aScope™ 3 Slim 3.8/1.2Flexible Videoscope - Single Use 402001000Recalls & reported problems
Ambu A/S · Ambu aScope 3 Slim - Video Bronchioscope Single-Use, Flexible and Sterile. Insertion cord dia 3.8 mm, channel width of 1.2 mm. Bending angle of 130°/130° Min. ordering 5 pcs. = 1 box. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05707480121960), received between Sep 8, 2017 and Sep 8, 2017 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Difficult to Remove1
- Sticking1
Patient problems
- Extubate1
- No Code Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 8, 2017Break; Difficult to Remove; Sticking
ASCOPE 3S WAS USED IN A VIDEO-ASSISTED INTUBATION PROCEDURE FOR PLACEMENT OF A 37FR DOUBLE LUMEN TUBE. ONCE THE DLT WAS PLACED IN THE PATIENT'S AIRWAY THE ANESTHESIOLOGIST WAS UNABLE TO WITHDRAW THE SCOPE FROM THE LUMEN OF THE DLT. THE SCOPE WAS STUCK IN THE TUBE AND WAS REMOVED WITH SIGNIFICANT FORCE. WHEN THE SCOPE WAS FROM THE DLT, THE ANESTHESIOLOGIST NOTICED THAT THE DISTAL END OF THE SCOPE HAD BROKEN OF AND THE BROKEN PART WAS STILL IN THE LUMEN OF THE DLT. A SECOND ASCOPE WAS USED TO CONFIRM THE PLACEMENT OF THE DISTAL TIP WITHIN THE LUMEN OF THE DLT. TO SOLVE THE SITUATION, THE DLT WAS WITHDRAWN FROM THE PATIENT AND PATIENT WAS RE-INTUBATED WITH A NEW TUBE.
Patient problems: Extubate; No Code Available
MAUDE report 6852181 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.