1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN ISRAEL IT WAS REPORTED THAT PATIENT FACED INFUSION SET CANNULA BREAK EVENT ON (B)(6) 2024. THE INFUSION SET WAS IN USE FOR ONE DAY. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22333921 ↗
  • InjuryApr 17, 2025Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem; Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN ISRAEL. IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN EVENT OF LOW BLOOD GLUCOSE RESULTING IN HOSPITALIZATION ON (B)(6) 2025. LENGTH OF HOSPITALIZATION WAS GREATER THAN 24 HOURS. TREATMENT FOR LOW BLOOD GLUCOSE WAS USING SYRINGE TO INCREASE SUGAR. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Hypoglycemia

    MAUDE report 21854093 ↗
  • MalfunctionNov 29, 2024Infusion or Flow Problem; Infusion or Flow Problem

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN ISRAEL. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2024. THE BLOCKAGE WAS IN THE TUBING. NO FURTHER INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20808743 ↗
  • MalfunctionAug 23, 2024Infusion or Flow Problem; Infusion or Flow Problem

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE ISRAEL. ON 25/JUL/2024, IT WAS OBSERVED THAT THE OCCLUSION OCCURED THAT PREVENTS THE FLOW OF INSULIN DURING THERAPY. IT WAS CONFIRMED THAT THE CUSTOMER ABLE TO REMOVE THE SET AT THIS TIME TO TEST THE SITE PORTION OF THE INFUSION SET. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20058035 ↗
  • MalfunctionAug 14, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    REFERENCE NUMBER (B)(4). EVENT OCCURRED IN ISRAEL. ON (B)(6) 2024, IT WAS REPORTED THAT THE PATIENT FACED INFUSION SET CANNULA WAS BENT AND BREAK AT ABDOMEN. THE PATIENT ALSO REPORTED THAT THE CANNULA WAS STILL ON THE SKIN AND DID NOT RECEIVE ANY MEDICAL TREATMENT BECAUSE OF CANNULA FRACTURING WHILE IN THE BODY. THE INFUSION SET WAS IN USE FOR 4 DAYS. NO FURTHER INFORMATION AVAILABLE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19984940 ↗

Other Set, Administration, Intravascular devices