Medtronic Extended MMT-441ARecalls & reported problems

Unomedical A/S · Single use infusion set for subcutaneous infusion. The infusion set must be used as described in the Instruction for Use. · Class 2

All models of Medtronic Extended (27)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Extended infusion set is a single-use set for delivering fluid under the skin. Most of the 838 reports describe the adhesive failing and the set coming off early, and some mention leaks. A few describe high blood sugar after the set came loose. The reports include 72 injuries and no deaths. Reports from the past year number 498, up from 52 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,162
Deaths reported
1
Injuries reported
106
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,162 adverse event reports that list this exact device (DI 05705244022805), received between Jun 21, 2022 and Sep 30, 2026 — classified as 1 death, 106 injuries, 1,055 malfunctions.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (732 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 50 reportsJun 2024: 69 reportsJul 2024: 101 reportsAug 2024: 36 reportsSep 2024: 15 reportsOct 2024: 10 reportsNov 2024: 9 reportsDec 2024: 3 reportsJan 2025: 4 reportsFeb 2025: 2 reportsMar 2025: 6 reportsApr 2025: 3 reportsMay 2025: 10 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 7 reportsNov 2025: 9 reportsDec 2025: 7 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 66 reportsApr 2026: 64 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 7 reportsAug 2026: 349 reportsSep 2026: 318 reportsOct 2026: 0 reports202420252026349
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of or Failure to Bond732
  • Premature Separation608
  • Leak/Splash132
  • Adverse Event Without Identified Device or Use Problem72
  • Detachment of Device or Device Component67
  • Use of Incorrect Control/Treatment Settings45
  • Fluid/Blood Leak23
  • Material Twisted/Bent22

Patient problems

  • No Clinical Signs, Symptoms or Conditions992
  • Hyperglycemia110
  • Elevated Ketones/ Diabetic Ketoacidosis19
  • Skin Inflammation/ Irritation17
  • Hypoglycemia12
  • Hemorrhage/Blood Loss/Bleeding5
  • Skin Infection4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2

Recent adverse event reports

The 25 most recent of 1,162 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported by the patient that tape was not sticking on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948219 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported by the patient that tape was not sticking on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948202 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that patient experienced infusion set tape not sticking on (b)(6) 2026 due to perspiration.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947289 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that patient experienced infusion set tape not sticking on (b)(6) 2026 due to swimming.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947276 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that patient experienced infusion set tape not sticking event on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945043 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that patient experienced infusion set tape not sticking event on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945041 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient experienced an issue with infusion set on (b)(6) 2026 where the infusion set tape was not sticking due to perspiration.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944455 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient experienced an issue with infusion set on (b)(6)2026 where the infusion set tape was not sticking due to swimming.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944443 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient experienced an infusion set tape not sticking event due to perspiration on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944250 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient experienced an infusion set tape was not sticking after insertion event on (b)(6)2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944244 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that patient experienced an infusion set adhesive failure on (b)(6)2026, where the tape not sticking due to perspiration.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944047 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient experienced an infusion set tape not sticking event due to perspiration on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944041 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that patient faced an event where infusion set tape was not sticking on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943408 ↗
  • MalfunctionSep 30, 2026Fluid/Blood Leak; Detachment of Device or Device Component

    It was reported that patient faced and event of infusion set tubing came apart after two days of use on (b)(6)2026. Insertion site was right arm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942826 ↗
  • MalfunctionSep 30, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that the patient experienced tape detached due to hot weather condition on (b)(6)2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942298 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935790 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935789 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935361 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935193 ↗
  • MalfunctionSep 29, 2026Loss of or Failure to Bond; Premature Separation

    It was reported that patient experienced an event on (b)(6) 2026 in which infusion set tape was not sticking.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934401 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Hyperglycemia

    MAUDE report 50930328 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Hyperglycemia

    MAUDE report 50929978 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Hyperglycemia

    MAUDE report 50929964 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced hyperglycemia event on (b)(6)2026 and was treated with manual injection.

    Patient problems: Hyperglycemia

    MAUDE report 50929948 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia

    MAUDE report 50923470 ↗

Other Set, Administration, Intravascular devices

All Set, Administration, Intravascular devices →